ORIGINAL ARTICLE
Year : 2019  |  Volume : 44  |  Issue : 4  |  Page : 254-262

Dose shadowing and prosthesis involvement for megavoltage photon In vivo diode dosimetry


Department of Medical Physics, University of the Free State, Bloemfontein, South Africa

Correspondence Address:
Dr. Nicholas Ade
Department of Medical Physics, University of the Free State, PO Box 339, Bloemfontein 9300
South Africa
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_59_19

Rights and Permissions

Aim: The aim of the study is to investigate the photon beam perturbations induced by an in vivo diode in combination with prosthesis involvement in a human-like phantom. Materials and Methods: Beam perturbations for 6 MV and 10 MV photons caused by an EDP-203G in vivo diode in combination with prosthesis involvement were studied in a unique water-equivalent pelvic phantom, equipped with bony structures and Ti prosthesis using single fields between 2 × 2 and 15 cm × 15 cm as well as 10 MV lateral opposing fields and a four-field plan. Dose distributions were measured with Gafchromic EBT3 films with and without the diode included in the beams on the prosthesis (prosthetic fields) and non-prosthesis (non-prosthetic fields) sides of the phantom. Differences between prosthetic and non-prosthetic field dose data were determined to assess the effect of the prosthesis on the diode-induced beam perturbations inside the phantom. Results: Photon beam dose perturbations ranged from 2% to 7% and from 5% to 12% for prosthetic and non-prosthetic fields, respectively, with relative differences between 2% and 4%. In addition, d50depths ranging from 8.7 to 11.5 cm and from 11.5 to 15 cm were acquired in the phantom for prosthetic and non-prosthetic fields, respectively, with relative differences between 2% and 5%. Conclusion: On the basis of accuracy requirements in radiotherapy noting that a small underdose to tumors could yield a decrease in the probability of tumor control, the diode-induced beam perturbations in combination with prosthesis involvement in the photon fields may affect treatment outcome, as there would be a reduction in the prescribed target dose during treatment delivery.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed188    
    Printed2    
    Emailed0    
    PDF Downloaded54    
    Comments [Add]    

Recommend this journal