Year : 2021  |  Volume : 46  |  Issue : 1  |  Page : 33-40
 

Establishment and dissemination of radiological standards in the field of diagnostic radiology in India


1 Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai, Maharashtra, India
2 School of Basic Medical Sciences, Department of Physics, Savitribai Phule Pune University, Pune, Maharashtra, India
3 Division of Remote Handling Robotics, Bhabha Atomic Research Centre, Mumbai, India

Date of Submission14-Oct-2020
Date of Decision21-Mar-2021
Date of Acceptance22-Mar-2021
Date of Web Publication5-May-2021

Correspondence Address:
Mrs. Vinatha Sumanth Panyam
Radiation Standards Section, Radiation Safety Systems Division, Bhabha Atomic Research Center, Trombay, Mumbai - 400 085, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_98_20

Rights and Permissions

 

   Abstract 

Background: The increased use of ionizing radiation for diagnostic purpose has resulted in an increase in the world population dose. Patient dosimetry in X-ray diagnostic radiology is required to establish diagnostic reference levels (DRLs) and to assess the average dose received by organs and tissues. International bodies have recommended DRLs to be based on dosimetric quantities. Aim: To cater to the increased requirement for dosimetry in diagnostic radiology, international guidelines are provided to establish and disseminate traceable calibration for dosimeters used in X-ray diagnostic radiology. X-ray diagnostic beams established are standardized using a diagnostic range free-air ionization chamber (DFAIC) (20-150 kV). Materials and Methods: Characterization of the DFAIC and determination of the correction factors for the air kerma measurements were evaluated experimentally and by theoretical calculations. Results: The paper details the establishment of 18 diagnostic beam qualities using DFAIC along with the associated uncertainties. The overall uncertainty for the air kerma measurements was within ±0.5% at 1 sigma level. Eight diagnostic range air kerma measurements using DFAIC were compared with the medium energy primary standard FAIC (50–300 kV) maintained in the laboratory. Conclusion: The air kerma rates agreed within ±1% and are within the overall standard uncertainty of both the chambers at the time of the comparison. Dissemination to the users in the field of diagnostic radiology in the country has been carried out by calibrating their ionization chambers and solid-state detector-based instruments against the DFAIC. The methodology followed to standardize the beams using DFAIC and calibration of dosimeters is presented in this work.


Keywords: Calibration, diagnostic radiology, free-air ionization chamber


How to cite this article:
Panyam VS, Rakshit S, Dhole SD, Patil BJ, Das AP, Kossery GA, Vilippalil S, Chaudhury P. Establishment and dissemination of radiological standards in the field of diagnostic radiology in India. J Med Phys 2021;46:33-40

How to cite this URL:
Panyam VS, Rakshit S, Dhole SD, Patil BJ, Das AP, Kossery GA, Vilippalil S, Chaudhury P. Establishment and dissemination of radiological standards in the field of diagnostic radiology in India. J Med Phys [serial online] 2021 [cited 2021 Jun 21];46:33-40. Available from: https://www.jmp.org.in/text.asp?2021/46/1/33/315513



   Introduction Top


The main component of the worldwide annual average dose from artificial sources of exposure is from medical diagnostic procedures. Since the early 1990s, there has been a significant rise (nearly 50%) in patients undergoing diagnostic medical examinations. With the advent of new and improved X-ray diagnostic tools such as computed tomography (CT), fluoroscopy, positron emission tomography, mammography, and other diagnostic tools, their use for diagnostic purpose has increased substantially.[1] Around 50% of the collective effective dose from X-ray diagnostic procedures is due to CT, interventional radiography, and angiography procedures.[2] It is observed that about 3.6 billion diagnostic X-ray examinations undertaken annually in the world.[1] To control the dose received by patients while undergoing diagnostic procedures, it is recommended internationally to apply the principle of optimization along with regulatory enforcement. Furthermore, it is imperative that the dose received by the patients should be measured using a calibrated dosimeter. Thus, there is an increased requirement to provide traceable calibration for instruments used in diagnostic radiology (CT, interventional radiology, mammography, and dental X-rays).[3],[4],[5]

ICRP 73, ICRU 74, ICRP-103, and IAEA BSS-2014 have recommended the establishment and use of diagnostic reference levels (DRLs) in diagnostic radiology.[6],[7],[8],[9] DRLs are defined in terms of air kerma. Free-air ionization chamber (FAIC) is the primary standard to realize air kerma in the energy range under consideration. The FAIC is a wall less ionization chamber where the primary photon beam and the secondary electron interact only with air to realize the definition of exposure and air kerma. There are basically two types of FAICs used in standards laboratories worldwide – the plane parallel type and the cylindrical type. Majority of the laboratories in the world make use of the plane parallel type although there is some advantage of the cylindrical FAIC over the plane parallel type FAICs. The laboratory maintains a primary standard plane parallel type medium energy FAIC (MFAIC) for which international equivalence has been established by participating in international intercomparison programs. The experience gained in developing the MFAIC was made use in developing a low-energy FAIC having a plane parallel geometry.

The laboratory's FAIC (MFAIC) is designed for medium energy X-ray beams. It can operate over the energy range from 50 kV to 300 kV. Although in principle the use of MFAIC for low-energy measurements is possible, it is not recommended. The magnitude of the correction for air attenuation and scatter and its associated uncertainties are more when the attenuation length is large for low-energy measurements.[10] Many of the standard laboratories maintain two chambers – one for lower energy (up to 50 kV) and the other for medium energy X-ray measurements (50–300 kV). To carry out all the diagnostic range X-ray measurements using a single chamber, a FAIC was designed to be used in the diagnostic energy range of 20 kV to 150 kV. FAIC of similar dimensions meant for low energy measurements have been established by international standards laboratories.[11],[12],[13],[14] The present study explains the design criteria of the chamber, characterization of the chamber, standardization of the diagnostic beams up to 150 kV, comparison with MFAIC of the laboratory, calibration of ionization chambers and X-ray meters, and evaluation of the associated uncertainties.


   Material and Methods Top


Diagnostic range beam qualities established in the laboratory

The X-ray machine maintained in the laboratory is a YXLON MG325 make with tungsten target. [Table 1] gives details of the diagnostic beam qualities and their possible applications. Diagnostic beam qualities-radiation qualities in radiation (RQR), radiation qualities based on aluminum added filter (RQA), radiation qualities based on copper added filter (RQT), and RQR beams for mammography (RQRM) were generated in the laboratory by determining the additional filtration required to generate the beams as per IEC 61267 and IAEA TRS-457 guidelines.[15]
Table 1: Diagnostic radiation beam qualities for the calibration of diagnostic dosimeters

Click here to view


Diagnostic range free-air ionization chamber

Parallel plate FAIC is a wall less ionization chamber which realizes the unit of the quantity exposure/air kerma from its definition. A schematic diagram of the diagnostic range FAIC (DFAIC) is shown in [Figure 1] (a) along with the images of the chamber in open (b) and shielded condition (c). The chamber is enclosed in a lead-lined box having dimensions: 321 mm×210 mm×212 mm. The chamber enclosure is made up of lead sheet of thickness 5 mm sandwiched between aluminum plates of thickness 15 mm and 3 mm. The lead-lined enclosure is provided so as to ensure that no scattered radiation enters the chamber volume and also to attenuate the primary beam falling on the front wall of the chamber.
Figure 1: (a) Schematic diagram of the diagnostic range free air chamber developed (b) Chamber in open condition (c) Chamber in shielded condition. Electrons e1 represent those electrons which deposit all its energy in the collecting region before getting collected at the electrodes, electrons e2 and e3 represent electrons that compensate the loss of charge produced in the collecting volume by charge particle equilibrium

Click here to view


The basic component of the DFAIC is a tungsten diaphragm aperture and aluminum plates. The X-ray beam passes through the diaphragm and interacts with the air in the active volume of the chamber. The chamber consists of three coplanar plates parallel to a high voltage plate. The central plate is the collecting electrode, and the other two coplanar plates are the guard electrodes. DFAIC plates were fabricated using 5-mm thick aluminum plates of grade AL6061T6. Coordinate measuring machine was used to measure the dimensions of the collector plate, the guard plate, and the HV plate. It was also used to measure the coplanarity of the collector with the guard plates. The collector length is 50.0 mm (tolerance 10 μm), and the coplanarity and flatness of the plates were measured with a tolerance of 10 μm. The collecting electrode is connected to an electrometer to measure the current produced due to ionization in the volume of interest.

The energy range in which the DFAIC can be used is based on the concept of charge particle equilibrium (CPE) and the requirement that the electrode separation should be such that the secondary electrons generated by primary interaction of photons in air do not reach the electrodes. The chamber is designed such that the distance of the plates from the X-ray beam is greater than the range of the most energetic secondary electrons originating in the beam.

Plate separation distance was fixed at 10 cm based on the range of continuous slowing down approximation (CSDA) range of the electrons in air and Klein-Nishina energy transfer equation for Compton interaction. The CSDA value for 50 keV photon is 4.08 cm in air, and as the plate separation is 10 cm, energies lower than 50 keV will expend all its energy before getting collected. As the energy range under consideration is predominantly in the photoelectric region, correction for loss of electron before it expends all its energy is more for energies >60 keV. The distance between the diaphragm and the center of the collecting region is 13.88 cm which satisfies the CPE condition, and the correction for the attenuation in air of the low-energy photons between the diaphragm and the collector is not substantial.

Electric field is generated by applying polarizing voltage to the high voltage plate and the collecting electrode maintained at ground/virtual ground potential. The guard electrodes are also maintained at ground potential. Appropriate voltage is applied between the plates so as to collect all the ion pairs generated by the interaction of photons in the air volume of interest. The electric field perturbation in the FAIC is reduced by encircling the FAIC with horizontal guard strips with reducing potential across each strip.

The volume of interest is defined by the collecting electrode length, the electric field lines, and the radius of the tungsten diaphragm.

Air kerma is calculated by the given equation (1)



Where I is the current measured, W/e is the mean energy expended by an electron of charge e to produce an ion pair in air, mair is the mass of air of the DFAIC volume defined by the aperture and the collecting plate, gair is the fraction of initial electron energy lost through radiative processes in air, and Πki corrects for the ion recombination loss, air attenuation, air density change, polarity effect, photon scatter contribution, electron loss, and photon transmission.

While designing the tungsten alloy diaphragm, it was ensured that the diameter of the aperture is greater in diameter than the focus of the X-ray tube (5.5 mm for YXLON MG325 tube). The diaphragm's aperture was measured to have a diameter of 9.90 mm with a tolerance of 10 μm. Basic chamber properties are given in [Table 2].
Table 2: Geometrical characteristics of the diagnostic range free-air ionization chamber

Click here to view


Characterization of diagnostic range free-air ionization chamber and evaluation of correction factor (experimentally)

Characterization of DFAIC and determination of correction factors, namely, ion recombination loss, air attenuation, air density change, polarity effect, photon scatter contribution, electron loss, and photon transmission were determined experimentally and explained in the following section.

Recombination correction factor ks

For air kerma measurements, although the voltage applied to the ionization chamber is in the saturation region, ion recombination still exists for which correction has to be applied. In an ionization chamber, recombination of ion pairs is due to two processes – the initial recombination between oppositely charged ions in a single ionizing particle track and the volume recombination due to recombination of ions formed in different ionizing tracks which is a dose rate dependent quantity. To determine the ion pairs lost due to recombination, measurements were carried out with a low noise triaxial cable connected to the FAIC and the electrometer. The high voltage bias was applied through a separate voltage source. Saturation studies carried out show 99% saturation on the application of bias voltage of 400 volts and above.

The loss of ion pairs in recombination even after applying adequate bias voltage has to be corrected. Recombination correction factor was evaluated by the method described by Boutillon.[16],[17] Ionization current IV and IV/n of the FAIC were measured at two voltages, V and V/n where n is between 2 and 4. The ionization currents were corrected for change in temperature and pressure from the reference value. The measurements were repeated for different values of the air kerma rates by different filtrations. The ratio of the current R = IV/IV/n was determined for the dose rate values standardized in the laboratory. The different values of R were plotted as a function of uncorrected ionization current IV, uncorrected (current which is not corrected for change in temperature and pressure, attenuation, etc.). The intercept of the linear fit of the data is denoted as a0 and the gradient a1.

The initial recombination coefficient kinit and the volume recombination coefficient kvol are evaluated using the equations:





The recombination correction factor ks for the ionization current Iv, day at polarizing voltage V on a given day is given by:



Where IV, day, is the ionization current for a particular day when a polarizing voltage V is applied for which ks has to be determined. Linear fit curve for the determination of the ion recombination correction factor was plotted for the dose rates generated.

Linearity of response with variation of tube current of the X-ray tube

The response linearity test was done to verify that the DFAIC response is linear with dose rate variation. The tube current was varied between 1 mA and 30 mA. The measurements were carried out with constant voltage and additional filtration for beam quality RQR 5. The measured ionization charge was normalized to the charge corresponding to 30 mA tube current.

Polarizing correction factor kpol

Polarity effect was studied by measuring charge with both positive and negative polarity and the polarity correction factor kpol calculated as:



Where M + and M are the charge obtained at positive and negative polarity, respectively, and M is the charge for the polarity routinely used. The polarity effect once determined is applied for subsequent measurements for each beam quality. The consistency of the value is checked at regular intervals to ensure that the value remains the same.

Air attenuation correction ka

This correction is applied to correct the air attenuation between the reference plane and the center of the collection volume. The attenuation is determined by applying the mass attenuation coefficient in air for the effective energy of the beam standardized. The air attenuation correction factor ka is given by:



Where

μ/ρ is the mass attenuation coefficient (cm2g-1)

L is the distance between the defining plane and the chamber center distance of 13.88 cm

The attenuation coefficient in air was taken from XCOM, published by NIST.

Air density correction factor kT,P

This correction factor corrects for the change in mass of air in the collecting volume due to change in ambient temperature and pressure. The ionization current was corrected to get current at reference temperature of 20°C and reference pressure of 1013.2 mbar. The air density correction factor is given by equation 7.



Where tavg and pavg are the average temperature and pressure values over the course of measurements.

Photon transmission correction factor k1

The diaphragm used in the DFAIC is made up of tungsten alloy and has a simple cylindrical aperture. An analytical diaphragm transmission correction factor k1[10],[18] is used to correct for the transmission of photons through the downstream edge of the diaphragm and is given by:



Where,

μ is the linear attenuation coefficient of the tungsten alloy for the effective energy of the beam and Z is the distance from the focal spot to the reference plane of the DFAIC. Diaphragm transmission correction factor for all the diagnostic beam qualities was determined and applied to the DFAIC measured current.

Transmission through front wall of the chamber

The beam interacts with the front wall having 5 mm thick lead sheet sandwiched between 15 mm and 3 mm of aluminum sheets. Measurements were carried out with the ISO narrow beam quality N100 to assure that the wall transmission is negligible for low-energy diagnostic range photon beams. Measurements were carried out with and without a suitable lead stopper in front of the aperture and wall transmission determined.

Electron loss correction factor ke

Inadequate plate separation “d” and smaller collecting electrode width “w” results in the secondary electrons reaching the collector plate before expending all its energy which results in the definition of exposure not being realized. A correction factor to account for any loss of charge resulting from d and w is accounted in the electron-loss correction factor ke.

The CSDA range for 55 keV electron is 4.81 cm in air. As the plate separation is 10 cm and the collecting electrode width is 12 cm, the electron loss correction factor for low-energy X-ray tube up to 50 kV is taken as 1. The percentage of electron loss for energies >50 keV outside a given radius from the X-ray beam was computed for a zero diameter beam in cylindrical coordinates and using the figures given by RITZ and NIST publications for additional filtration of 3 mm Al.[19],[20],[21] The electron loss corrections were determined up to 100 kV as a function of the half-value layer of aluminum and given by equation 9.



Photon scatter correction factor ksc

The collection of any charge due to the interaction of secondary photons in air is not to be included in the definition of air kerma. The correction factor for this contribution was calculated based on an empirical relationship given by Burns,[10],[22] and the values of ksc for some of the low and medium energy beam qualities were determined.

Air kerma measurements and comparison with MFAIC

Air kerma measurements were carried out with the DFAIC chamber for all the diagnostic beam qualities established in the laboratory. RQR beam qualities were compared with MFAIC for tube potential varying from 50 kV to 150 kV.


   Results and Discussion Top


The DFAIC designed and developed in the laboratory satisfies the dual criteria of a FAIC – the attenuation length should be greater than the electron range for the maximum photon energy under consideration and the plate separation should be such that the secondary electrons expend all its energy before reaching either of the electrodes.

The first step in characterization of DFAIC was to determine the voltage at which all the ion pairs produced in the chamber are collected. [Figure 2] shows measured ionization current as a function of applied voltage. Saturation of more than 99% is achieved when bias voltage of 400 volts and above is applied to the chamber.
Figure 2: Ionization saturation curve of the DFAIC

Click here to view


Collection of all the ion pairs produced in the chamber volume is not possible as a finite amount of ion pairs is always lost to recombination. Recombination loss correction factor ks was determined for the dose rates generated in the laboratory by using the slope and intercept of the linear fit data plotted in [Figure 3] and equations 2 - 4.
Figure 3: Ion recombination linear fit curve

Click here to view


Application of 2000 volts and above results in an ion collection efficiency better than 99.96%. The linearity of response of the chamber with X-ray tube current is shown in [Figure 4], and the linear correlation coefficient 0.99998 demonstrates the performance of the DFAIC with increasing dose rate.
Figure 4: Linearity of response with the variation in tube current of X-ray tube

Click here to view


Polarity and air density correction factor is determined and the DFAIC current corrected for it. Air attenuation and diaphragm transmission correction factors were determined for the effective energies of the diagnostic beams. The transmission of photons through the front wall of thickness 5 mm lead for the DFAIC is 0.09%. It is generally recommended that the transmitted photons should be less than 0.1% of the initial beam.[23] Electron loss correction factor varied from 0.02% at 60 kV to 0.5% for 100 kV X-ray potential. For X-ray beams with tube potential higher than 100 kV and also additional filtration >3 mm, electron loss correction factor of 0.5% was applied as the calculations are based on RITZ and NIST publications for tube potential up to 100 kV and additional filtration of 3 mm aluminum. Photon scatter correction was <0.5% for X-ray potential up to 150 kV. [Table 3] gives the correction factors applied to the DFAIC measured current.
Table 3: Correction factors for the diagnostic range free-air ionization chamber for diagnostic beam qualities

Click here to view


Internal comparison of DFAIC with the MFAIC was conducted as a preliminary verification of the metrology characteristics of the DFAIC. [Table 4] shows the air kerma rate values of RQR beam qualities measured using both the FAICs. The deviation between the two chambers is within the uncertainty of measurement of each chamber.
Table 4: Air kerma rate measurements for radiation qualities in radiation beam qualities using diagnostic range free-air ionization chamber

Click here to view


RQA, RQT, and RQRM beam qualities generated were standardized using DFAIC, and [Table 5] gives the air kerma rates along with the details of the additional filtration used to generate the beam qualities and the HVL of the beams.
Table 5: Air kerma rate measurements of the radiation qualities based on aluminum added filter, radiation qualities based on copper added filter, and radiation qualities in radiation beams for mammography beam qualities using diagnostic range free-air ionization chamber

Click here to view


Uncertainty of measurements in the standardization of the diagnostic beam qualities for DFAIC is given in [Table 6]. The uncertainty for MFAIC measurements is 0.51% at 1 sigma level. Uncertainty in the measurement was evaluated using the document “Evaluation of measurement data- Guide to the expression of uncertainty in measurement (GUM)(JCGM 100:2008)”[24] for DFAIC and the primary standard MFAIC.
Table 6: Uncertainties associated with the diagnostic range free-air ionization chamber for the air kerma measurements of diagnostic beam qualities

Click here to view


Dissemination of standards to the users in the field of diagnostic radiology

The aim of establishing and standardizing diagnostic beam qualities is to provide traceable calibration to the users in the field of diagnostic radiology. Calibration of solid-state detectors/instruments and ionization chambers from different diagnostic radiology facilities have been carried out, and the values compared with the certificate values of the manufacturer which are traceable to primary standard FAICs of international standards laboratories. Calibration was carried out by substitution technique where the detector's reference point is placed at the point where the air kerma rate is measured using the DFAIC. [Table 7] gives details of the beam qualities compared, detectors used, and the deviation between the reference values established in the laboratory with the user instrument readout value. The uncertainty in calibrating/testing of the instrument at 95% confidence level (K = 2) is ±3%. [Table 7] gives details of the uncertainty of each instrument quoted by the manufacturer traceable to an international standard. For air kerma measurements, compliance with the manufacturer quoted limit of the values is based on ILAC-G8:03/2009.[25] Of the 11 beam qualities tested, two measurements plus its uncertainty were not complying with the limits quoted by the manufacturer even though its measurements were within the limit.
Table 7: Comparison of the air kerma rate measured with commercially available detector system

Click here to view



   Conclusion Top


Diagnostic range free-air ion chamber is designed and developed in the laboratory for 20-150 kV beam standardization. The establishment of diagnostic range air kerma standard and its dissemination to the users in the field of diagnostic radiology is achieved in this work. An air kerma-based calibration facility for the calibration of diagnostic range instruments is established in the country. Characterization of DFAIC and comparison of the chamber with the primary standard MFAIC maintained in the laboratory are done with satisfactory results. Experimental measurements were carried out to determine the various correction factors for DFAIC. Uncertainty associated in air kerma measurements have been evaluated for the DFAIC. The chamber will be used as an absolute standard for air kerma measurements in diagnostic range. The evaluation of correction factors by Monte Carlo technique and participating in international intercomparison will aid in establishing the chamber as a primary standard in the X-ray diagnostic energy range.

Acknowledgment

Authors would like to thank Shri R M Suresh Babu, Director, Health Safety and Environment Group for his support and encouragement. Authors are also thankful to Ashish D Raut, Sachin P Bobade, and Dipali Mhadeshwar for their help in arranging the setups and carrying out the experiments.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR), Sources and Effects of Ionizing Radiations (Report to the General Assembly with Scientific Annexes A and B), New York: United Nations Scientific Committee on the Effects of Atomic Radiation; 2008.  Back to cited text no. 1
    
2.
United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR). Sources and Effects of Ionizing Radiations (Report to the General Assembly with Scientific Annexes). New York: United Nations Scientific Committee on the Effects of Atomic Radiation, UN; 2000.  Back to cited text no. 2
    
3.
Wagner LK, Fontenla DP, Kimme-Smith C, Rothenberg LN, Shepard J, Boone JM. Recommendations on performance characteristics of diagnostic exposure meters: report of AAPM Diagnostic X-Ray Imaging Task Group No. 6, Med Phys 1992;19:231-41.   Back to cited text no. 3
    
4.
International Electro Technical Commission (IEC) 61267. Medical Diagnostic Equipment – Radiation Conditions for Use in the Determination of Characteristics: Geneva, IEC; 2005.  Back to cited text no. 4
    
5.
International Atomic Energy Agency (IAEA), Technical Reports Series No. 457; Dosimetry in Diagnostic Radiology: An International Code of Practice, Vienna, IAEA; 2007.  Back to cited text no. 5
    
6.
International Commission on Radiological Protection (ICRP) Publication 73. Radiological Protection and Safety in Medicine. Oxfordshire, England: Pergamon; 1996.  Back to cited text no. 6
    
7.
International Commission on Radiation Units and Measurements (ICRU). Report 74, Patient Dosimetry for X-rays Used in Medical Imaging: Bethesda, MD: ICRU; 2005.  Back to cited text no. 7
    
8.
International Commission on Radiological Protection. Annals of the ICRP; Publication 103, Volume 37 Nos. 2-4; ICRP, 2007.  Back to cited text no. 8
    
9.
International Atomic Energy Agency (IAEA). Safety Standards Series No. GSR Part 3, Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards; Vienna, IAEA; 2014.  Back to cited text no. 9
    
10.
Burns DT, Büermann L. Free-air ionization chambers. Metrologia 2009;46:S9-23.  Back to cited text no. 10
    
11.
Seltzer SM. Correction Factors for the NIST Free-Air Ionization Chambers Used to Realize Air Kerma from W-Anode X-Ray Beams, National Institute of Standards and Technology, NISTIR 7887, National Institute of Standards and Technology of U.S; 2012. p. 1-41.  Back to cited text no. 11
    
12.
Šolc J, Sochor V. Characterization of the new free-air primary standard for low energy X-rays at CMI. Radiat Phys Chem 2014;104:328-31.  Back to cited text no. 12
    
13.
Seyed Mostafa Mohammadi, Hossein Tavakoli-Anbarana. Design of the free-air ionization chamber, FAC-IR-150, for X-ray dosimetry: Radiation Physics and Chemistry 2018;144:344-50.  Back to cited text no. 13
    
14.
Kessler C, Roger P, Burns DT. Establishment of Reference Radiation Qualities for Mammography: BIPM Rapport-2010/01; Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312 Sevres Cedex , 2010.   Back to cited text no. 14
    
15.
Panyam VS, Rakshit S, Dhole SD, Patil BJ, Greeshma KA, Kulkarni MS. Methodology adopted to establish diagnostic X-ray beam qualities. Appl Radiat Isotopes 2019;150:164-7.  Back to cited text no. 15
    
16.
Boutillon M. Volume recombination parameter in ionization chambers. Phys Med Biol 1998;43:2061-72.  Back to cited text no. 16
    
17.
Burns DT, Kessler C, Roger P, El-Sersy AR. Preliminary Characterization of the NIS Free-Air Chamber Standard at the BIPM: Rapport BIPM-09/02; Bureau International des Poids et Mesures, F-92312 Sevres, July, 2009.  Back to cited text no. 17
    
18.
Simons HA. The calculation of gamma ray penetration of the walls of cylindrical and conical collimating holes. Phys Med Biol 1962;6:56-76.  Back to cited text no. 18
    
19.
Ritz VH. Design of Free-Air Ionization Chambers for the Soft X-Ray Region (20–100 kV): Radiology, 73;911-22.  Back to cited text no. 19
    
20.
Paul J. Lamperti, Michelle O'Brien. Calibration of X-Ray and Gamma-Ray Measuring Instruments; NIST Special Publication; NIST, 2001. p. 250-8.  Back to cited text no. 20
    
21.
Paul J. Lamperti, Thomas P. Loftus, Robert Loevinger. Calibration of X-Ray and Gamma-Ray Measuring Instruments; NBS Special Publication; NBS, 1988. p. 250-16  Back to cited text no. 21
    
22.
Burns DT. Free-Air Chamber Correction Factors for Electron Loss, Photon Scatter, Fluorescence and Bremsstrahlung: Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312 Sevres Cedex, CCRI (I)/01-36; 2001.  Back to cited text no. 22
    
23.
Wyckoff HO, Kirn FS. Standard ionization-chamber requirements for 250- to 500-kilovolt X-rays. J Res Natl Bur Stand 1957;58:111-7.  Back to cited text no. 23
    
24.
Joint Committee for Guides in Metrology. Evaluation of measurement data - Guide to the expression of uncertainty in measurement: JCGM 100:2008 GUM 1995 with minor corrections; JCGM.  Back to cited text no. 24
    
25.
Accreditation Committee International Laboratory Accreditation Cooperation (ILAC). International Guidelines on the Reporting of Compliance with Specification: ILAC-G8:03, 2009; ILAC, Australia.  Back to cited text no. 25
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (873 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
   Material and Methods
    Results and Disc...
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed107    
    Printed0    
    Emailed0    
    PDF Downloaded31    
    Comments [Add]    

Recommend this journal