TECHNICAL NOTE |
|
|
|
Year : 2014 | Volume
: 39
| Issue : 2 | Page : 121-126 |
|
Comparative analysis of volumetric-modulated arc therapy and intensity-modulated radiotherapy for base of tongue cancer
L Nithya1, N Arunai Nambi Raj2, Arulraj Kumar1, Sasikumar Rathinamuthu1, Manish Bhushan Pandey1
1 Department of Radiation Oncology, Dharamshila Hospital and Research Centre, Vasundhara Enclave, Delhi, India 2 School of Advanced Sciences, VIT University, Vellore, Tamil Nadu, India
Date of Submission | 19-Dec-2013 |
Date of Decision | 13-Mar-2014 |
Date of Acceptance | 17-Mar-2014 |
Date of Web Publication | 23-Apr-2014 |
Correspondence Address: L Nithya Chief Medical Physicist and RSO, Dharamshila Hospital and Research Centre, Vasundhara Enclave, Delhi 110 096 India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0971-6203.131288
Abstract | | |
The aim of this study was to compare the various dosimetric parameters of dynamic multileaf collimator (MLC) intensity modulated radiation therapy (IMRT) plans with volumetric modulated arc therapy (VMAT) plans for base of tongue cases. All plans were done in Monaco planning system for Elekta synergy linear accelerator with 80 MLC. IMRT plans were planned with nine stationary beams, and VMAT plans were done for 360° arc with single arc or dual arc. The dose to the planning target volumes (PTV) for 70, 63, and 56 Gy was compared. The dose to 95, 98, and 50% volume of PTV were analyzed. The homogeneity index (HI) and the conformity index (CI) of the PTV 70 were also analyzed. IMRT and VMAT plan showed similar dose coverage, HI, and CI. Maximum dose and dose to 1-cc volume of spinal cord, planning risk volume (PRV) cord, and brain stem were compared. IMRT plan and VMAT plan showed similar results except for the 1 cc of PRV cord that received slightly higher dose in VMAT plan. Mean dose and dose to 50% volume of right and left parotid glands were analyzed. VMAT plan gave better sparing of parotid glands than IMRT. In normal tissue dose analyses VMAT was better than IMRT. The number of monitor units (MU) required for delivering the good quality of the plan and the time required to deliver the plan for IMRT and VMAT were compared. The number of MUs for VMAT was higher than that of IMRT plans. However, the delivery time was reduced by a factor of two for VMAT compared with IMRT. VMAT plans yielded good quality of the plan compared with IMRT, resulting in reduced treatment time and improved efficiency for base of tongue cases.
Keywords: Base of tongue, intensity-modulated radiation therapy, volumetric modulated arc therapy
How to cite this article: Nithya L, Raj N A, Kumar A, Rathinamuthu S, Pandey MB. Comparative analysis of volumetric-modulated arc therapy and intensity-modulated radiotherapy for base of tongue cancer. J Med Phys 2014;39:121-6 |
How to cite this URL: Nithya L, Raj N A, Kumar A, Rathinamuthu S, Pandey MB. Comparative analysis of volumetric-modulated arc therapy and intensity-modulated radiotherapy for base of tongue cancer. J Med Phys [serial online] 2014 [cited 2023 Mar 24];39:121-6. Available from: https://www.jmp.org.in/text.asp?2014/39/2/121/131288 |
Introduction | |  |
Intensity-modulated radiation therapy (IMRT) has replaced the conventional three-dimensional-conformal radiotherapy (3D-CRT) techniques for head and neck cancers. Already it has been proved by many researchers that IMRT gives better dose coverage to the target and sparing of organ at risk (OAR) [1] than 3D-CRT. IMRT gives less radiation-related toxicity than 3D-CRT for head and neck cancers. [2] Using IMRT we are able to achieve good uniformity of dose to concave-shaped head and neck tumor and dose conformal to target volumes. Treatment planning of head neck cases are challenging due to the large number of OARs and the proximity of OARs with tumors. The planner faces the difficulty in sparing the OARs with good coverage and uniformity of the dose to the tumor volumes.
Volumetric-modulated arc therapy (VMAT) is the next generation and enhanced version of IMRT technique. Most of the centers are adapting VMAT technique due to its fastest and efficient delivery than IMRT. [3] Quick delivery reduces the inconvenience to the patients and also increases the throughput of the linear accelerator. Many reports have explained the dosimetric superiority and similarity of VMAT with IMRT in head and neck cases. [3],[4] Most of the cases compared multiple head and neck sites like oropharynx, hypopharynx, nasopharynx, and larynx. [5] In this study, we would like to evaluate the IMRT and VMAT plans for base of tongue case.
Materials and Methods | |  |
Patient selection
A total of 13 base of tongue cases were selected for this study. The median age was 61 with 12 males and one female. The clinical stage distribution was T2 in four patients, T3 in six patients, and T4 in four patients with node stage N1-N3.
Eight of the patients received the treatment with IMRT plan and five of them received treatment with VMAT plan. VMAT and IMRT plans were created with same isocenter retrospectively. Both IMRT and VMAT plans were done in the same computed tomography (CT) and structure set.
Linac and record and verify system
All IMRT and VMAT plans were created using the same 6-MV photon beams for Elekta Synergy linear accelerator (Elekta Ltd, Crawley, UK), which was equipped with 80 multileaf collimator (MLC) with 1-cm resolution. Maximum leaf speed was 2.0 cm/s, minimum leaf gap between opposite banks was 0.5 cm, the total leaf travel distance was 32.5 cm, and the leaves did not interdigitate. In the upper jaws and backup jaws covering full 40 × 40 cm 2 , the maximum gantry speed was 6°/s and the variable dose rate upto 500 MU (monitor units)/min. The combination of the dose rate, gantry speed, and leaf speed was automatically selected optimally by the linear accelerator control system Precise Desktop 7.01 during VMAT delivery.
MOSAIQ Version 1.60X6 (IMPAC Medical Systems, Inc, Sunnyvale, CA, USA) was used as record and verify system. CT images from CT machine was transferred to treatment planning system (TPS) via Digital imaging and Communications in Medicine (DICOM), and treatment plan from TPS was transferred to MOSAIQ via DICOM-Radiotherapy (DICOM-RT).
Imaging
The patients were immobilized with thermoplastic sheet. The CT images were acquired for all patients in Biograph positron emission tomography-CT (PET-CT) (Siemens AG, Medical solutions, Germany) with 3-mm slice thickness, field of view 50 cm. The images were imported in Monaco planning system.
Monaco planning system
Monaco planning system (Elekta Ltd, Crawly, UK) version 3.20.02 utilizes not only physical effects of radiation but also biological properties of the tissue. It has three biological constraints such as target EUD, parallel, and serial and six physical constraints such as target penalty, quadratic overdose, overdose DVH, underdose DVH, maximum dose, and quadratic underdose.
The user has an option to set the cell sensitivity of Poisson's cell kill model. OAR can be set as serial or parallel constraints depending on the properties of the tissue. The user is able to influence the amount of volume of a structure, which can be sacrificed and also affect the strengths of the limits of the constraints.
The system uses a two step optimization process. In the first phase of optimization, the system creates user-specified segments, which also operate as control points at the treatment units during treatment. Then the calculation algorithm of the first phase optimizes the fluence distributions for the segments. At this phase, fluence distribution is calculated by pencil beam algorithm (PB), which considers dose to be produced by the sum of narrow cylindrically symmetry beams. The dose at any point is calculated by summing the calculated dose distribution of all pencil beams to the examination point. Because the algorithm is kernel-based two-dimensional method, so the accuracy is limited especially in the presence of heterogeneities.
In the second calculation phase, the system takes into account the deliverability of the accelerator. The optimal field shapes are then generated in order to produce as smooth composite dose delivery as possible. As a result, plan quality might be affected in comparison with the fluence distribution calculated in the first phase. The second phase calculation can be done by PB or voxel-based Monte Carlo method. The Monte Carlo method enables more accurate dose distribution calculation than the analytical algorithms. [6] The Monte Carlo method calculates the number of electrons created in each voxel by primary photon beam and particles created in interactions. The final calculated dose distribution is formed by most presumable distribution of the absorbed dose. The user can change the calculation accuracy and time by modifying some parameters like Monte Carlo grid spacing and variance.
Contouring
The contouring of tumor volumes and normal structures was done by the oncologists.
Target volumes
The clinical target volume (CTV 70) encompasses the gross tumor volume (GTV) and 5-mm margin. The planning target volume (PTV 70 ) for CTV 70 was created with 3-mm margin for CTV 70 to account for the set up uncertainties. The secondary target was the CTV of lymph node groups with high risk or surgical neck levels at risk of subclinical disease named as CTV 63. PTV 63 was created with 3-mm margin to CTV 63. The lymph node with low risk was contoured as CTV 56. The 3-mm margin that was given to CTV 56 to create PTV 56 . PTV 63 and PTV 56 were cropped 2 mm inside the body contour automatically by the system. No manual correction was done for the PTVs.
Critical structures
The normal structures included the spinal cord, parotid glands, and brain stem. The spinal cord was expanded by 5 mm to create planning risk volume (PRV), which was called PRV cord. Both the parotid glands were contoured and parotid minus PTV was created using 3-mm outer margin to PTV. Parotid minus PTV was utilized for optimization, and evaluation was done for the whole parotid.
Normal tissues
The body contour minus all tumor volumes and critical structures was taken as normal tissue.
Treatment planning and evaluation criteria
Both VMAT and IMRT plans were generated for a treatment in 35 fractions, to deliver total dose of 70 Gy PTV 70 , 63 Gy to PTV 63 , and 56 Gy to PTV 56 . The goal of treatment planning was to cover at least 95% of the volume of PTV 70 , PTV 63 , and PTV 56 with 98% of the prescribed dose (68.6, 61.74, and 54.88 Gy, respectively) and to restrict the volume of PTV 70 receiving more than 110% of the prescribed dose (77 Gy) should be below 10% volume of PTV 70 . The maximum allowed point dose to spinal cord was 45Gy; PRV cord was 50 Gy and for brain stem was 54 Gy. Also, the mean dose to both parotid glands should be below 26 Gy. All parameters are summarized in [Table 1].
Planning criteria in TPS
All IMRT and VMAT plans were planned with following calculation properties: Grid spacing was selected as 3 mm, and Monte Carlo variance was 3%. Monte Carlo algorithm was selected as secondary algorithm for second stage dose calculation, i.e., final dose calculation. The global parameter: Beamlet width was selected as 2 mm, target margin and avoidance margin was 3-4 mm. Auto flash margin was 2 mm, surface margin was 3 mm. Minimum electron density for filled option was selected as 1.000 HU and minimum CT number for clear option as 200 HU. The dose was calculated to the medium not to the water. For all plans, inhomogeneity correction was applied.
IMRT treatment planning
All IMRT plans were created using nine fields with 6-MV photon beam. The gantry angle started from 200° to 160° with 40° intervals. The collimator angle was kept as 0°. For all plans, dynamic MLC (dmlc) method was selected. In sequential parameters, the nominal dose rate was set between 120 and 150 MU/min. Dose rate was user selectable. The vendor recommended the dose rate range for Elekta machine was between 120-360 MU/min. For complex plan, low dose rate was recommended to achieve good quality of plan with high MU efficiency; thus very low dose rate was selected. The minimum segment width was kept as 5 mm and the fluence smoothing level was at medium level.
VMAT treatment planning
All VMAT plans were planned with 360° arc with 20° increments (angular spacing between sampled fluence profiles in stage 1) with 6-MV photon. All VMAT plans were started with single arc. If good plan was not achieved with single arc, double arc plan was done. A total of four patients were planned with single arc in clockwise direction, and nine patients were planned with double arc in clockwise and counter clockwise direction. Segment shape optimization (SSO) method was used. In SSO method, user could not be able to select the dose rate. The optimizer controlled the dose rate. This results the better plan quality. Minimum segment width was kept as 5 mm and the fluence smoothing level was at medium level.
Plan comparison
Evaluation parameters
The comparison of IMRT and VMAT plans was evaluated using the following terms:
- Homogeneity index (HI): ( D 2% -D 98% )/D 50% , a ratio evaluating the dose homogeneity in PTV where D 2% , D 98% , and D 50% are the minimum dose delivered to 2,98, and 50% volume of the PTV, respectively, [7]. HI of zero indicates the dose distribution is homogeneous
- Conformity index (CI): Vpres/PTVp, a ratio evaluating the coverage of the prescription dose in treatment plans, where Vpres was the volume of body receiving the prescribed dose and PTVp was the volume of PTV receiving the prescription dose. [8] CI of one indicates the good dose conformity
- Target volumes: D 98% , D 95% , and D 50% for PTV 70 , PTV 63 , and PTV 56 were analyzed, where D 98% , D 95% , and D 50% were minimum dose delivered to 98, 95, and 50% volume of PTV, respectively. A total of 110% (77 Gy) of dose to target volume was for most of the patients were zero. So Dmax dose to PTV was analyzed
- Critical structures: Dmax and D 1cc for spinal cord, PRV cord, and brainstem were analyzed, where Dmax was the maximum dose to the normal structure and D 1cc maximum dose received by the 1-cc volume of the normal structure. Parotid glands were analyzed with mean dose to whole glands and D 50% of the glands, i.e., minimum dose delivered to 50% of parotid glands
- Normal tissues: The volume of normal tissue receiving ≥2, ≥5, and ≥10 Gy were analyzed
- Determination of effective delivery time: MU/fraction and delivery time for each plan were compared. The delivery time for all test plans (IMRT and VMAT) was recorded by the same technologist. In this study, the patient setup time was excluded for both the plans. The delivery time for IMRT included the beam data transfer time from MOSAIQ to Precise Desktop, beam on time, and gantry rotation time. Gantry was rotated through ASU for each angle. For VMAT, the beam data transfer time and beam on time was taken into consideration.
- Statistical analyses: Statistical tests of the significance to calculate the differences between IMRT and VMAT plans were done using Wilcoxon matched-pair signed-rank test (two tailed, P ≤ 0.05).
Results | |  |
For all 13 patients, both IMRT and VMAT plans were accepted after fulfilling the planning objective. The plans were evaluated by the same oncologist. The isodose distribution was evaluated with DVH. [Figure 1] shows the comparative isofill distribution of IMRT and VMAT. The results of the analysis of DVH were given in [Table 2]. | Figure 1: Isofill comparison of IMRT (left side) and VMAT (right side) for one base of tongue case. Red color: 68 Gy; yellow color: 63 Gy; cyan color: 56 Gy
Click here to view |
Target volumes
We were able to achieve the good dose coverage in both IMRT and VMAT plans. Dose to all the target volumes were comparable. VMAT gave very minimal and statistically insignificant improvement in HI {VMAT (0.0979 ± 0.012); IMRT (0.1035 ± 0.019)} and CI {VMAT (1.074 ± 0.26); IMRT (1.12 ± 0.24)}. The results were given in [Table 2]. The DVH comparison of PTVs was shown in [Figure 2]. | Figure 2: DVH comparison of IMRT and VMAT of PTVs. Solid lines: VMAT plan; dashed line: IMRT plan
Click here to view |
Organ at risks
Spinal cord: The planning objective was met in both IMRT and VMAT plans. No statistical significant differences were observed in Dmax {VMAT (41.9 ± 1.28); IMRT (42.3 ± 1.07)} Gy as well as for 1 cc of spinal cord {VMAT (38.8 ± 1.65); IMRT (39.8 ± 1.05)} Gy.
PRV cord
Both plans met the planning objective. Dmax values were almost similar in VMAT (51.7 ± 1.85) Gy and IMRT (51.3 ± 1.91) Gy. The dose received by 1 cc of spinal cord was higher in VMAT (47.3 ± 1.21) Gy than in IMRT (46.8 ± 1.02) Gy, which showed significant difference statistically.
Brain stem
Planning objective for brain stem was easily met in both the plans. Dmax and dose to 1-cc volume of brain stem were analyzed. Dmax was lower in VMAT (48.5 ± 2) Gy compared with IMRT (49.4 ± 1.6) Gy and also 1 cc of brain stem {VMAT (43.7 ± 2); IMRT (44.3 ± 2.2)} Gy. Dmax as well as 1cc of brainstem did not show any significant statistic differences.
Parotid glands
In the case of both parotid glands (right (Rt) and left (Lt)), the planning objective was not fulfilled. When we tried to fulfill the planning objective of parotid gland, we could not achieve the objective of target volumes. Hence, the dose to the parotid glands was compromised. The mean dose received in VMAT {Rt (32.5 ± 2.61); Lt (32.8 ± 2.63)} was less than IMRT {Rt (35.2 ± 3.10); Lt (35.8 ± 7.07)}. Also the D 50% dose received in VMAT plan {Rt (30.3 ± 2.94); Lt (30.3 ± 3.30)} was less than with that of IMRT plans {Rt (35.0 ± 3.80); Lt (35 ± 7.97)}.
The dose differences and the P values are tabulated in [Table 1]. The DVH analyses of the OARs were shown in [Figure 3] and [Figure 4]. | Figure 3: DVH comparison of spinal cord, brain stem, and PRV cord. Solid lines: VMAT; dashed lines: IMRT; blue color: Spinal cord; brown color: PRV cord; orange color: Brainstem
Click here to view |
 | Figure 4: DVH comparison of parotid glands. Solid lines: VMAT; dashed lines: IMRT; blue color: left parotid; green color: right parotid. DVH: Dose volume histogram, IMRT: Intensity-modulated radiation therapy, VMAT: Volumetric-modulated arc therapy
Click here to view |
Normal tissue dose
The volume of normal tissue received ≥2 Gy was less in VMAT than IMRT. The volume of normal tissue received ≥5 Gy and ≥10 Gy was similar in both VMAT and IMRT.
Delivery efficiency
The MUs were compared. The average MU for VMAT (804 ± 96) was higher than that of IMRT plans (700 ± 52). The delivery time was compared between IMRT and VMAT. The plan delivery for IMRT was double (10 min:18 s) than that of VMAT delivery time (5 min:06 s).
Discussion | |  |
This study compared VMAT with dynamic MLC IMRT for 13 patients of base of tongue cases. The main issue in the treatment of head and neck cases is the concave shape of the tumor and the proximity of normal structures to the tumor. IMRT has been the standard treatment techniques for head and neck cancer. It gives the advantage of multiple prescriptions, normal tissue sparing, and good dose conformity. VMAT is the technique in which IMRT plans were delivered with reduced delivery time. [3],[4],[9] Many papers have been published using Eclipse or pinnacle TPS comparing VMAT and IMRT plans. However, comparison studies of VMAT and IMRT using Monaco planning system are very few compared with the above-mentioned system. This paper compares IMRT and VMAT in Monaco planning system.
Dose coverage in both the plans was similar. The HI and CI did not show any statistical differences. Spinal cord dose and brain stem dose were similar in both VMAT and IMRT plans. Volume 1 cc of PRV cord was slightly higher in VMAT compared with IMRT. However, the doses were within the acceptable limit. Both the parotid glands had better sparing in VMAT than IMRT. The volume of normal tissue dose received in VMAT is less than IMRT, which will reduce late effect of low dose radiation (radiation-induced carcinomas).
The number of MUs required to deliver a prescribed dose was higher in VMAT (both single and dual arc) than IMRT. It differed from other previous studies that showed a reduction in the MUs of VMAT plans. [1],[5],[7] In IMRT plan, low dose rate was selected to achieve the good quality of plan that gives the higher MU efficiency. Also, Elekta linear accelerator works on variable dose rate in dynamic MLC mode. However, the treatment delivery time was reduced to half for VMAT compared with IMRT. This definitely gives the advantage of treating more patients in the stipulated time in the busy department.
Conclusion | |  |
This study compared dynamic MLC IMRT and VMAT plans for 13 base of tongue cases with the same target volumes and OARs. In both the plans, we were able to achieve good target coverage and normal tissue sparing. The main difference between VMAT and IMRT was the delivery time. Owing to reduced delivery time and improved efficiency, the machine throughput may be increased. We conclude in this study that VMAT gives good quality of the plan as IMRT with less delivery time and improved efficiency for base of tongue cases.
References | |  |
1. | Verbakel WF, Cuijpers JP, Hoffmans D, Bieker M, Slotman BJ, Senan S. Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: A comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys 2009;74:252-9.  |
2. | Anand AK, Jain J, Negi PS, Chaudhoory AR, Sinha SN, Choudhury PS, et al. Can dose reduction to one parotid gland prevent xerostomia?-A feasibility study for locally advanced head and neck cancer patients treated with intensity-modulated radiotherapy. Clin Oncol (R Coll Radiol) 2006;18:497-504.  |
3. | Lee TF, Chao PJ, Ting HM, Lo SH, Wang YW, Tuan CC, et al. Comparative analysis of SmartArc-based dual arc volumetric-modulated arc radiotherapy (VMAT) versus intensity-modulated radiotherapy (IMRT) for nasopharyngeal carcinoma. J Appl Clin Med Phys 2011;12:3587.  |
4. | Holt A, Van Gestel D, Arends MP, Korevaar EW, Schuring D, Kunze-Busch MC, et al. Multi-institutional comparison of volumetric modulated arc therapy vs. intensity-modulated radiation therapy for head-and-neck cancer: A planning study. Radiat Oncol 2013;8:26.  |
5. | Vanetti E, Clivio A, Nicolini G, Fogliata A, Ghosh-Laskar S, Agarwal JP, et al. Volumetric modulated arc radiotherapy for carcinomas of the oro-pharynx, hypo-pharynx and larynx: A treatment planning comparison with fixed field IMRT. Radiother Oncol 2009;92:111-7.  |
6. | Krieger T, Sauer OA. Monte Carlo-versus pencil-beam-/collapsed-cone-dose calculation in a heterogeneous multi-layer phantom. Phys Med Biol 2005;50:859-68.  |
7. | ICRU Report 83. Prescribing, recording, and reporting photon-beam intensity- modulated radiation therapy (IMRT). International Commission on Radiation Units and Measurements, Bethesda, 2010.  |
8. | Pathak P, Vashisht S. A quantitative analysis of intensity-modulated radiation therapy plans and comparison of homogeneity indices for the treatment of gynecological cancers. J Med Phys 2013;38:67-73.  [PUBMED] |
9. | Sankaralingam M, Glegg M, Smith S, James A, Rizwanullah M. Quantitative comparison of volumetric modulated arc therapy and intensity modulated radiotherapy plan quality in sino-nasal cancer. J Med Phys 2012;37:8-13.  [PUBMED] |
[Figure 1], [Figure 2], [Figure 3], [Figure 4]
[Table 1], [Table 2]
This article has been cited by | 1 |
Dosimetric Evaluation of Low-Dose Spillage Volumes for Head and Neck Cancer Using Intensity-Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Treatment Techniques |
|
| Gourav Kumar, Manindra Bhushan, Lalit Kumar, Vimal Kishore, Kothanda Raman, Pawan Kumar, Soumitra Barik, Sandeep Purohit | | Progress in Medical Physics. 2021; 32(3): 70 | | [Pubmed] | [DOI] | | 2 |
Dosimetric comparison of coplanar and non-coplanar volumetric-modulated arc therapy in head and neck cancer treated with radiotherapy |
|
| Sanjib Gayen, Sri Harsha Kombathula, Sumanta Manna, Sonal Varshney, Puneet Pareek | | Radiation Oncology Journal. 2020; 38(2): 138 | | [Pubmed] | [DOI] | | 3 |
The effect of rectal gas on dose distribution during prostate cancer treatment using full arc and partial arc Volumetric Modulated Arc Therapy (VMAT) treatment plans |
|
| Motoharu Sasaki, Hitoshi Ikushima, Akira Tsuzuki, Wataru Sugimoto | | Reports of Practical Oncology & Radiotherapy. 2020; 25(6): 974 | | [Pubmed] | [DOI] | | 4 |
3D conformal, IMRT and VMAT for the treatment of head and neck cancer: a brief literature review |
|
| Kazi T. Afrin, Salahuddin Ahmad | | Journal of Radiotherapy in Practice. 2020; : 1 | | [Pubmed] | [DOI] | |
|
 |
|
|
|
|