ORIGINAL ARTICLE
Year : 2020  |  Volume : 45  |  Issue : 4  |  Page : 226-233
 

Is synchronous bilateral breast irradiation using flattening filter-free beam-based volumetric-modulated arc therapy beneficial? A dosimetric study


1 Department of Physics, School of Advanced Science, Vellore Institute of Technology, Vellore, Tamil Nadu; Department of Radiation Oncology, Yashoda Hospitals, Hyderabad, Telangana, India
2 Department of Physics, School of Advanced Science, Vellore Institute of Technology, Vellore, Tamil Nadu, India

Date of Submission02-May-2020
Date of Decision23-Oct-2020
Date of Acceptance29-Oct-2020
Date of Web Publication2-Feb-2021

Correspondence Address:
Dr. K Veluraja
Department of Physics, School of Advanced Science, Vellore Institute of Technology, Vellore - 632 014, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_32_20

Rights and Permissions

 

   Abstract 

Objective: The aim of this study is to validate the clinical use of flattening filter-free (FFF) beam-based volumetric-modulated arc therapy (VMAT) in synchronous bilateral breast carcinoma (SBBC) patient treatments and to compare with flattening filtered (FF) beam-based VMAT. Materials and Methods: Computed tomography images of 15 SBBC patients were taken for this study. A dose of 50 Gy in 25 fractions was prescribed to planning target volume (PTV). VMAT plans were generated using both FFF and FF 6 MV X-ray beams in Eclipse treatment planning system. PTV and organs at risk (OARs) doses were analyzed quantitatively using dose–volume histograms (DVHs) to meet plan objectives. Pretreatment point and planar dosimetry were performed. Results: The findings were reported as mean ± 1 standard deviation. PTV volume receiving 95% of the prescribed dose was 95.71% ± 0.65% for FF-VMAT and 95.45% ± 1.33% for FFF-VMAT (P = 0.743). Conformity index was 1.12 ± 0.31 (FF-VMAT) and 1.12 ± 0.02 (FFF-VMAT). Right lung mean dose was 10.95 ± 1.33 Gy (FF-VMAT) and 10.60 ± 98.5 (FFF-VMAT). Left lung mean dose was 9.73 ± 1.56 (FF-VMAT) and 9.61 ± 1.53 Gy (FFF-VMAT). Tumor control probability (TCP) was 99.68% ± 0.02% (FF-VMAT) and 99.67% ± 0.01% (FFF-VMAT) (P = 0.390). Uncomplicated TCP was 98.72% ± 0.02% (FF-VMAT) and 98.72% ± 0.01% (FFF-VMAT) (P = 0.508). Conclusion: The planning objective parameters achieved using FFF-based VMAT showed that FFF can also be used clinically to treat bilateral breast carcinomas and the low-dose lung volumes were still lesser with FFF-VMAT plans than FF-VMAT.


Keywords: Bilateral breast, breast radiotherapy, flattening filter free, SBBC, synchronous bilateral breast, volumetric-modulated arc therapy


How to cite this article:
Nagaraj J, Veluraja K. Is synchronous bilateral breast irradiation using flattening filter-free beam-based volumetric-modulated arc therapy beneficial? A dosimetric study. J Med Phys 2020;45:226-33

How to cite this URL:
Nagaraj J, Veluraja K. Is synchronous bilateral breast irradiation using flattening filter-free beam-based volumetric-modulated arc therapy beneficial? A dosimetric study. J Med Phys [serial online] 2020 [cited 2021 May 14];45:226-33. Available from: https://www.jmp.org.in/text.asp?2020/45/4/226/308610



   Introduction Top


The incidence of breast cancer is one of the highest among women.[1],[2] In breast cancer itself, the incidence of cancer in both the breasts (bilateral) is very rare. Bilateral breast cancers are divided into metachronous and synchronous depending on the time gap between detection of cancer in both the breasts. If detected less than a year gap, it is called synchronous bilateral breast carcinoma (SBBC); otherwise, it is called metachronous. The incidence rate of SBBC is lesser than metachronous. Many studies are indicating the range of synchronous bilateral breast carcinoma to be from 0.4% to 2.8% of total breast cancers.[3],[4],[5],[6] Compared to other breast cancers, the incidence of distant metastasis is higher and also disease-free survival is significantly less in synchronous bilateral breast carcinoma patients.[4],[5]

Surgery followed by adjuvant radiation therapy is the well-known treatment process of early-stage breast cancers. The radiation therapy techniques used for synchronous bilateral breast irradiation are the same as unilateral breast cancers, except the beam orientation, and the junction of both breasts makes the planning more complex. Furthermore, achieving the dose constraints of total lung and heart is very critical in the treatment of SBBC.[7],[8],[9] Many studies compared three-dimensional conformal radiotherapy (3DCRT) techniques with intensity-modulated radiotherapy (IMRT) for SBBC and shown that IMRT was more optimal with respect to organs at risk (OAR) dose and planning target volume (PTV) dose conformity in the selected cases.[5] The disadvantage of conventional IMRT is that the time required to deliver dose was much higher than 3DCRT and hence reduces the patient throughput in any busy radiotherapy department.

The introduction of volumetric-modulated arc therapy (VMAT) makes the special form of IMRT to deliver the dose in time comparable to 3DCRT and hence providing optimal radiation therapy treatments to breast cancer patients.[10],[11],[12] The advantage of VMAT has been more pronounced in the current radiotherapy practices. Helical tomotherapy is also another rotational method of treatment, but lags in the treatment time compared to VMAT.[13]

The radiation beams used mostly are flattening filtered (FF) beam from the accelerator head. The use of flattening filter-free (FFF) beam has increased more rapidly in recent years. Technical evolution of high-performance computers and accurate dose calculation algorithms enhance the treatment planning systems (TPSs) to counter the complex intensity modulation involving FFF. With FFF beam, the treatment time is considerably reduced for stereotactic treatments where higher dose per fraction is employed.

Vendors are coming up with linear accelerators enabled with FFF alone in recent times [e.g., Varian Edge™ manufactured by Varian Medical Systems (Palo Alto, USA), providing the option of buying high-intensity FFF beams alone]. The extent of usefulness of FFF is well established in hypofractionated treatments in the form of SRS and SBRT where the tumor size is smaller and the treated volumes are much lower than conventional RT with larger size tumors (with or without positive lymphatic nodes).[14],[15],[16],[17] SBBC is one such site, which requires large-field arrangements.

The advantage of using flattened beam-based VMAT for SBBC had been reported by Nicolini et al.[5] There are very scarce papers in literature, which suggest the role of FFF beam-based VMAT in the treatment of SBBC. Hence, we decided to validate the usefulness of FFF beam-based VMAT (FFF-VMAT) to treat SBBCs in comparison with FF beam-based VMAT (FF-VMAT) for conventional dose of 2 Gy per fraction.


   Materials and Methods Top


Patient characteristics

In this study, 15 synchronous bilateral breast carcinoma patients registered in our institute from 2010 to 2018 were selected. Among these 15 patients, 5 were right side total mastectomy, 6 were left side total mastectomy, and 4 were both breast conservative surgery performed patients. Of these 15 patients, 13 were treated using FF-VMAT and 2 were treated using forward IMRT.

Simulation

All patients were immobilized by following the institutional protocol using thermoplastic masks and VacLok with positions supine, head straight, and hands above head. All the patients were simulated in free breathing condition. Both plain and contrast-enhanced three-dimensional computed tomography (CT) images were acquired using 16-slice Siemens Somatom Scope (Siemens AG, Germany) with a slice thickness of 3 mm for contouring and treatment planning. Among these 15 patients, six patients underwent PET-CT and these images were rigid registered with the planning CT images to assist target delineation.

Planning

In 2013, we had upgraded our existing C-Series linear accelerator with FFF beam mode of delivery along with Eclipse TPS version 11 (Varian medical systems, Palo Alto, USA). For all 15 patients, new contours and plans (FF-VMAT and FFF-VMAT) were created regardless of old plan and technique in Eclipse TPS version 11. The Radiation Therapy Oncology Group (RTOG) breast contouring guidelines[18] were followed for contouring. The structures such as CTV (clinical target volume), PTV, PTVEval (PTV cropped by 5 mm from body for evaluation), right lung, left lung, heart, spinal cord, and healthy tissue were contoured. VMAT treatment plans were created using upgraded Clinac iX (Varian medical systems, Palo Alto, USA) linear accelerator, which is capable of delivering 6 MV FF and FFF beams. It is equipped with millennium 120 multileaf collimator and On-Board imager where kV and MV planar images and cone-beam computed tomography volume images can be acquired to verify the patient position before treatments.

The dose prescription to PTV was 50 Gy in 25 fractions. Planning goals were set as follows: for PTV, minimum 90% of the PTV volume should receive 95% of the prescribed dose and no volume should receive dose more than 115% of the prescribed dose as well as the global dose maximum should be inside the PTV. For OAR, lung dose constraints were 20 Gy volume should be <30% and mean lung dose should be <15 Gy; heart dose to be minimized as low as possible (mean <5 Gy was suggested in RTOG 1005) and spinal cord maximum dose should be <45 Gy.[19]

For all the patients, 6 MV FF VMAT plans were created in the TPS using two isocenters with four partial coplanar arcs, with two partial arcs placed at each isocenter as shown in [Figure 1]. The first isocenter was placed on the right side breast PTV and the second isocenter was placed in the same transverse plane on the left side breast PTV by moving lateral coordinate without changing vertical and longitudinal plane. For the right side target, the arc angle ranged from 230° to 60° clockwise (CW) and from 60° to 230° counterclockwise (CCW) with collimator rotation of 15° or 345°. For the left side target, it was 300° to 130° CW as well as CCW with collimator rotation of 15° or 345°.
Figure 1: Isocenter placement and volumetric-modulated arc therapy: arc beam geometry in axial plane

Click here to view


The direct aperture optimizer progressive resolution optimizer version 3 in Eclipse allows VMAT optimization of multiple isocenters as well as multiple arcs simultaneously in the same plan. The VMAT optimization was performed to get optimal dose distribution by giving lower and upper dose constraints along with priority. Both lower and upper constraints were used for PTV, while only upper constraints were used for critical structures. Priority ranks of 1, 2, 3, and 4 were given to PTV, lungs, heart, and spine, respectively. Normal tissue objective function available in Eclipse was used to reduce the dose to normal tissues. The dose was calculated on planning CT using analytical anisotropic algorithm in Eclipse TPS with calculation grid size of 2.5 mm. The final plan was taken as the reference. In the same way, FFF-based VMAT plans were created by keeping the same planning objectives as of the reference FF-VMAT plans only by changing the energy from 6 MV FF to 6 MV FFF, and the highest dose rate available for the respective energy was used. These two plans were compared using dosimetric and biological parameters.

Dosimetric evaluation

Cumulative dose–volume histograms (DVHs) were used to compare the dose coverage and plan quality parameters. For PTV, V95% (% of volume receiving 95% of the prescribed dose that is 47.5 Gy), V90% (volume receiving 90% of the prescribed dose), and D1cc (dose received by 1cc of the PTV volume) were used to compare dose coverage to PTV.[5] Conformity index (CI) and heterogeneity index (HI) were used to compare the quality of plans. The following formulas were used to compute the CI and HI.

CI95% = V95%/VPTV (1)

where V95% = volume covered by 95% of the dose (cc), VPTV = volume of the PTV (cc)

HI = D5%/D95% (2)

where D5% = dose received by 5% volume of PTV (Gy), D95% = dose received by 95% volume of PTV (Gy). To compare dose to lungs, the parameters V5 Gy (% of volume receiving dose of 5 Gy), V10 Gy (% of volume receiving dose of 10 Gy), V20 Gy (% of volume receiving dose of 20 Gy), and Dmean (mean dose) were used. For heart, parameters such as Dmean, D2% (dose received by 2% of volume), V10 Gy (% of volume receiving dose of 10 Gy), and V45 Gy (% of volume receiving dose of 45 Gy) were used.

Healthy tissue (body subtracted from PTV) dose was compared using Dmean, V3 Gy (% of volume receiving dose of 3 Gy), V10 Gy (% of volume receiving dose of 10 Gy), external volume index (EI), and integral dose (ID).[5] EI was calculated as VD/VPTV, where VD is the volume of healthy tissue receiving dose more than the prescribed dose and VPTV is the volume of PTV.

Biological evaluation

Biological parameters, tumor control probability (TCP) and normal tissue complication probability (NTCP), were also calculated to check the quality of plans using cumulative DVHs. The cumulative DVHs were exported with dose bin size of 1cGy in text format and converted to tabular format in Microsoft Excel for calculating equivalent uniform dose (EUD), TCP, and NTCP. The dose in DVH was converted to equivalent dose of 2 Gy (EQD2 Gy) using L-Q model[20] to find the EUD. The formula used to calculate EQD2 Gy is as follows:

EQD2 Gy = D(α/β + D/n)/(α/β + 2) (3)

where D = nd (d is the dose per fraction and n is the number of fractions); α/β = ratio that gives the dose at which linear (α) and quadratic (β) components of cell killing are equal. EUD was calculated to convert the heterogeneous dose distribution to homogeneous dose that produces the same biological effect based on the Niemierko's model.[21]

EUD= (ΣiviDia)1/a (4)

where Di is the equivalent dose of 2 Gy corresponding to ith bin and vi is the partial volume receiving the dose Di, a is the model parameter.

The calculated EUD was used to calculate the TCP of PTV and NTCP of normal structures based on Niemierko's models.[22],[23],[24],[25],[26] The formulas used were as follows:

TCP = 1 / [1 + (TCD50 / EUD)^4γ50] (5)

NTCP = 1 / [1 + (TD50/5 / EUD)^ 4γ50] (6)

Where γ50 = slope of the dose response curve at a dose of 50% complication or control probability; TCD50 = tumor dose for 50% TCP; and TD50/5 = normal tissue dose for 50% complication probability in 5 years. The values used in the calculation of EQD, EUD, TCP, and NTCP are given in [Table 1]. The therapeutic gain can be obtained using uncomplicated TCP (UTCP). UTCP was calculated as follows:
Table 1: Biological parameters for planning target volume and organs at risks

Click here to view


UTCP = (TCP × πi (1 − NTCPi)) (7)

where NTCPi is the NTCP of organ i (e.g., if i = lung, NTCPi = NTCP of organ lung).

Pretreatment verification

Pretreatment plan verification was done by point dose and planar dose measurements. Verification plans were created for both plans without resetting the collimator angles and with preset MU values as of original plans in TPS for point dose and planar dose measurements. Both FF-VMAT and FFF-VMAT plans were executed on verification phantoms as like actual plan setup just by moving the couch lateral values as shown in [Figure 2].
Figure 2: Phantom and detector setup for point and planar dosimetry

Click here to view


Point dose

Point doses were measured at isocenter for all fields using CC13 (IBA Dosimetry, Gmbh, Germany) (cavity volume 0.13cc) cylindrical chamber and slab phantom made up of PMMA plastic material (30 cm × 30 cm × 10 cm). The point of measurement was kept in the low-gradient high-dose regions. Two points from the left side and right side high dose as well as low gradient were identified in the lateral direction and measured the dose in those points by keeping the couch separation same as of the original plan. The cumulative variations of measured dose from the TPS-predicted dose for both techniques were tabulated.

Planar dose

Multicube phantom with I-matrix array detectors (31.5 cm × 34 cm × 22 cm) and OmniPro IMRT software (IBA Dosimetry, Gmbh, Germany) were used for planar dosimetry. The detector plane in coronal view was taken for comparison and dose plane was exported in DICOM format to Omnipro IMRT software to compare with the corresponding measured dose maps. In this process, a setup field was introduced in the verification plan at coordinates as of user origin in the middle of the detector as shown in [Figure 2] and used to align all the fields when exporting the plan in DICOM format. The 2D gamma evaluation was done to evaluate both the plans on the usable detector area of central 24 cm × 24 cm. Distance to agreement (DTA) and dose difference (DD) criteria of 3 mm and 3%, 2 mm and 3%, and 2 mm and 2% were used to compute the gamma value for comparison of both the plans. The percentage of points passing the 2D gamma value of ≤1 was calculated and tabulated.

Efficiency of plan

Efficiency of both the plans was compared using the total number of monitor units (MU) and beam ON time for each plan by considering the rest of the setup and simulation procedures were the same for both the techniques.

Statistical analysis

Statistical analyses were made using Microsoft Excel spreadsheet. To find the significance of difference, Student's paired t-test with two tails was used. The null hypothesis set was that both FF-VMAT and FFF-VMAT were having the same mean with 95% confidence limit. Thus, if the probability value (P) is ≤0.05, the differences in the two techniques are statistically significant.


   Results Top


The plans which satisfied the clinical goals with acceptable limitations for both FF-VMAT and FFF-VMAT plans were taken for comparison. The dose distribution achieved by both the plans for a sample patient is shown in [Figure 3] with dose levels of 47.5, 45 , and 20 Gy.
Figure 3: Dose distribution using flattening filtered volumetric-modulated arc therapy (left) and flattening filter-free volumetric-modulated arc therapy (right) for the sample patient

Click here to view


The mean results of all patients were reported with one standard deviation for all evaluating parameters. The results of physical and biological parameters for PTV are given in [Table 2] with their P values. Dosimetric and biological parameters for OARs and normal healthy tissue are given in [Table 3]. The corresponding P values were also given for each parameter under evaluation for both the techniques. The UTCP was 98.72% ± 0.02% (FF-VMAT) and 98.72% ± 0.01% (FFF-VMAT) (P = 0.508).
Table 2: Planning target volume: dosimetric and biological comparison results

Click here to view
Table 3: Organs at risk: dosimetric and biological comparison results

Click here to view


The gamma analysis window of both the techniques of a sample patient is shown in [Figure 4] for different gamma index analysis (GIA) criteria. The analysis was made first by taking the entire measured area (GIA total) and then by keeping the region of interest in junction where the overlap occurs usually (GIA at junction) as indicated with red rectangle in [Figure 4]. Patient's plan verification measurement results were tabulated along with the number of MUs and treatment beam ON time in [Table 4].
Figure 4: Gamma analysis window of different sets of analysis criteria of the sample patient's planar dosimetry for both the techniques

Click here to view
Table 4: Delivery parameters result

Click here to view



   Discussion Top


In contrast to unilateral breast irradiation, the treatment planning to SBBC is complex due to the involved critical organ's doses. It was already proved that VMAT showed better distribution than IMRT with FF for SBBC.[5] In this dosimetric study, the feasibility of using FFF beam-based VMAT to treat SBBC was investigated and compared with FF beam-based VMAT. The results of this study showed that the FFF-VMAT plans were as good as FF-VMAT plans.

Planning target volume and dose-volume comparison

Dose distributions achieved with respect to PTV in both FF-VMAT and FFF-VMAT plans were equivalent and the differences were very minimal [Figure 2]. The maximum doses to all patients were less than 110% of the prescribed dose, and in these patients, the variations of maximum doses for PTV between both the plans were from 1% to 2%. The DVH data also showed that the PTV coverage for both the plans were similar and the differences were statistically insignificant for parameters such as PTV V95%, V90%, D1cc, CI95%, and HI [Table 2]. It was established that the PTV results were equivalent to FF-VMAT results of single-side breast cancer.[27],[28] The average DVH of all 15 patients for each structure under evaluation (PTV, healthy tissue, lungs, heart, and spinal cord) are given in [Figure 5] for comparison.
Figure 5: Average dose–volume histograms of all 15 patients under study

Click here to view


Lung dose–volume histogram comparison

With respect to lungs, the left side lung showed no statistically significant differences in the mean lung dose, 20 Gy lung volume and 5 Gy lung volume between both FF-VMAT and FFF-VMAT. The P values were much higher than 0.05 as shown in [Table 3]. The mean dose and V20Gy for right lung from both the plans showed that the differences were insignificant as shown in [Table 3]. Whereas, FFF-VMAT showed statistically significant reduction in V5Gy volume for right lung compared to FF-VMAT with P value of 0.039.

Although both the lungs are having the mean dose lesser than 15 Gy, the mean doses are higher for right side than the left side for all patients. The lung volume near to the PTV is always less in the left side due to the presence of heart and hence reducing the mean dose to left side lung.

Heart, spine, and healthy tissue

Heart mean doses showed random variations for both the plans [Figure 5] and statistically insignificant with P value of 0.869 [Table 3]. However, the mean heart dose (<5 Gy) was not achieved in all cases since we attempted to keep as low as possible without compromising the PTV coverage. The maximal heart distance and maximal heart length were found to be more in free breathing condition for these patients, which contributed significantly to high mean dose to heart.[28] It can be reduced further using the combinations of breath-hold techniques and gated delivery systems.[29]

Spinal cord maximum dose was always <45 Gy in all patients for both the plans. The P value of 0.954 for spine maximum dose showed that there was no significant difference between the two plans.

Healthy tissue 3 Gy volume and 10 Gy volume showed a statistically significant difference and FFF-VMAT was superior to FF-VMAT. The difference may be because of the reduced average energy of FFF. On the other hand, there are no significant difference between two techniques in EI and ID.

Biological comparisons

Biological parameter EUD for all structures showed no statistically significant difference with P value much higher than 0.05. For PTV, the TCP values for all the patients were above 99% for both the plans, and the differences were insignificant statistically. NTCP and UTCP values also suggested that there were no statistically significant differences between FFF-VMAT and FF-VMAT. Hence, these two plans can be considered equivalent to each other based on biological and physical parameters irrespective of shape of the beams before intensity modulation.

Efficiency comparisons

Lesser MU and lesser beam ON time are the desired characteristics of a better plan of similar distributions. The plan which has lesser overall treatment time is considered to be the best plan with respect to patient throughput. From [Table 4], FF-VMAT was better with respect to lesser MU than FFF-VMAT with a P value of 0.006. At the same time, beam ON time showed no statistically significant difference between both plans, which means that both the plans can be delivered in the same time with more number of MUs for FFF-VMAT. The availability of high dose rates in FFF mode compensates the time of beam ON for FFF-VMAT. We had expected relatively less treatment time with FFF-VMAT, but because of the conventional dose of 2 Gy per fraction, the maximum dose rate was not boosting more than the conventional level of 600 MU/min.[30] Hence, the overall treatment time remainsed the same for both the plans.

Deliverability comparison

The plan deliverability check with point dose and planar dose measurements also showed no significant difference between these two VMAT plans. Point dose variations were well within the institutional protocol limit of 3% for all patients in both techniques. Furthermore, the gamma passing rates were more than 95% for both the plans with clinically accepted GIA criteria of 3 mm and 3%. This indicates that both plans deliverability are equivalent in accordance with TPS.

Future study

Although the standard conventional dose of 50 Gy in 25 fractions has been followed in our center for SBBCs, the FAST-Forward trial (26 Gy in 5 fractions)[31] is being adopted in many centers across the world for breast cancers. The introduction of FFF may reduce the treatment time for the dose of 5.2 Gy per fraction and we want to study its significance in our future work and we would like to elaborate this study with deep inspiration breath-hold gated therapy.


   Conclusion Top


There is always a conflict in using FFF beams for large-field treatments and our results showed that there were no dosimetric differences between FFF-VMAT and FF-VMAT plans and also proved that FFF beams could also be used to treat large-field breast cancers. It is inferred that even though the MU for FFF beams were higher than FF beams, the beam ON time remained the same for conventional dose of 2 Gy per fraction. Moreover, the low-dose lung volumes are quite less with FFF beams. Thus, FFF-based VMAT delivery can also be preferred equally for SBBCs as of FF-VMAT.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Gupta S. Breast cancer: Indian experience, data, and evidence. South Asian J Cancer 2016;5:85-6.  Back to cited text no. 1
[PUBMED]  [Full text]  
2.
Ferlay J, Soerjomataram I, Dikshit R, Eser S, Mathers C, Rebelo M, et al. Cancer incidence and mortality worldwide: Sources, methods and major patterns in GLOBOCAN 2012. Int J Cancer 2015;136:E359-86.  Back to cited text no. 2
    
3.
Wadasadawala T, Visariya B, Sarin R, Upreti RR, Paul S, Phurailatpam R. Use of tomotherapy in treatment of synchronous bilateral breast cancer: Dosimetric comparison study. Br J Radiol 2015;88:20140612. doi:10.1259/bjr.20140612.  Back to cited text no. 3
    
4.
Heron DE, Komarnicky LT, Hyslop T, Schwartz GF, Mansfield CM. Bilateral breast carcinoma: Risk factors and outcomes for patients with synchronous and metachronous disease. Cancer 2000;88:2739-50.  Back to cited text no. 4
    
5.
Nicolini G, Clivio A, Fogliata A, Vanetti E, Cozzi L. Simultaneous integrated boost radiotherapy for bilateral breast: A treatment planning and dosimetric comparison for volumetric modulated arc and fixed field intensity modulated therapy. Radiat Oncol 2009;4:27.  Back to cited text no. 5
    
6.
Kheirelseid EA, Jumustafa H, Miller N, Curran C, Sweeney K, Malone C, et al. Bilateral breast cancer: Analysis of incidence, outcome, survival and disease characteristics. Breast Cancer Res Treat 2011;126:131-40.  Back to cited text no. 6
    
7.
Marks LB, Bentzen SM, Deasy JO, Kong FM, Bradley JD, Vogelius IS, et al. Radiation dose-volume effects in the lung. Int J Radiat Oncol Biol Phys. 2010 Mar 1;76(3 Suppl):S70-6. doi: 10.1016/j.ijrobp.2009.06.091. PMID: 20171521; PMCID: PMC3576042.  Back to cited text no. 7
    
8.
Munshi A, Talapatra K, Dutta D. Breast cancer radiotherapy and cardiac risk. Oncol Rev 2010;4:137-46.  Back to cited text no. 8
    
9.
Darby SC, Ewertz M, McGale P, Bennet AM, Blom-Goldman U, Brønnum D, et al. Risk of ischemic heart disease in women after radiotherapy for breast cancer. N Engl J Med 2013;368:987-98.  Back to cited text no. 9
    
10.
Yu CX, Li XA, Ma L, Chen D, Naqvi S, Shepard D, et al. Clinical implementation of intensity-modulated arc therapy. Int J Radiat Oncol Biol Phys 2002;53:453-63.  Back to cited text no. 10
    
11.
Otto K. Volumetric modulated arc therapy: IMRT in a single arc. Med Phys 2008;35:310-7.  Back to cited text no. 11
    
12.
Balaji K, Subramanian B, Yadav P, Anu Radha C, Ramasubramanian V. Radiation therapy for breast cancer: Literature review. Med Dosim 2016;41:253-7.  Back to cited text no. 12
    
13.
Yu CX. Intensity-modulated arc therapy with dynamic multileaf collimation: An alternative to tomotherapy. Phys Med Biol 1995;40:1435-49.  Back to cited text no. 13
    
14.
Hrbacek J, Lang S, Graydon SN, Klöck S, Riesterer O. Dosimetric comparison of flattened and unflattened beams for stereotactic ablative radiotherapy of stage I non-small cell lung cancer. Med Phys. 2014;41:031709. doi: 10.1118/1.4866231. PMID: 24593713.  Back to cited text no. 14
    
15.
Pönisch F, Titt U, Vassiliev ON, Kry SF, Mohan R. Properties of unflattened photon beams shaped by a multileaf collimator. Med Phys 2006;33:1738-46.  Back to cited text no. 15
    
16.
Subramanian SV, Subramani V, Swamy ST, Gandhi A, Chilukuri S, Kathirvel M. Is 5 mm MMLC suitable for VMAT-based lung SBRT? A dosimetric comparison with 2.5 mm HDMLC using RTOG-0813 treatment planning criteria for both conventional and high-dose flattening filter-free photon beams. J Appl Clin Med Phys 2015;16:112-24.  Back to cited text no. 16
    
17.
Subramaniam S, Thirumalaiswamy S, Srinivas C, Gandhi GA, Kathirvel M, Kumar KK, et al. Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams. Strahlenther Onkol 2012;188:484-90.  Back to cited text no. 17
    
18.
RTOG-Endorsed Consensus Guidelines for Delineation of Target and Normal Structures for Breast Cancer. Available from: http://www.rtog.org/CoreLab/ContouringAtlases/BreastCancerAtlas.aspx. [Last accessed on 2019 Dec 05].  Back to cited text no. 18
    
19.
A Phase III Trial Of Accelerated Whole Breast Irradiation With Hypofractionation Plus Concurrent Boost Versus Standard Whole Breast Irradiation Plus Sequential Boost For Early-Stage Breast Cancer, RTOG 1005; 2014.  Back to cited text no. 19
    
20.
Rana S, Rogers K. Radiobiological evaluation of dose calculation algorithms in RapidArc planning of esophageal cancer treatment plans. J Solid Tumors 2013;3:44.  Back to cited text no. 20
    
21.
Niemierko A. Reporting and analyzing dose distributions: A concept of equivalent uniform dose. Med Phys 1997;24:103-10.  Back to cited text no. 21
    
22.
Niemierko A, Goitein M. Implementation of a model for estimating tumor control probability for an inhomogeneously irradiated tumor. Radiother Oncol 1993;29:140-7.  Back to cited text no. 22
    
23.
Niemierko A, Goitein M. Modeling of normal tissue response to radiation: The critical volume model. Int J Radiat Oncol Biol Phys 1993;25:135-45.  Back to cited text no. 23
    
24.
Oinam AS, Singh L, Shukla A, Ghoshal S, Kapoor R, Sharma SC. Dose volume histogram analysis and comparison of different radiobiological models using in-house developed software. J Med Phys 2011;36:220-9.  Back to cited text no. 24
[PUBMED]  [Full text]  
25.
Okunieff P, Morgan D, Niemierko A, Suit HD. Radiation dose-response of human tumors. Int J Radiat Oncol Biol Phys 1995;32:1227-37.  Back to cited text no. 25
    
26.
Yarnold J, Bentzen SM, Coles C, Haviland J. Hypofractionated whole-breast radiotherapy for women with early breast cancer: Myths and realities. Int J Radiat Oncol Biol Phys 2011;79:1-9.  Back to cited text no. 26
    
27.
Aly MM, Glatting G, Jahnke L, Wenz F, Abo-Madyan Y. Comparison of breast simultaneous integrated boost (SIB) radiotherapy techniques. Radiat Oncol 2015;10:139.  Back to cited text no. 27
    
28.
Shim JG, Kim JK, Park W, Seo JM, Hong CS, Song KW, et al. Dose-volume analysis of lung and heart according to respiration in breast cancer patients treated with breast conserving surgery. J Breast Cancer 2012;15:105-10.  Back to cited text no. 28
    
29.
Swamy ST, Radha CA, Kathirvel M, Arun G, Subramanian S. Feasibility study of deep inspiration breath-hold based volumetric modulated arc therapy for locally advanced left sided breast cancer patients. Asian Pac J Cancer Prev 2014;15:9033-8.  Back to cited text no. 29
    
30.
Sangaiah A, Ganesh KM, Ramalingam K, Karthikeyan K, Jagadheeskumar N. Dosimetric validation of volumetric modulated arc therapy (VMAT) using AAPM TG-119 benchmark plans in an upgraded CLINAC 2100CD for flattening filter free (FFF) photon beams. Asian Pac J Cancer Prev 2017;18:2965-70.  Back to cited text no. 30
    
31.
Murray Brunt A, Haviland JS, Wheatley DA, Sydenham MA, Alhasso A, Bloomfield DJ, et al. Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial. Lancet 2020;395:1613-26.  Back to cited text no. 31
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,879 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed310    
    Printed8    
    Emailed0    
    PDF Downloaded99    
    Comments [Add]    

Recommend this journal