ORIGINAL ARTICLE
Year : 2021  |  Volume : 46  |  Issue : 1  |  Page : 7-15
 

Dosimetric evaluation of simplified knowledge-based plan with an extensive stepping validation approach in volumetric-modulated arc therapy-stereotactic body radiotherapy for lung cancer


1 Department of Radiation Oncology, Faculty of Medicine, Kindai University, Osaka, Japan
2 Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osakasayama, Osaka, Japan

Date of Submission30-Jul-2020
Date of Decision05-Mar-2021
Date of Acceptance05-Mar-2021
Date of Web Publication5-May-2021

Correspondence Address:
Mr. Hajime Monzen
377-2 Onohigashi, Osakasayama, Osaka 589-8511
Japan
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_67_20

Rights and Permissions

 

   Abstract 

Purpose: We investigated the performance of the simplified knowledge-based plans (KBPs) in stereotactic body radiotherapy (SBRT) with volumetric-modulated arc therapy (VMAT) for lung cancer. Materials and Methods: For 50 cases who underwent SBRT, only three structures were registered into knowledge-based model: total lung, spinal cord, and planning target volume. We performed single auto-optimization on VMAT plans in two steps: 19 cases used for the model training (closed-loop validation) and 16 new cases outside of training set (open-loop validation) for TrueBeam (TB) and Halcyon (Hal) linacs. The dosimetric parameters were compared between clinical plans (CLPs) and KBPs: CLPclosed, KBPclosed-TB and KBPclosed-Hal in closed-loop validation, CLPopen, KBPopen-TB and KBPopen-Hal in open-loop validation. Results: All organs at risk were comparable between CLPs and KBPs except for contralateral lung: V5 of KBPs was approximately 3%–7% higher than that of CLPs. V20 of total lung for KBPs showed comparable to CLPs; CLPclosed vs. KBPclosed-TB and CLPclosed vs. KBPclosed-Hal: 4.36% ± 2.87% vs. 3.54% ± 1.95% and 4.36 ± 2.87% vs. 3.54% ± 1.94% (P = 0.54 and 0.54); CLPopen vs. KBPopen-TB and CLPopen vs. KBPopen-Hal: 4.18% ± 1.57% vs. 3.55% ± 1.27% and 4.18% ± 1.57% vs. 3.67% ± 1.26% (P = 0.19 and 0.27). CI95 of KBPs with both linacs was superior to that of the CLP in closed-loop validation: CLPclosed vs. KBPclosed-TB vs. KBPclosed-Hal: 1.32% ± 0.12% vs. 1.18% ± 0.09% vs. 1.17% ± 0.06% (P < 0.01); and open-loop validation: CLPopen vs. KBPopen-TB vs. KBPopen-Hal: 1.22% ± 0.09% vs. 1.14% ± 0.04% vs. 1.16% ± 0.05% (P ≤ 0.01). Conclusions: The simplified KBPs with limited number of structures and without planner intervention were clinically acceptable in the dosimetric parameters for lung VMAT-SBRT planning.


Keywords: Knowledge-based plan, non-small cell lung cancer, plan quality, stereotactic body radiotherapy


How to cite this article:
Wada Y, Monzen H, Tamura M, Otsuka M, Inada M, Ishikawa K, Doi H, Nakamatsu K, Nishimura Y. Dosimetric evaluation of simplified knowledge-based plan with an extensive stepping validation approach in volumetric-modulated arc therapy-stereotactic body radiotherapy for lung cancer. J Med Phys 2021;46:7-15

How to cite this URL:
Wada Y, Monzen H, Tamura M, Otsuka M, Inada M, Ishikawa K, Doi H, Nakamatsu K, Nishimura Y. Dosimetric evaluation of simplified knowledge-based plan with an extensive stepping validation approach in volumetric-modulated arc therapy-stereotactic body radiotherapy for lung cancer. J Med Phys [serial online] 2021 [cited 2021 Jun 16];46:7-15. Available from: https://www.jmp.org.in/text.asp?2021/46/1/7/315509



   Introduction Top


Stereotactic body radiotherapy (SBRT) is an effective and noninvasive treatment for early-stage non-small cell lung cancer in patients who are medically inoperable or refuse surgery.[1] The local control rate at 3 years exceeds 90% with mild toxicity.[2],[3],[4] Intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT) can achieve higher conformity and have greater sparing of normal organs than conventional three-dimensional conformal radiotherapy (3D-CRT) with multiple noncoplanar beams.[5],[6],[7],[8],[9] The use of SBRT with IMRT/VMAT techniques is also well-tolerated and effective for lung tumors.[4],[10] However, differences in the skill and experience levels of planners and institutions can cause large variations in IMRT/VMAT plan quality.[11],[12],[13]

The use of a knowledge-based plan (KBP) can improve the consistency of treatment planning and reduce the variation of plan quality between planners or institutions.[14] A KBP model is devised to create KBPs based on previously delivered clinical plans (CLPs), including dose distributions and geometric characteristics such as the anatomical features of the target and organs at risk (OARs) and beam arrangement. This model trains the geometric characteristics and associated dose-volume histogram (DVH) of the plans included in the model using the combination of principal component analysis and regression techniques to predict the achievable DVH range for OARs of new patients. This range is produced by the mean estimated DVH ± one standard deviation (SD). Then, the KBP model generates dose-volume objectives below the lower boundary of the estimated DVH automatically, and the auto-optimization is performed.

The utility of KBP with IMRT/VMAT has been reported at various treatment regions (e.g., head and neck, breast, prostate, and hepatocellular cancer).[14],[15],[16],[17],[18],[19],[20],[21],[22],[23],[24],[25],[26],[27],[28] In SBRT of lung cancer, the doses to the lungs and heart can be reduced more in KBPs than CLPs.[29] However, some reports have trained KBP models using 9 or 10 structures.[10],[30] The use of many structures can cause large variations between planners because exact delineation of the same structures by different planners is challenging.[31] Thus, simplified KBPs to include only a limited number of structures and generating the plans with a single auto-optimization are crucial for the elimination of large between-planners' variation. The aim of this study was to investigate whether the simplified KBPs could be produced by a single auto-optimization and then used for lung SBRT in the clinical situations. We validated both closed- and open-loop KBP models to verify their usability and employed two different radiotherapy machines to compare their effects on plan quality.


   Materials and Methods Top


This study was approved by the Institutional Review Board (Approval No. 29-133) of the Kindai University Faculty of Medicine, in accordance with the Declaration of Helsinki. The requirement for informed consent was waived for all analyzed patients.

Clinical plans for training of the knowledge-based plan model

The computed tomography (CT) images and treatment plans of 50 sequential cases in which patients underwent SBRT for lung cancer at our institution between October 2013 and November 2018 were enrolled to train the KBP model. The patients' characteristics are shown in [Table 1]. The number of CLPs that employed 3D-CRT, IMRT, and VMAT were 9, 16, and 25, respectively. The locations of the lung tumors (e.g., left or right lung, central or peripheral) were not classified in the database. All patients received four-dimensional CT with 10 respiration phases for planning. The internal tumor volume (ITV) was evaluated on CT images in the every phase and delineated. The planning target volume (PTV) was defined as an automatic expansion by 5 mm from the ITV. The OARs of the total lung, ipsilateral lung for tumor, contralateral lung for tumor, heart, and esophagus, and planning organ at risk volume (PRV) added to the spinal cord were delineated by experienced radiation oncologists.
Table 1: Patients' characteristics in the training set, closed-loop validation group, and open-loop validation group

Click here to view


The CLPs prescribed 40–52 Gy in 4–5 fractions were mixed in the model; a higher priority assigned to homogeneity between October 2013 and April 2018 and a relatively high acceptable maximum dose to the PTV (up to 140% of the prescribed dose) between May 2018 and February 2020. Therefore, all plans were re-normalized to the prescribed dose of 48 Gy in 4 fractions covering 95% of the PTV using the Eclipse treatment planning system ver. 15.0 (Varian Medical Systems, Palo Alto CA, USA). All CLPs used Varian anisotropic analytical algorithm (AAA) for lung treatment planning clinically. This purpose is to unify the prescribed dose. The most CLPs were created by coplanar technique, and the all KBPs were created by coplanar technique. The CLPs were optimized for the TrueBeam linear accelerator (linac) (Varian Medical Systems, Palo Alto CA, USA), and the KBPs were optimized automatically for the TrueBeam and Halcyon linacs (Varian Medical Systems, Palo Alto CA, USA). The dose distributions in CLPs and KBPs were calculated with the AAA for the comparison with CLPs to eliminate any effect of the calculation algorithm.[32] The KBPs with both the TrueBeam and Halcyon used 6-MV flatting filter free beams. The beam and collimator angles of the CLPs were arranged to decrease the doses to the OARs. In contrast, to eliminate the subjectivity of the planners, the beam angles for the KBPs were determined using two full arcs for the TrueBeam (gantry angles: clockwise [181°–179°] and counterclockwise [179°–181°]; collimator angles: 30° and 330° for each arc) and four full arcs for the Halcyon (gantry angles: clockwise [181°–179°], counterclockwise [179°–181°], clockwise [181°–179°], and counterclockwise [179°–181°]; collimator angles: 281°, 326°, 11°, and 56° for each arc). The irradiation time with 4 arcs for Halcyon is almost equal to that with 2 arcs for the TrueBeam. The dose constraints for the PTV and OARs were based on the previous report, as shown in [Table 2].[33]
Table 2: Dosimetric constraints of planning target volume and organs at risk based on Japan clinical oncology group trial 1408

Click here to view


Closed- and open-loop knowledge-based plan model validations

The closed- and open-loop KBP model validations[34] were performed, as shown in [Figure 1]:
Figure 1: Schema of the validation method. (a) Closed-loop validation based on 19 volumetric-modulated arc therapy plans: clinical plan (CLP) vs. knowledge-based plan using the TrueBeam (KBPclosed-TB) vs. Halcyon (KBPclosed-Hal) for the cases enrolled in the database. (b) Open-loop validation based on 16 volumetric-modulated arc therapy plans: CLP vs. knowledge-based plan using the TrueBeam (KBPopen-TB) vs. Halcyon (KBPopen-Hal) for the cases not enrolled in the database

Click here to view


  1. Closed-loop validation: 19 VMAT plans in the database performed between August 2017 and November 2018 were used to generate the KBP. This step aimed to evaluate the ability to reproduce or improve the plans used for model training retrospectively.[34] The KBPs calculated with the TrueBeam and Halcyon linacs were denoted as KBPclosed-TB and KBPclosed-Hal, respectively
  2. Open-loop validation: 16 VMAT plans performed between January 2019 and February 2020 were enrolled. These were new cases that were not used in the database. This step aimed to test the capability of the KBP model to generate high-quality plans for new patient cases prospectively.[34] The KBPs calculated with the TrueBeam and Halcyon linacs were denoted as KBPopen-TB and KBPopen-Hal, respectively.


All KBPs were generated with a single auto-optimization using the KBP model. The patients' characteristics in each group are shown in [Table 1]. Only three structures were registered into the KBP model: the PTV, the total lung, and the PRV of the spinal cord. The objectives for all OARs were generated automatically [Table 3].
Table 3: Model objectives and priorities

Click here to view


The following parameters were compared between the CLPs and KBPs in the closed- and open-loop validations;

  1. Dose to 2% (D2%), 50% (D50%), and 98% (D98%) of the PTV
  2. Homogeneity index (HI) defined as the indicators of dose uniformity for the PTV (HI = 100× [D2%−D98%]/D50%)[35]
  3. Conformity index (CI95) defined as the indicators of dose convergence for the PTV (CI95 = V95%/VPTV)[36]
  4. Dmean and the volumes receiving 40 Gy (V40), 20 Gy (V20), 15 Gy (V15), and 5 Gy (V5) of the total lung
  5. Dmean and the volume receiving 20 Gy (V20) and 5 Gy (V5) of the ipsilateral and contralateral lungs
  6. Maximum dose (Dmax) to the spinal cord
  7. Volume receiving 30 Gy (V30) of the heart
  8. Volumes receiving 35 Gy (V35) and 40 Gy (V40) of the esophagus
  9. Ratio of the volume of the 50% of prescription isodose curve to PTV (R50)[37]
  10. Total monitor units (MU).


Statistical analysis

Data are reported as mean values ± one SD. The Mann–Whitney U-test was used to make the comparisons between the CLPs and KBPs. Statistical analysis was performed using StatFlex ver. 7 (Artec Inc., Osaka, Japan), and differences were considered statistically significant at P < 0.05.


   Results Top


Closed-loop knowledge-based plan model validation

[Table 4] compares the parameters between the CLPs and KBPs in closed-loop validation. The HI of the PTV for the KBPclosed-Hal was significantly higher than those of the CLP, but there was no significant difference between the KBPclosed-TB and CLP. The CI95 of the KBPclosed-TB and KBPclosed-Hal was significantly lower than that of the CLP. For the OARs, almost all dose-volume parameters were comparable between the CLP, KBPclosed-TB, and KBPclosed-Hal, whereas the V5 of the contralateral lung in the both KBPclosed-TB and KBPclosed-Hal was higher than those of the CLP. The R50 of KBPclosed-TB and KBPclosed-Hal was significantly lower than that of the CLP. The MU of the KBPclosed-TB was significantly higher than that of the CLP, while that of the KBPclosed-Hal was not significantly different from that of the CLP.
Table 4: Comparison of dose-volume parameters of the planning target volume, lung, spinal cord (PRV), and monitor unit between the clinical plan, knowledge-based plan closed-TrueBeam, and knowledge-based plan closed-Halcyon

Click here to view


Open-loop knowledge-based plan model validation

[Table 5] compares the parameters between the CLPs and KBPs in open-loop validation. [Figure 2] shows the dose distributions of the CLP, KBPopen-TB, and KBPopen-Hal for one selected patient. There were no significant differences in the HI of the PTV between the CLP, KBPopen-TB, and KBPopen-Hal. The CI95 of the KBPopen-TB and KBPopen-Hal was significantly lower than that of the CLP. Most dose-volume parameters of the OARs were comparable between the CLP, KBPopen-TB, and KBPopen-Hal, but Dmean and V5 of the contralateral lung were higher in both the KBPopen-TB and KBPopen-Hal than that in the CLP (shown in [Figure 2] as the extent of the low-dose distribution in both KBPs). In this patient, V5 of the contralateral lung in the CLP, KBPopen-TB, and KBPopen-Hal were 0.6%, 4.8%, and 6.4%, respectively. The CI95 of the CLP, KBPopen-TB, and KBPopen-Hal were 1.34, 1.16, and 1.19, respectively. The R50 of KBPopen-TB and KBPopen-Hal were significantly lower than that of the CLP. The MU of the KBPopen-TB was significantly higher than that of the CLP, but there was no significant difference between the KBPopen-Hal and CLP. A significant difference was observed only in the MU value between KBPopen-TB and KBPopen-Hal.
Table 5: Comparison of dose-volume parameters of the planning target volume, lung, spinal cord (PRV), and monitor unit between the clinical plan, knowledge-based plan open-TrueBeam, and knowledge-based plan open-Halcyon

Click here to view
Figure 2: Example dose distributions for the clinical plan (CLP) and knowledge-based plans using the TrueBeam and Halcyon in open-loop validation (KBPopen-TB and KBPopen-Hal). Compared with the clinical plans, both the KBPopen-TB and KBPopen-Hal had more extended low-dose distribution but higher conformity to the planning target volume

Click here to view


[Figure 3] compares the dose-volume parameters of the PTV between the CLPs and KBPs in closed- and open-loop validations. Many dose-volume parameters of the PTV showed significant differences between the CLPs and KBPs in closed-loop validation. The CLPs had significantly lower D50 and higher CI95 than the KBPs had in open-loop validation.
Figure 3: Comparisons of dose-volume parameters of the planning target volume between the clinical plans (CLPs) and knowledge-based plans (KBPs) using the TrueBeam and Halcyon in closed- and open-loop validations (KBPclosed-TB, KBPclosed-Hal, KBPopen-TB, and KBPopen-Hal, respectively). Central line of each box: median line; lower line: first quartile; upper line: third quartile. Whisker ranges do not contain outliers, which are plotted as individual points. N.S.; not significant, *P < 0.05, **P < 0.01, ***P < 0.001

Click here to view



   Discussion Top


In this study, we investigated whether the simplified KBPs using only three structures and a single auto-optimization could be used clinically to create lung VMAT-SBRT plans. Three structures (the PTV, peripheral lung region of the target, and spinal cord as serial organ) were employed for optimization of the KBPs to control the peripheral dose distribution of the target and decrease the maximum dose to the OARs.[22] The simplified KBPs were able to generate comparable plan quality to the CLPs, and the dosimetric parameters showed no major differences between the closed- and open-loop KBP model validations. In addition, the dose distribution with coplanar technique using the simplified KBPs for Halcyon linac could be applied to clinical practice as shown [Table 4] and [Table 5] and [Figure 2]. The locations of the lung tumors (e.g., left or right lung, central or peripheral) were not classified to simplify the KBP model in this study, although several KBP models have been created to perform classification by the location in some studies.[10] The dose distributions of the KBPs were not dependent on the linac for any parameters except for MU. Thus, the simplified KBPs using both radiotherapy machines in both loop validations could generate high-quality plans without any intervention by planners, as SBRT for lung cancer has few complex anatomical characters, such as almost no proximity or overlap between the PTV and OARs.[13],[15] We clarified that the simplified KBPs may have advantages in cases of treatment plans with few or no overlapping structures. KBP models that do not classify the ipsilateral and contralateral OARs or right and left target separately could be trained and established with little or no overlapping structures in SBRT or other precision treatments. Such simplified KBPs could be used clinically to create lung VMAT-SBRT plans.

The variations of the dose-volume parameters of the PTV in the KBPs produced in closed- and open-loop validation groups were lower than those in the CLPs, and the dose–volume parameters of the KBPs were similar between both validations. The planning policy regarding SBRT for lung cancer was not unified between the CLPs in this model, whereas all KBPs employed the same objective parameters in both validations. This suggests that the simplified KBPs minimized the variation in VMAT-SBRT plan quality. The HI of the PTV for the KBPclosed-Hal was significantly higher than those of the CLP; however, there was no significant difference between the KBPclosed-TB and CLP. The number of arcs could affect the plan quality. The numbers of arcs were 2-arcs and 4-arcs were used for TrueBeam and Halcyon in this study, respectively. Michiels et al. described the DVH parameters for the Halcyon with 3 arcs were superior to those for the TrueBeam with 2 arcs significantly.[38]

The V5 of the contralateral lung in the KBPs produced using both closed- and open-loop validations were higher than that of the CLPs because full arc VMAT was employed in the KBPs to eliminate planner subjectivity in the beam arrangement. Previous reports showed that the V5 of the lungs is unlikely to affect the grade of radiation pneumonitis.[39],[40] V20 is a more well-known risk factor for symptomatic radiation pneumonitis than V5.[41] The KBPs generated lower V20 than the CLPs in both the closed- and open-loop validations, although there was no significant difference. R50 is commonly used to evaluate the intermediate-to-low dose spread and dose falloff outside the target. The R50 of the KBPs was superior or comparable to that of the CLPs in closed- and open-loop validations, which indicated the KBPs were enough to suppress the spread of dose distribution.[37] Therefore, the simplified KBPs are considered to be applicable clinically even if full arc VMAT is employed.

The MU of the KBPs-TB was significantly higher than that of the CLPs and KBPs-Hal. The difference in the number of arcs (two- and four-arc VMAT for the TrueBeam and Halcyon linacs, respectively) and multi-leaf collimator type between the radiotherapy machines could increase the MU and the complexity of multi-leaf collimator motion. These have been described as the characteristics of KBP in several reports.[14],[42] On the other hand, the reduced MU ensures plan deliverability.[43] Thus, the increase of MU should be noted carefully when using KBP.

The simplified approach of KBP with minimized effects of planners' subjectivity (e.g., limited number of structures, no classification of lung tumor location, single auto-optimization, and full arc VMAT without beam arrangement) is crucial to reduce variation caused by differences in planners' or institutions' skill and experience levels. Snyder et al. showed the KBP with a full set of OARs could generate the superior or comparable plan quality to the CLP.[10] In this study, the simplified KBP can generate the comparable plans to the CLP without a full set of OARs in the model. This approach is expected to increase the ease of KBP model sharing and standardize plan quality between many institutions, including community medical organizations.


   Conclusions Top


In SBRT with full arc VMAT for lung cancer, the proposed simplified KBPs with a single auto-optimization are comparable or better than the corresponding CLPs. The simplified KBPs had low V20 of the total lung without deterioration of the dose to the PTV. The simplified KBPs can be employed effectively for the clinical use in many institutions because of their low variation of plan quality.

Acknowledgment

We would like to thank Richard Lipkin, PhD, from Edanz Group (https://en-author-services.edanzgroup.com/) for editing a draft of this manuscript.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Li H, Shen Y, Wu Y, Cai S, Zhu Y, Chen S, et al. Stereotactic body radiotherapy versus surgery for early-stage non-small-cell lung cancer. J Surg Res 2019;243:346-53.  Back to cited text no. 1
    
2.
Nagata Y, Negoro Y, Aoki T, Mizowaki T, Takayama K, Kokubo M, et al. Clinical outcomes of 3D conformal hypofractionated single high-dose radiotherapy for one or two lung tumors using a stereotactic body frame. J Radiat Oncol Biol Phys 2002;52:1041-6.  Back to cited text no. 2
    
3.
Timmerman R, Papiez L, McGarry R, Likes L, DesRosiers C, Frost S, et al. Results of a phase I study in medically inoperable stage I non-small cell lung cancer. Chest 2003;124:1946-55.  Back to cited text no. 3
    
4.
Yamashita H, Haga A, Takahashi W, Takenaka R, Imae T, Takenaka S, et al. Volumetric modulated arc therapy for lung stereotactic radiation therapy can achieve high local control rates. Radiat Oncol 2014;9:243.  Back to cited text no. 4
    
5.
Mix M, Tanny S, Nsouli T, Alden R, Chaudhari R, Kincaid R, et al. Outcomes following stereotactic body radiotherapy with intensity-modulated therapy versus three-dimensional conformal radiotherapy in early stage non-small cell lung cancer. Lung Cancer (Auckl) 2019;10:151-9.  Back to cited text no. 5
    
6.
Rauschenbach BM, Mackowiak L, Malhotra HK. A dosimetric comparison of three-dimensional conformal radiotherapy, volumetric-modulated arc therapy, and dynamic conformal arc therapy in the treatment of non-small cell lung cancer using stereotactic body radiotherapy. J Appl Clin Med Phys 2014;15:147-61.  Back to cited text no. 6
    
7.
Bai X, Shan G, Chen M, Wang B. Approach and assessment of automated stereotactic radiotherapy planning for early stage non-small-cell lung cancer. Biomed Eng Online 2019;18:101.  Back to cited text no. 7
    
8.
Chang JH, Poon I, Erler D, Zhang L, Cheung P. The safety and effectiveness of stereotactic body radiotherapy for central versus ultracentral lung tumors. Radiother Oncol 2018;129:277-83.  Back to cited text no. 8
    
9.
Nagai A, Shibamoto Y, Yoshida M, Inoda K, Kikuchi Y. Safety and efficacy of intensity-modulated stereotactic body radiotherapy using helical tomotherapy for lung cancer and lung metastasis. Biomed Res Int 2014;2014:473173.  Back to cited text no. 9
    
10.
Snyder K, Kim J, Reding A, Fraser C, Gordon J, Ajlouni M, et al. Development and evaluation of a clinical model for lung cancer patients using stereotactic body radiotherapy (SBRT) within a knowledge-based algorithm for treatment planning. J Appl Clin Med Phys 2016;17:263-75.  Back to cited text no. 10
    
11.
Williams MJ, Bailey MJ, Forstner D, Metcalfe PE. Multicentre quality assurance of intensity- modulated radiation therapy plans: A precursor to clinical trials. Australas Radiol 2007;51:472-9.  Back to cited text no. 11
    
12.
Bohsung J, Gillis S, Arrans R, Bakai A, Wagter CD, Knoos T, et al. IMRT treatment planning: A comparative inter-system and inter-centre planning exercise of the QUASIMODO group. Radiother Oncol 2005;76:354-61.  Back to cited text no. 12
    
13.
Hussein M, Heijmen BJM, Verellen D, Nisbet A. Automation in intensity modulated radiotherapy treatment planning – A review of recent innovations. Br J Radiol 2018;91:20180270.  Back to cited text no. 13
    
14.
Kubo K, Monzen H, Ishii K, Tamura M, Kawamorita R, Sumida I, et al. Dosimetric comparison of RapidPlan and manually optimized plans in volumetric modulated arc therapy for prostate cancer. Phys Med 2017;44:199-204.  Back to cited text no. 14
    
15.
Tamura M, Monzen H, Matsumoto K, Kubo K, Otsuka M, Inada M, et al. Mechanical performance of a commercial knowledge-based VMAT planning for prostate cancer. Radiat Oncol 2018;13:163.  Back to cited text no. 15
    
16.
Tol JP, Delaney AR, Dahele M, Slotman BJ, Verbakel WF. Evaluation of a knowledge-based planning solution for head and neck cancer. Int J Radiat Oncol Biol Phys 2015;91:612-20.  Back to cited text no. 16
    
17.
Fogliata A, Wang PM, Belosi F, Clivio A, Nicolini G, Vanetti E, et al. Assessment of a model based optimization engine for volumetric modulated arc therapy for patients with advanced hepatocellular cancer. Radiat Oncol 2014;9:236.  Back to cited text no. 17
    
18.
Wang J, Hu W, Yang Z, Chen X, Wu Z, Yu X, et al. Is it possible for knowledge-based planning to improve intensity modulated radiation therapy plan quality for planners with different planning experiences in left-sided breast cancer patients? Radiat Oncol 2017;12:85.  Back to cited text no. 18
    
19.
Uehara T, Monzen H, Tamura M, Ishikawa K, Doi H, Nishimura Y. Dose-volume histogram analysis and clinical evaluation of knowledge-based plans with manual objective constraints for pharyngeal cancer. J Radiat Res 2020;61:499-505.  Back to cited text no. 19
    
20.
Inoue E, Doi H, Monzen H, Tamura M, Inada M, Ishikawa K, et al. Dose-volume histogram analysis of knowledge-based volumetric-modulated arc therapy planning in postoperative breast cancer irradiation. In Vivo 2020;34:1095-101.  Back to cited text no. 20
    
21.
Tamura M, Monzen H, Matsumoto K, Kubo K, Ueda Y, Kamima T, et al. Influence of cleaned-up commercial knowledge-based treatment planning for volumetric modulated arc therapy of prostate cancer. J Med Phys 2020;45:71-7.  Back to cited text no. 21
  [Full text]  
22.
Kamima T, Ueda Y, Fukunaga J, Shimizu Y, Tamura M, Ishikawa K, et al. Multi-institutional evaluation of knowledge-based planning performance of volumetric modulated arc therapy ( VMAT ) for head and neck cancer. Phys Med 2019;64:174-81.  Back to cited text no. 22
    
23.
Ueda Y, Miyazaki M, Sumida I, Ohira S, Tamura M, Monzen H, et al. Knowledge-based planning for oesophageal cancers using a model trained with plans from a different treatment planning system. Acta Oncol 2020;59:274-83.  Back to cited text no. 23
    
24.
Ueda Y, Fukunaga J, Kamima T, Adachi Y, Nakamatsu K, Mozen H. Evaluation of multiple institutions' models for knowledge-based planning of volumetric modulated arc therapy (VMAT) for prostate cancer. Radiat Oncol 2018;13:46.  Back to cited text no. 24
    
25.
Ueda Y, Monzen H, Fukunaga J, Ohira S, Tamura M, Suzuki O, et al. Characterization of knowledge-based volumetric modulated arc therapy plans created by three different institutions' models for prostate cancer. Rep Pract Oncol Radiother 2020;25:1023-8.  Back to cited text no. 25
    
26.
Monzen H, Tamura M, Ueda Y, Fukunaga J, Kamima T, Muraki Y, et al. Dosimetric evaluation with knowledge-based planning created at different periods in volumetric-modulated arc therapy for prostate cancer: a multi-institution study. Radiol Phys Technol 2020;13:327-35.  Back to cited text no. 26
    
27.
Kubo K, Monzen H, Ishii K, Tamura M, Nakasaka Y, Kusawake M, et al. Inter-planner variation in treatment-plan quality of plans created with a knowledge-based treatment planning system. Phys Med 2019;67:132-40.  Back to cited text no. 27
    
28.
Ito T, Tamura M, Monzen H, Matsumoto K, Nakamatsu K, Harada T, et al. Impact of Aperture Shape Controller on Knowledge-based VMAT Planning of Prostate Cancer. Nihon Hoshasen Gijutsu Gakkai Zasshi 2021;77:23-31.  Back to cited text no. 28
    
29.
Kavanaugh JA, Holler S, DeWees TA, Robinson CG, Bradley JD, Iyengar P, et al. Multi-institutional validation of a knowledge-based planning model for patients enrolled in RTOG 0617: Implications for plan quality controls in cooperative group trials. Pract Radiat Oncol 2019;9:e218-27.  Back to cited text no. 29
    
30.
Hof SV, Delaney AR, Hilal T, Jos T, Slotman BJ, Senan S, et al. Knowledge-Based Planning for Identifying High-Risk Stereotactic Ablative Radiation Therapy Treatment Plans for Lung Tumors Larger Than 5 cm. Int J Radiat Oncol Biol Phys 2019;103:259-67.  Back to cited text no. 30
    
31.
Brouwer CL, Steenbakkers RJ, van den Heuvel E, Duppen JC, Navran A, Bijl HP, et al. 3D Variation in delineation of head and neck organs at risk. Radiat Oncol 2012;7:32.  Back to cited text no. 31
    
32.
Tajaldeen A, Ramachandran P, Alghamdi S, Geso M. On the use of AAA and AcurosXB algorithms for three different stereotactic ablative body radiotherapy (SABR) techniques: Volumetric modulated arc therapy (VMAT), intensity modulated radiation therapy (IMRT) and 3D conformal radiotherapy (3D-CRT). Rep Pract Oncol Radiother 2019;24:399-408.  Back to cited text no. 32
    
33.
Kawahara D, Ozawa S, Kimura T, Saito A, Nishio T, Nakashima T, et al. Marginal prescription equivalent to the isocenter prescription in lung stereotactic body radiotherapy: Preliminary study for Japan Clinical Oncology Group trial (JCOG1408). J Radiat Res 2017;58:149-54.  Back to cited text no. 33
    
34.
Castriconi R, Fiorino C, Broggi S, Cozzarini C, Muzio ND, Calandrino R, et al. Comprehensive intra-institution stepping validation of knowledge-based models for automatic plan optimization. Phys Med 2019;57:231-7.  Back to cited text no. 34
    
35.
Wu H, Jiang F, Yue H, Zhang H, Wang K, Zhang Y. Applying a RapidPlan model trained on a technique and orientation to another: A feasibility and dosimetric evaluation. Radiat Oncol 2016;11:108.  Back to cited text no. 35
    
36.
Hussein M, South CP, Barry MA, Adams EJ, Jordan TJ, Stewart AJ, et al. Clinical validation and benchmarking of knowledge-based IMRT and VMAT treatment planning in pelvic anatomy. Radiother Oncol 2016;120:473-9.  Back to cited text no. 36
    
37.
Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Brian K, et al. Stereotactic body radiation therapy: The report of AAPM Task Group 101. Med Phys 2010;37:4078-101.  Back to cited text no. 37
    
38.
Michiels S, Poels K, Crijns W, Delombaerde L, Roover RD, Vanstraelen B, et al. Volumetric modulated arc therapy of head-and-neck cancer on a fast-rotating O-ring linac: Plan quality and delivery time comparison with a C-arm linac. Radiother Oncol 2018;128:479-84.  Back to cited text no. 38
    
39.
Ueyama T, Arimura T, Takumi K, Nakamura F, Higashi R, Ito S, et al. Risk factors for radiation pneumonitis after stereotactic radiation therapy for lung tumours: Clinical usefulness of the planning target volume to total lung volume ratio. Br J Radiol 2018;91:20170453.  Back to cited text no. 39
    
40.
Badellino S, Muzio JD, Schivazappa G, Guarneri A, Ragona R, Bartoncini S, et al. No differences in radiological changes after 3D conformal vs VMAT-based stereotactic radiotherapy for early stage non-small cell lung cancer. Br J Radiol 2017;90:20170143.  Back to cited text no. 40
    
41.
Chun SG, Hu C, Choy H, Komaki RU, Timmerman RD, Schild SE, et al. Impact of intensity-modulated radiation therapy technique for locally advanced non-small-cell lung cancer: A secondary analysis of the NRG oncology RTOG 0617 randomized clinical trial. J Clin Oncol 2017;35:56-62.  Back to cited text no. 41
    
42.
Wall PD, Fontenot JD. Evaluation of complexity and deliverability of prostate cancer treatment plans designed with a knowledge-based VMAT planning technique. J Appl Clin Med Phys 2020;21:69-77.  Back to cited text no. 42
    
43.
Scaggion A, Fusella M, Agnello G, Bettinelli A, Pivato N, Roggio A, et al. Limiting treatment plan complexity by applying a novel commercial tool. J Appl Clin Med Phys 2020;21:27-34.  Back to cited text no. 43
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,284 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed435    
    Printed14    
    Emailed0    
    PDF Downloaded102    
    Comments [Add]    

Recommend this journal