Probability distribution of pixel intensities of ebt3 films and its application in the correction of uncertainty budget


1 Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai, Maharashtra, India
2 Radiological Physics and Advisory Division, Bhabha Atomic Research Centre; Homi Bhabha National Institute, Mumbai, Maharashtra, India

Date of Submission05-Oct-2020
Date of Decision22-Mar-2021
Date of Acceptance22-Mar-2021
Date of Web Publication5-May-2021

Correspondence Address:
Mr. Rahul Kumar Chaudhary
Radiological Physics and Advisory Division, CT and CRS Building, Bhabha Atomic Research Centre, Anushaktinagar, Mumbai - 400 094, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_94_20

Rights and Permissions

 

   Abstract 

Background and Aim: Modern radiotherapy modalities, such as Intensity-Modulated Radiotherapy and Volumetric Modulated Arc Therapy involve complex dose delivery. The dose delivery is complex as it involves beam modulation, hence, manual dose calculations for these techniques are not possible. Film dosimetry is commonly used method of dose verification for these modalities because of the advantages associated with it. The quantification of uncertainty associated with a film dosimetry system under clinical use becomes important for accurate dosimetry. The spread in the distribution of the pixel values (PV) of the irradiated film contributes to the uncertainty. The probability distribution (PD) of the PV was studied for the clinical photon beam energies of 6, 10, and 15 MV. Methods and Materials: Gafchromic EBT3 film and EPSON 10000XL flatbed scanner were used for this purpose and using the resulting PD, the uncertainty budgets for these energies in the red, green and blue color channels were estimated. Results: The PV of exposed films for the energies studied follows t-distribution, the sum of the squares of the deviation of the measured data from the fitted value was of the order of 10−7, this indicates the goodness of fit. The “t” value corrected combined standard uncertainty (CSU) at 1σ confidence level for exposed film and dose measurement at 200 cGy were 1.42%, 1.48%, and 1.63% and 1.99%, 3.23%, and 5.08% for 6, 10, and 15 MV energies, respectively, in the red colour channel. Conclusion: In the case of the limited number of measurements of a quantity, the SU values must be corrected using the “t” value to get the correct CSU.


Keywords: EBT3 films, film dosimetry, flatbed scanners, uncertainty budget


How to cite this article:
Chaudhary RK, Pathan M, Kumar R, Sharma S D, Sapra B K. Probability distribution of pixel intensities of ebt3 films and its application in the correction of uncertainty budget. J Med Phys 2021;46:26-32

How to cite this URL:
Chaudhary RK, Pathan M, Kumar R, Sharma S D, Sapra B K. Probability distribution of pixel intensities of ebt3 films and its application in the correction of uncertainty budget. J Med Phys [serial online] 2021 [cited 2021 Jun 21];46:26-32. Available from: https://www.jmp.org.in/text.asp?2021/46/1/26/315512



   Introduction Top


Modern radiotherapy techniques, such as intensity-modulated radiotherapy and volumetric modulated arc therapy involve complex dose delivery. Manual calculations of the dose are not possible for these modalities. Hence, the pre-treatment dose verification of these techniques is often done by comparing the planned dose distribution, calculated by the treatment planning system with the measured dose distributions and performing the gamma analysis. The measurement of the dose distribution is done by two-dimensional arrays or radiochromic films.

Radiochromic films such as Gafchromic EBT3 (Ashland International Speciality Products Advanced Materials, Wayne, NJ, USA) are frequently used in radiotherapy dosimetry because of the known advantages associated with it, such as, nearly tissue equivalent composition,[1] weak energy dependence,[1],[2],[3],[4] dose rate independence of response[5] and high spatial resolution.[6] The radiochromic films undergo polymerization and develop self-darkening when irradiated by the ionizing radiations. The darkening developed in the film as a result of the irradiation is quantified in the terms of net optical density (NOD) which is the measure of absorbed dose.

In order to digitize the films, film scanners are used. Film scanner consists of the light source and a charge-coupled device to receive the light transmitted through the irradiated film. The radiochromic film along with the scanner constitutes the film dosimetry system (FDS).

Over the years, different models and types of film scanners have been used for film dosimetry, for example, roller-based (VIDAR VXR Dosimetry Pro Advantage, VIDAR Systems Corporation, VA, USA) and commercially available flatbed scanners (FBS) such as EPSON 10000XL (Seiko Epson Corporation, Nagano, Japan), etc. These scanners have been studied in detail by various groups,[7],[8],[9] FBS are widely used with Gafchromic EBT film types because of the better results shown, as compared to the roller-based scanners.[10]

Film dosimetry requires precautions during the handling, irradiation and scanning to minimize the sources of uncertainty and to improve the accuracy. Therefore, uncertainty quantification of the FDS for clinical energies becomes necessary for the success of treatment.

It has been observed that, during the analysis of scanned film image, there is a spread in the pixel values (PV) of the irradiated film, this introduces uncertainty in the measurement of the absorbed dose. The spread of the PV is described by the probability distribution (PD) curve and its nature is important in the determination of uncertainty. In the present study, the nature of PD of the PV has been studied for the dose of 200 cGy, delivered by 6, 10, and 15 MV photon beams.

The resulting PD (t-distribution [TD]) was applied to estimate the uncertainty budgets of 6, 10, and 15 MV photon beams, which were otherwise, estimated assuming the Gaussian distribution of measurement (or sample) data.[6],[7],[8] In the absence of sufficient data, this assumption may decrease the confidence in the value of reported combined standard uncertainty (CSU). TD has been used to correct the SU values, associated with individual components of radiochromic film dosimetry in the cases, involving the limited number of measurements of a quantity.


   Materials and Methods Top


Film handling and sample preparation

EBT3 films have an expiry period of about 30 months from the date of manufacturing. EBT3 films within the expiry date were used for dosimetry purposes. The films were handled either using gloves or vacuum tweezers, films were kept in the dark when not in use and were handled in the interior room light as per the recommendations of the EBT3 film specifications manual.[11]

Square film pieces of 5 cm × 5 cm were used for the study, film samples were numbered at the top right corners to maintain the identity of orientation at the time of placing the film samples on the glass plate of FBS.

Flatbed scanner settings and film scan

Flatbed scanner settings

The EPSON scan software (version 3.49 A) was used for the scan of the film on the FBS. The films were scanned in 48 bit red, green and blue (RGB) colour mode at 72 dots per inch resolution.[12] The image enhancement features were turned off and scanned images were saved in Tiff format.

Film scan

The EPSON 10000XL scanner's lamp turns on at the time of preview or scan. In order to stabilize the response of scanner, the practice of ignoring first seven scans was followed before starting the actual measurements.[8] Since portrait orientation of the film has higher dose-response and lesser uncertainty in determining the dose than landscape orientation,[13] the films were scanned in the portrait orientation at the center of glass plate of FBS to minimize the effect of non-uniformity of response of the scanner.[14] This also ensures the reproducibility of film positioning.

The compression glass plate is recommended above the film to ensure flat-film scanning.[15] This removes the uncertainty arising due to the non-contact of the film with the glass plate of the scanner. [Figure 1] shows an example of the placement of the glass plate to make the film in perfect contact with the scanner's glass plate. The films were read after the waiting period of 48 h post irradiation.[16]
Figure 1: Photograph of the placement of glass sheet over the film to make it in perfect contact with the glass plate

Click here to view


Film irradiation and image analysis

Irradiation geometry and uncertainty in the dose delivery

The irradiations were carried out on the True Beam® linear accelerator (Varian Medical Systems, Palo Alto, USA). Prior to the irradiation of the films on the linear accelerator, the uncertainties in the outputs of 6, 10 and 15 MV beams were estimated following the procedure mentioned in IAEA-TECDOC-1585.[17] The film irradiation and ionization chamber measurements were done using the slab phantom in the isocentric geometry at 10 cm depth by 10 cm × 10 cm field size. Twelve calibrations doses extending through the optimum range of EBT3 films, i.e., 0.2–10 Gy were delivered for the calibration of EBT3 films.[18]

Image analysis

Image J software (Version 1.51a, National Institute of Health, Bethesda, Maryland, USA) was used for the image analysis. The region of interest (ROI) of 4 cm × 4 cm was chosen for reading the PV in the RGB colour channels.[18]

Probability distribution of the pixel values

The nature of spread in the PV of films, irradiated by 6, 10 and 15 MV beams to the dose of 200 cGy was determined by plotting the PD function of PV in the RGB channels for above-mentioned energies. The probabilities of PV were obtained by taking the ratio of the counts of particular PV to the total pixels counts in the ROI. For better readability of the data, corresponding normalized histograms and 1σ confidence levels have also been plotted. The data was fitted with TD and the goodness of fit was estimated by the sum of squares of the deviation of measured data from the fitted value.

The “t” value of TD was used for determining confidence intervals (CI) and hence, SU of the components of the exposed films, this was used further in determining the CSU (total uncertainty) of the exposed films and the experiment.

The CI of the mean PV of the data following TD is given by



here, μ = Mean of the samples

σ (Sigma) = Standard deviation of the mean of samplest = TD critical value for the given degrees of freedom (DF) and probability, available from standard tables

n = sample size

In this study, the SU of the components of the exposed film has been obtained from error (±) term of eq. (1), expressed as the percentage of the mean PV.

Film calibration and curve fitting

Calibration curves were plotted between the NOD and delivered doses for the three colour channels. The expression used for the calculation of NOD[19] is as follows:



here, Iexp, Iunexp and Ibckg are the PV for exposed, unexposed films and dark sheet, respectively.

Non-linear relationship was used to obtain dose from the measured NOD,[20]

D = a.NOD + b.NODc (3)

Here, a and b are fitting parameters and D is the delivered dose. Parameter c has the value that gives the best fitting results and generally lies between 2 and 3. For our experimental data, the value of 2.5 gave the best fitting results. Levenberg-Marquardt optimization algorithm was used for the fitting of experimental data, this is because this algorithm is adapted for the fitting of multiple parameters and non-linear functions.[21]

Uncertainty estimation

The uncertainty in the film dosimetry is divided into two parts (1) experimental uncertainty and (2) uncertainty due to curve fitting. The overall uncertainty is the sum in quadrature of relative experimental uncertainty and relative fitting uncertainty.[21]

Relative experimental uncertainty

The relative experimental uncertainty is given by the expression[19]



Here, σNOD is the SU in the NOD value which is calculated by using the error propagation expression,[22]



here, , and are the SU of measured Iunexp, Iexp and Ibckg respectively, estimated using eq. (1). Dfit is the dose estimated as the result of fitting of the calibration data. The Ibckg was measured by placing a completely dark sheet over the scanner's glass plate and taking the average of the scan values over the whole glass surface. These values remained constant throughout the experiment. [Figure 2] shows the arrangement to scan the black sheet for the measurement of Ibckg. Similarly, Iunexp was obtained by computing the mean of the response of unexposed film samples of size 5 cm × 5 cm. These values also remained constant throughout the experiment.[21]
Figure 2: Arrangement to scan the black sheet to measure the Ibckg

Click here to view


To find the CSU of the exposed films, the SU due to the individual components (intra sheet, inter sheet, intra scanner uniformity, scanner reproducibility, scanner warm up effect, output of linac, film orientation, effect of reading delay, film positioning reproducibility, gap between film and glass plate of the scanner) were assumed independent of each other and the CSU of the exposed film was obtained by adding the SU of components in the quadrature. [Table 1] describes the method to estimate the uncertainty due to the individual components (except, the output of linac) as mentioned above.[21]
Table 1: Components of the uncertainty of an exposed film and the methods to estimate them

Click here to view


Fitting uncertainty

The relative fitting uncertainty is given by,



Here, σa and σb are the uncertainties associated with fitting parameters a and b, respectively.[19]


   Results Top


Film calibration and curve fitting

[Figure 3],[Figure 4],[Figure 5] show the calibration curves for 6, 10, and 15 MV energies, respectively in the RGB color channels. [Table 2] shows the values and uncertainties of the fitting parameters of the fitted non-linear equations in the RGB channels for the photon energies under the study.
Figure 3: Calibration curves for 6 MV energy in the red, green and blue colour channels

Click here to view
Figure 4: Calibration curves for 10 MV energy in the red, green and blue colour channels

Click here to view
Figure 5: Calibration curves for 15 MV energy in the red, green and blue colour channels

Click here to view
Table 2: Values of the fitting parameters and associated uncertainties for 6, 10 and 15 MV energies in red, green and blue colour channels

Click here to view


Uncertainty in the output of the linear accelerator

The uncertainty in the output of linear accelerator was quantified by the ionization chamber measurements. The uncertainty in the output was <1% for all the photon beams under the study, which is in agreement with the other studies.[23] This value was considered for the uncertainty budget estimation.

Probability distribution of the pixel values

[Figure 6],[Figure 7],[Figure 8] show the PD curves of the PV along with the normalized histograms of the films in the RGB colour channels, irradiated to the dose of 200 cGy by 6, 10 and 15 MV energies, respectively. It can be observed that, the PD curves are symmetrical about the mean PV for all the color channels for the studied energies. The sum of squares of the difference between measured and TD fitted data was of the order of 10−7, this shows the goodness of fit of the TD. The 1σ CI of mean is shown by shaded area on the PD curves.
Figure 6: Probability distribution curves, normalized histograms and 1σ CI of red (top), green (middle) and blue (bottom) colour channels for EBT3 films irradiated by 6 MV energy

Click here to view
Figure 7: Probability distribution curves, normalized histograms and 1σ CI of red (top), green (middle) and blue (bottom) colour channels for EBT3 films irradiated by 10 MV energy

Click here to view
Figure 8: Probability distribution curves, normalized histograms and 1σ CI of red (top), green (middle) and blue (bottom) colour channels for EBT3 films irradiated by 15 MV energy

Click here to view


Uncertainty budget

The uncertainty budget of exposed film is shown in [Table 3]. The relative experimental uncertainty and fitting uncertainties at 1σ confidence level for 6, 10 and 15 MV energies in RGB color channels are shown in [Table 4].
Table 3: Uncertainty budget of the exposed film for 6, 10 and 15 MV energies in red, green and blue colour channels at 1σ confidence level

Click here to view
Table 4: Relative experimental and fitting uncertainties for 6, 10 and 15 MV energies in red, green and blue colour channels at 1σ confidence level

Click here to view


[Table 3] shows that the CSU of the exposed films are 1.42%, 1.48% and 1.63% for 6, 10 and 15 MV energies, respectively in the red color channel. Red color channel has the highest CSU for the exposed film, whereas, rest of the color channels have almost similar uncertainties for all the energies studied.

It can be observed from [Table 4] that, total uncertainty in the dose determination is 1.99%, 3.23% and 5.08% for 6, 10 and 15 MV energies respectively, in the red color channel.


   Discussions Top


The PD of the PV in the EBT3 film dosimetry was studied, and it was found that the PV PD fits well TD. The TD approaches the Gaussian behavior for large sample size or degrees of freedom.

Earlier studies have estimated the uncertainty budget associated with radiochromic film dosimetry assuming the Gaussian distribution. This assumption may underestimate the confidence level for the limited number of measurement data. Application of the TD in the estimation of the SU corrects the uncertainty budget.

TD has been applied here because, the data of any film dosimetry experiment is assumed to be the sample from Gaussian distribution, whereas the sample size (generally 3 or 5 readings) itself is too small to be normally distributed, in these situations TD corrects for this assumption, this is achieved by including the “t” value in the expression for CI (eq. [1]).

The uncertainty budget for EBT3 film and EPSON 10000XL FBS FDS has been estimated on the basis of TD. Wherever possible, the sources of the uncertainty were reduced to ≤0.01% and were therefore excluded in the estimation of uncertainty budget. The remaining sources of the uncertainty were kept as minimum as possible following the procedures mentioned in this study. The uncertainty budget presented here is usable and is valid for the FDS of the current study. The method described here for correcting the uncertainty budget in the absence of the sufficient data is independent of the FDS.

The red color channel has the highest uncertainty for all the energies and color channels studied, this is because of the highest sensitivity of response with delivered dose for the red color channel. As a result of this, any error impacts the uncertainty in this channel the most. The increase in total uncertainty in dose determination with the energy of the beam is consistent with the findings of the past study.[21]

For the completeness of the study, the most commonly used 9 MeV electron energy was also studied (not described here) and it was found that the total uncertainty behavior was similar to 6 MV energy for the RGB color channels studied.


   Conclusions Top


The PD of the PV of the irradiated film fits TD well, this distribution approaches to normal behavior for large sample sizes. CI must be corrected by “t” value in the case of radiotherapy dosimetry, where, limited measurements of the quantity of interest are made and distribution of the measured values is assumed to be Gaussian.

 
   References Top

1.
Spelleken E, Crowe SB, Sutherland B, Challens C, Kairn T. Accuracy and efficiency of published film dosimetry techniques using a flatbed scanner and EBT3 film. Australas Phys Eng Sci Med 2018;41:117-28.  Back to cited text no. 1
    
2.
Chiu-Tsao ST, Ho Y, Shankar R, Wang L, Harrison LB. Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies. Med Phys 2005;32:3350-4.  Back to cited text no. 2
    
3.
Butson MJ, Cheung T, Yu PK. Weak energy dependence of EBT Gafchromic film dose response in the 50 kVp-10 MVp X-ray range. Appl Radiat Isot 2006;64:60-2.  Back to cited text no. 3
    
4.
Fuss M, Sturtewagen E, De Wagter C, Georg D. Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol 2007;52:4211-25.  Back to cited text no. 4
    
5.
Rink A, Vitkin IA, Jaffray DA. Characterization and real-time optical measurement of the ionizing radiation dose response for a new radiochromic medium. Med Phys 2005;32:2510-16.  Back to cited text no. 5
    
6.
Saur S, Frengen J. GafChromic EBT film dosimetry with flatbed CCD scanner: A novel background correction method and full dose uncertainty analysis. Med Phys 2008;35:3094-101.  Back to cited text no. 6
    
7.
Matney JE, Parker BC, Neck DW, Henkelmann G, Rosen II. Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry. J Appl Clin Med Phys 2010;11:198-208.  Back to cited text no. 7
    
8.
Ferreira BC, Lopes MC, Capela M. Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry. Phys Med Biol 2009;54:1073-85.  Back to cited text no. 8
    
9.
Paelinck L, Neve WD, Wagter CD. Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry. Phys Med Biol 2007;52:231-42.  Back to cited text no. 9
    
10.
Wilcox E, Daskalov G, Nedialkova L. Comparison of the Epson Expression 1680 flatbed and the vidar VXR-16 Dosimetry PRO™ film scanners for use in IMRT dosimetry using Gafchromic and radiographic film. Med Phys 2007;34:41-8.  Back to cited text no. 10
    
11.
GafChromic EBT3 Film Specifications. Available from: http://www.gafchromic.com. [Last accessed on 2020 Sep 15].  Back to cited text no. 11
    
12.
Epson Expression® 10000XL Color Flatbed Scanner User's Guide. Available from: https:files.support.epson.com. [Last accessed on 2021 Apr 04].  Back to cited text no. 12
    
13.
Marroquin EY, Gonzalez JA, Lopez MA, Barajas JE, Garcia-Garduno OA. Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density. J Appl Clin Med Phys 2016;17:466-81.  Back to cited text no. 13
    
14.
Poppinga D, Schoenfeld AA, Doerner KJ, Blanck O, Harder D, Poppe B. A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry. Med Phys 2014;41:021707.  Back to cited text no. 14
    
15.
Palmer AL, Bradley DA, Nisbet A. Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning. J Appl Clin Med Phys 2015;16:425-31.  Back to cited text no. 15
    
16.
Borca VC, Pasquino M, Russo G, Grosso P, Cante D, Sciacero P, et al. Dosimetric characterization and use of Gafchromic EBT3 film for IMRT dose verification. J Appl Clin Med Phys 2013;14:158-71.  Back to cited text no. 16
    
17.
IAEA-TECDOC-1585. Measurement Uncertainty. A Practical Guide for Secondary Standards Dosimetry Laboratories. ISBN 978-92-0-104408-2, ISSN 1011-4289, Printed by the IAEA in Austria in May 2008.  Back to cited text no. 17
    
18.
Bouchard H, Lacroix F, Beaudoin G, Carrier JF, Kawrakow I. On the characterization and uncertainty analysis of radiochromic film dosimetry. Med Phys 2009;36:1931-46.  Back to cited text no. 18
    
19.
Devic S, Seuntjens J, Hegyi G, Podgorsak EB, Soares CG, Kirov AS, et al. Dosimetric properties of improved GafChromic films for seven different digitizers. Med Phys 2004;31:2392-401.  Back to cited text no. 19
    
20.
Devic S, Seuntjens J, Sham E, Podgorsak EB, Ross Schmidtlein C, Kirov AS, et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys 2005;32:2245-53.  Back to cited text no. 20
    
21.
Sorriaux J, Kacperek A, Rossomme S, Lee JA, Bertrand D, Vynckier S, et al. Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams. Phys Med 2013;29:599-606.  Back to cited text no. 21
    
22.
Bevington PR, Robinson DK. Data Reduction and Error Analysis for the Physical Sciences. Boston: WCB/McGraw–Hill; 1992.  Back to cited text no. 22
    
23.
Chan MF, Li Q, Tang X, Li X, Li J, Tang G, et al. Visual analysis of the daily QA results of photon and electron beams of a trilogy linac over a five-year period. Int J Med Phys Clin Eng Radiat Oncol 2015;4:290-9.  Back to cited text no. 23
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,954 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussions
   Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed74    
    Printed0    
    Emailed0    
    PDF Downloaded38    
    Comments [Add]    

Recommend this journal