Journal of Medical Physics
 Home | Search | Ahead of print | Current Issue | Archives | Instructions | Subscription | Login  The official journal of Association of Medical Physicists of India      
 Users online: 246  Home  EMail this page Print this page Decrease font size Default font size Increase font size 
ORIGINAL ARTICLE
Year : 2019  |  Volume : 44  |  Issue : 4  |  Page : 239-245

Estimation of backscatter from internal shielding in electron beam radiotherapy using Monte Carlo simulations (EGSnrc) and Gafchromic film measurements


1 Radiological Physics and Internal Dosimetry Group, Institute of Nuclear Medicine and Allied Sciences; Department of Radiotherapy, Army Hospital(Research and Referral), New Delhi, India
2 Department of Radiotherapy, Army Hospital(Research and Referral), New Delhi, India
3 Batra Hospital and Medical Research Center, New Delhi, India

Correspondence Address:
Dr. Sukhvir Singh
Department of Radiotherapy, Army Hospital (Research and Referral), Delhi Cantt, New Delhi - 110 010
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_21_19

Rights and Permissions

Purpose: The purpose of the study was to estimate the backscatter electron dose in internal shielding during electron beam therapy using Monte Carlo (MC) simulations and Gafchromic film measurements. Materials and Methods: About 6 and 9 MeV electron beams from a Varian 2100C linac were simulated using BEAMnrc MC code. Various clinical situations of internal shielding were simulated by modeling water phantoms with 2 mm lead sheets placed at different depths. Electron backscatter factors (EBF), a ratio of dose at tissue-shielding interface to the dose at the same point without the shielding, were estimated. The role of 2 mm aluminum in reduction of backscatter was investigated. The measurements were also performed using Gafchromic films and results were compared with MC simulations. Results: For particular beam energy, the EBF value initially increased with depth in the buildup region and then decreased rapidly. The highest value of EBF for both the energies is nearly same though at different depths. Decreased EBF was observed for 9 MeV beam in comparison to the 6 MeV beam for the same depth of shielding placement. Two millimeter aluminum reduced the backscatter by nearly 25% at maximum backscatter condition for both the energies, though the effectiveness slightly decreased at higher energy. The range of backscatter electrons was varying from 5 to 12 mm in the upstream direction from the interface. The Gafchromic film-measured EBF and MC-simulated EBF were matching well within the clinically acceptable limits except in close vicinity of tissue-lead interface. Conclusions: This study provides an important clinical data to design internal shielding at the local clinical setup and confirms applicability of MC simulations in backscatter dose calculations at interfaces where physical measurements are difficult to perform.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed351    
    Printed7    
    Emailed0    
    PDF Downloaded67    
    Comments [Add]    

Recommend this journal