J Med Phys Close
 

Figure 2: Differences between the fluence reconstructed from the treated and planned positions for two treatment fields before (left) and after correction with 2D array measurements (right). The differences remain after correction. Colors represent the % local dose difference. The scales are in pixels

Figure 2: Differences between the fluence reconstructed from the treated and planned positions for two treatment fields before (left) and after correction with 2D array measurements (right). The differences remain after correction. Colors represent the % local dose difference. The scales are in pixels