ORIGINAL ARTICLE |
|
|
|
Year : 2021 | Volume
: 46
| Issue : 1 | Page : 1-6 |
|
Use of a checklist approach on a telecobalt in an attempt to reduce human errors in radiotherapy delivery and improve therapeutic ratio
Hari Mukundan1, Sankalp Singh2, Nishant Lohia2, Manoj Kumar Semwal3, Vijendra Kumar1, Sharad Bhatnagar3, Arti Sarin4, Gaurav Trivedi2
1 Department of Radiation Oncology, Command Hospital (AF), Bengaluru, Karnataka, India 2 Department of Radiation Oncology, Command Hospital (CC), Lucknow, Uttar Pradesh, India 3 Department of Radiation Oncology, Army Hospital (R&R), New Delhi, India 4 Department of Radiation Oncology, Indian Naval Hospital Ship Asvini, Mumbai, Maharashtra, India
Date of Submission | 27-Oct-2020 |
Date of Decision | 21-Mar-2021 |
Date of Acceptance | 22-Mar-2021 |
Date of Web Publication | 5-May-2021 |
Correspondence Address: Dr. Sankalp Singh Department of Radiation Oncology, Command Hospital (CC), Lucknow, Uttar Pradesh India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/jmp.JMP_106_20
Abstract | | |
Background: The process of radiotherapy treatment planning and delivery involves multiple steps and professionals causing it to be prone to errors. Radiotherapy centers equipped with old telecobalt machines have certain peculiar challenges to workflow. We designed and tested a checklist for radiotherapy technicians (RTTs) to reduce chances of error during treatment delivery on a telecobalt machine. Materials and Methods: A physical checklist was designed for RTTs to use in the pretreatment pause using a template advocated by the American Association of Physicists in Medicine. It was tested on 4 RTTs over 1000 radiotherapy delivery sittings. Results: The checklist helped to rectify 41 documentary lapses and 28 errors in radiotherapy treatment parameters while also identifying 12 instances where treatment plan modifications were due and 30 where the patient was due for review by the radiation oncologist. The average time to go through the checklist was between 2.5 and 3 min. Conclusions: The development and use of the checklist has helped in reducing errors and also improving workflow in our department. It is recommended to utilize such physical checklists in all radiotherapy centers with telecobalt machines. The success of the checklist depends upon leadership, teamwork, acceptance of a need to inculcate a “safety culture,” with voluntary error-reporting and a willingness to learn from such errors.
Keywords: Checklist, error, radiotherapy, telecobalt
How to cite this article: Mukundan H, Singh S, Lohia N, Semwal MK, Kumar V, Bhatnagar S, Sarin A, Trivedi G. Use of a checklist approach on a telecobalt in an attempt to reduce human errors in radiotherapy delivery and improve therapeutic ratio. J Med Phys 2021;46:1-6 |
How to cite this URL: Mukundan H, Singh S, Lohia N, Semwal MK, Kumar V, Bhatnagar S, Sarin A, Trivedi G. Use of a checklist approach on a telecobalt in an attempt to reduce human errors in radiotherapy delivery and improve therapeutic ratio. J Med Phys [serial online] 2021 [cited 2023 Mar 31];46:1-6. Available from: https://www.jmp.org.in/text.asp?2021/46/1/1/315505 |
Introduction | |  |
Cancer is a leading cause of death and disability in the world today. Radiation Oncology is the field of medicine which utilizes ionizing radiation safely and effectively in the treatment, palliation, and cure of malignant diseases. Ionizing radiation is generated by highly sophisticated, complex, and expensive medical machines of essentially two types: linear accelerators (LINACs) and cobalt teletherapy units or telecobalts.
Radiotherapy treatment process has multiple steps carried out by a team of varied professionals with specific skills and responsibilities working together toward a common goal. The Radiotherapy Technician/Radiation Therapist (RTT) is the team member who executes the treatment plan on a day-to-day basis [Figure 1]. | Figure 1: Different steps in a radiotherapy treatment workflow. The figure also depicts the team members primarily responsible for each respective step
Click here to view |
A radiotherapy “error” can be described as a nonconformance where there is an unintended divergence between a radiotherapy process followed and that defined as correct by local protocol.[1] An error may or may not lead to a radiotherapy ”incident,” which is any unintended event that has consequences that are not negligible from the point of view of protection or safety of the patient.[2]
The multidisciplinary and multistep nature of radiotherapy makes it prone to errors at every step. Complex technologies, computer applications, multiple interphases, and the stressful and time-constrained milieu of modern medicine are some of the factors that may abet these errors. At the same time, the multistep nature of the radiotherapy (RT) process also endows multiple opportunities for detecting and correcting the errors in a modular and less confounding manner.
Errors in radiotherapy have been detected at every stage from documentation, simulation, treatment planning, treatment delivery, and also during transfer of information and instructions between various stages. While some can be blamed on machine failures and faulty software, the majority (60% or more) can be traced to human inattention and mistakes. Human errors can be caused by failures of communication, nonadherence to standard protocols, poor documentation, and oversight.
Checklists have been successfully used to prevent errors and improve performance in complex environments such as aviation, aeronautical and space engineering, architecture, and product manufacturing.[3],[4] They have been used in the field of medicine for the last 30 years and have repeatedly proven to be effective in reducing complications in high intensity fields such as critical care and surgery.[4],[5],[6],[7] Even in radiation oncology, they have been widely introduced in simulation, treatment planning, treatment delivery, and in audits of radiotherapy processes.[8],[9],[10],[11],[12] With the widespread use of computers in radiotherapy, most departments have multiple electronic checklists integrated into the radiotherapy information system/software.
Radiotherapy centers in developing countries like India face some peculiar challenges. Most departments in government sector hospitals treat a high volume of patients while the RTTs at many of these centers are understaffed and overworked. Furthermore, India still has 180 functioning Telecobalts which up one-thirds of all teletherapy machines in India.[13],[14] The majority of these are a decade old and are not compatible with electronic checklists.
Under these demanding and error-prone circumstances, we felt that there was an unmet need for a physical checklist for treatment delivery on the aging telecobalt. Physical checklists have been used successfully for radiotherapy audits and treatment simulation,[9],[12] though we could not find one in literature designed for treatment delivery on a telecobalt.
We have designed and tested a checklist for the last but very critical step of the RT process: treatment execution on a telecobalt unit by the RTTs. The RTTs were assigned the responsibility of conducting the checklist before treatment of every patient. We present the results of our exercise.
Materials and Methods | |  |
The American Association of Physicists in Medicine guidelines for developing safety checklists in radiotherapy[3] were used as a template to design the checklist. The step at which the checklist would be gone through was just before treatment delivery by the RTT. The development team included radiation oncologists, medical physicists, and RTTs. The first checklist that was developed consisted of three sections [Figure 2]a. The first section was used to check the recording of patient's details and consent on the treatment card to avoid regulatory lapses. As treatment delivery is the final step of the entire radiotherapy process, the second section was designed as a final check of all parameters of treatment planning and delivery as entered in the treatment card. The language was kept simple, direct, and unambiguous. Questions with only yes or no type responses were used and color coding was utilized for easy marking and auditing of detected faults. The third section was developed to help the RTT in detecting if the treatment plan was due for a modification or if the patient needed to be reviewed by the physician. | Figure 2: (a) First version of checklist used for pilot testing (see text). (b) Final iteration of the checklist used in the study (see text)
Click here to view |
The checklist would be run by two RTTs posted on the telecobalt unit with one completing all tasks while the other would read out and check each element during the pretreatment pause. We then conducted limited pilot testing of the checklist for 100 radiotherapy sittings or fractions, after a preimplementation training and familiarization exercise for the RTTs. Feedback was taken after this pilot phase and checklist design was re-evaluated. The major problems identified were the time taken (between 6 and 10 min per patient) to go through the list and some questions having neither yes nor no as an appropriate response.
After multiple such iterations, the final draft was accepted after incorporating several changes [Figure 2]b. Instead of three, the list was divided into two sections with section A having all questions relevant only on the 1st day of treatment. Section B contained questions for all treatment days.
The color coding was further simplified and some questions reframed to make them simpler and reduce confusion. A “not applicable” option for certain questions was also provided. After retraining four RTTs this final version was tested for 1000 treatment fractions over a period of 6 weeks. The average time taken to go through the list was also recorded for first 10 and last 10 fractions. The observations made during these 1000 fractions are now presented.
Results | |  |
Out of the 1000 checklists that were run, 902 had no observations with all parameters that were checked being in green or acceptable zone. In the remaining 98 checklists, there were 111 observations (89 had one observation, 6 had two, 2 had three, and 1 had four observations). The distribution of observations into types and the actual observations made are displayed in [Figure 3]a,[Figure 3]b,[Figure 3]c,[Figure 3]d. | Figure 3: (a) Distribution of observations made into types. (b) Observations of documentary lapses made. (c) Observations of treatment parameters requiring rectification that were made. (d) Observations of treatment plan modifications and physician's review that were due and were timely identified
Click here to view |
In the first 10 fractions, the average time taken to go through the list was 6 min on the 1st day of treatment and 5 min on other days. During the last 10 fractions this had improved to 3 min and 2.5 min, respectively.
Discussion | |  |
Though appearing complicated and error-prone, radiotherapy is one of the safest fields of medicine today.[15] The error rate during radiotherapy process reported in an American study was as low as 0.005%.[16] Others have reported it to vary between 0.017% when using computerized systems to 0.2% without the same.[17] In a safety review of radiotherapy facilities by the World Health Organization (WHO) and International Atomic Energy Agency (IAEA), 3125 patients suffered adverse effects due to radiotherapy treatment errors over 30 years.[18] Furthermore, 90% of radiotherapy errors reported in literature are clinically insignificant. However, due to the nature of ionizing radiation, it always has the potential to cause severe and lasting damage leading to permanent disability or even death. In the WHO review, 1% (38) of the errors did lead to death of the patient as a consequence of radiation toxicity.[18]
Studies on radiation error rates primarily originate from the developed parts of the world. negligible radiotherapy incidents are reported by most developing nations of Asia and Africa, including India though some studies have shown otherwise. A postal audit of Thermoluminescent dosimeter (TLD) between 1997 and 2004 by IAEA/WHO in developing countries suggested that only 84% patients received radiotherapy dose within the acceptable range.[19] Recently, a paper on an audit of radiotherapy processes at Tata Memorial Centre has been published which detected an error rate of 0.16 per 100 fractions or 4.1 per 100 patients.[13] Clearly, errors occur but are underreported.
In our study, we have focused on reducing errors and improving performance during treatment delivery using a checklist which can be a simple but effective tool at every step of the radiotherapy workflow.
Use of the checklist in our department brought about some obvious and positive improvements. First among these was the standardization and streamlining of workflow. The majority of the observations in our study were documentary lapses (41% or 37%). While they may be put down to simple clerical errors, a tendency of working with incomplete patient records shows an inattentive and careless workforce and can be hazardous under some circumstances. Missing details like disease diagnosis and laterality can make the process of doing an audit meaningless. Incomplete consent forms and absent signatures of doctors and physicists can lead to regulatory and medicolegal issues as well [Figure 3]b.
The checklist ensured all our treatment records were meticulously checked and corrected. It also increased the overall inclination of all team members toward maintaining completed records even before the treatment card is subjected to a checklist. The number of documentary errors per week fell from 15 in the 1st week to none in the last week following a constant downward trend [Figure 4].
A total of 28 discrepancies in treatment parameters were detected in our study constituting 25% of observations made. Several minor omissions like mentioning of treatment position or depth of dose prescription or marking of treatment fields and shielding blocks on treatment cards, which were affecting the workflow and protocol of the department as well as the quality of care were identified [Figure 3]c. Thanks to the strict following of the checklist protocol, all such mistakes were sent back for immediate rectification. This certainly enhanced the quality of care delivered at our center. There was a downward trend with the passage of weeks on the number of treatment parameters requiring correction [Figure 4].
The reduction in both documentary and treatment planning mistakes after the introduction of the checklist might be reflective of its positive effect on the streamlining of workflow, standardization of department practices, and improved compliance with laid-down protocols and procedures. These are benefits that would help any medical practice in improving patient safety and clinical outcomes.
The checklist also acted as a memory guide reminding the RTTs about timely treatment plan modifications and radiation oncologists' review of the patients whenever they were due, reducing the chances of missing out on these critical events.
While the checklist is simple and effective, its success cannot be attributed only to the use of a memory aid to avoid forgetting tasks. It was important to use it in the background of sensitization and empowerment of all stakeholders in the RT process to take responsibility for identifying and accepting mistakes in a professional manner. The idea to be categorically conveyed was that the ultimate goal of the checklist was to find errors without laying blame, and to improve the overall quality of workflow.
Finally, a subjective opinion of the RTTs on the use of the checklist. In the postimplementation feedback, the RTTs were unanimous in their approval. They felt a greater sense of confidence in their own work while treating patients with the checklist as they felt sure they had not missed any minor details. The feeling of empowerment and responsibility that came with using the checklist also added to their confidence. They felt less stressed while carrying out large volumes of repetitive tasks. They also submitted that their channel of communication with the doctors and physicists had also improved. Their only complaint was about the additional work, as expected.
Limitations of the study
Nothing comes without a price, and in this case, it was the extra time. As mentioned previously, changes were made to the checklist specifically to reduce the time taken to run it. By the end of the trial, the RTTs were accustomed to using the list and without compromising on deliberations could finish the list within 3½ min on the 1st treatment day and 2½ min on other days. Two and a half minutes does not sound like much but when done for 40 consecutive patients it works out to an hour and 40 min of additional work. Most departments in India have many more patients than that per shift. The other hardship faced was auditing all of the checklists which can be tedious and time consuming as an additional task.
A possible solution to this according to the authors is the use of an electronic checklist on a standalone computer terminal or even a hand-held device. With the ubiquity of smartphones and Wi-Fi internet, this should not be a problem. There are several proprietary software available in the market that allow one to customize their own checklist. With an electronic checklist, not only is the requirement of printed paper reduced but with the user-friendly interface the checklist tends to be run through faster also. And of course, auditing the lists becomes much simpler and quicker. We have developed such a checklist using a freely available software but are yet to test or validate it.
The true evidence of the success of an intervention is in the measurement of clinical outcomes before and after the intervention is introduced in a closed setting. However, to do so with checklists in radiotherapy can be challenging. This is due to the time it takes for clinical outcomes and toxicities to become evident in our field. On the other hand, detection of “near misses” and “radiation incidents” before and after the intervention would require detailed and extensive audits of the department and should be done after the checklist has been used for a longer period of at least 6 months. As a result, we could not use these parameters to assess the impact of the checklist on our practice. However, it can be planned for a subsequent longer study.
Conclusions | |  |
The development and use of the checklist has helped in reducing errors and also improving workflow in our department. In institutes that use telecobalts and do not use computerized systems of planning and treatment the chances of radiotherapy errors are higher. We recommend that till the time that computer-based treatment planning systems and modern LINACS replace the older telecobalts, our checklist can be used as an effective way of improving quality of care at such centers. Longer trials may be able to detect improvement in clinical outcomes as well. Finally, the checklist was effective because of good leadership, a positive sense of teamwork and an organizational acceptance of a need to inculcate a “safety culture” with voluntary error-reporting and to be willing learners from such errors.
Acknowledgment
Malignant Disease Treatment Centre (MDTC) Command Hospital (AF), Bengaluru and MDTC, Command Hospital (CC), Lucknow, India.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
References | |  |
1. | Towards Safer Radiotherapy. The Royal College of Radiologists, Society and College of Radiographers, Institute of Physics and Engineering in Medicine, National Patient Safety Agency, British Institute of Radiology. London: The Royal College of Radiologists; 2008. |
2. | International Atomic Energy Agency. IAEA Safety Glossary: Terminology Used in Nuclear Safety and Radiation Protection. Vienna, Austria: International Atomic Energy Agency; 2007. |
3. | Fong de Los Santos LE, Evans S, Ford EC, Gaiser JE, Hayden SE, Huffman KE, et al. Medical physics practice guideline 4.a: Development, implementation, use and maintenance of safety checklists. J Appl Clin Med Phys 2015;16:5431. |
4. | Hales BM, Pronovost PJ. The checklist – A tool for error management and performance improvement. J Crit Care 2006;21:231-5. |
5. | Pronovost P, Needham D, Berenholtz S, Sinopoli D, Chu H, Cosgrove S, et al. An intervention to decrease catheter-related bloodstream infections in the ICU. N Engl J Med 2006;355:2725-32. |
6. | Haynes AB, Weiser TG, Berry WR, Lipsitz SR, Breizat AH, Dellinger EP, et al. A surgical safety checklist to reduce morbidity and mortality in a global population. N Engl J Med 2009;360:491-9. |
7. | Gawande A. The Checklist Manifesto: How to Get Things Right. New York: Metropolitan Books; 2009. |
8. | |
9. | Asnaashari K, Gholami S, Khosravi HR. Lessons learnt from errors in radiotherapy centers. Int J Radiat Res 2014;12:361-7. |
10. | Albuquerque KV, Miller AA, Roeske JC. Implementation of electronic checklists in an oncology medical record: Initial clinical experience. J Oncol Pract 2011;7:222-6. |
11. | Berry SL, Tierney KP, Elguindi S, Mechalakos JG. Five years' experience with a customized electronic checklist for radiation therapy planning quality assurance in a multicampus institution. Pract Radiat Oncol 2018;8:279-86. |
12. | Meier T, Suh JH, Barrett P, Stevens EM, Hugebeck L, Chao ST. Presimulation procedure safety checklist/time out reduces simulation related unintended deviations. Int J Radiat Oncol Biol Phys 2017;99:E559. |
13. | Agarwal JP, Krishnatry R, Panda G, Pathak R, Vartak C, Kinhikar RA, et al. An audit for radiotherapy planning and treatment errors from a low-middle-income country centre. Clin Oncol (R Coll Radiol) 2019;31:e67-74. |
14. | Munshi A, Ganesh T, Mohanti BK. Radiotherapy in India: History, current scenario and proposed solutions. Indian J Cancer 2019;56:359-63.  [ PUBMED] [Full text] |
15. | Ganesh T. Incident reporting and learning in radiation oncology: Need of the hour. J Med Phys 2014;39:203-5. [Full text] |
16. | |
17. | Pennsylvania Patient Safety Authority. Errors in radiation therapy. Pa Patient Safety Advis 2009;6:87-92. |
18. | Shafiq J, Barton M, Noble D, Lemer C, Donaldson LJ. An international review of patient safety measures in radiotherapy practice. Radiother Oncol 2009;92:15-21. |
19. | Izewska J, Andreo P, Vatnitsky S, Shortt KR. The IAEA/WHO TLD postal dose quality audits for radiotherapy: A perspective of dosimetry practices at hospitals in developing countries. Radiother Oncol 2003;69:91-7. |
[Figure 1], [Figure 2], [Figure 3], [Figure 4]
|