Clinical evaluation of a two-dimensional liquid-filled ion chamber detector array for verification of high modulation small fields in radiotherapy


1 Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
2 Department of Radiation Oncology, University of Arkansas for Medical Sciences, Little Rock, AR, USA
3 Department of Radiation Oncology, University of North Carolina, Chapel Hill, NC, USA

Correspondence Address:
Dr. Miljenko Markovic
Department of Radiation Oncology, University of Texas Health San Antonio, 7979 Wurzbach Rd, San Antonio, TX 78229
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_4_19

Rights and Permissions

Introduction: Clinical evaluation of a two-dimensional (2D) liquid-filled ion chamber detector array used in the verification of highly modulated small beams of stereotactic body radiation therapy (SBRT) has been conducted. Materials and Methods: Measurements with the Octavius 1000 SRS (PTW, Freiburg, Germany) detector with 977 liquid-filled ion chambers were compared against EDR2 film and PTW Octavius Seven29. The performance of detector array has been evaluated on ten SBRT patient plans. Dose profiles of individual and composite fields' calculated using Pinnacle3 treatment planning system were compared against measurements with Octavius 1000 SRS detector array, EDR2 film, and Octavius Seven29 detector. Gamma index and profile comparison were used in the evaluation and assessment of the detector's performance. Results: The Gamma index measurements show agreement between Pinnacle3 computations and Octavius 1000 SRS array, PTW Octavius Seven29, and EDR2 film for >90% of the points using 2%, 2 mm tolerance criteria. Profiles obtained with the Octavius 1000 SRS were in agreement with the EDR2 film profiles, demonstrating the detector's superior sampling rate. Conclusions: The Octavius 1000 SRS is a dosimetrically accurate device to perform quality assurance checks in SBRT treatments. The broad range of measurements performed in this study quantified the dosimetric accuracy of Octavius 1000 SRS detector in the clinical setup of the small fields in radiotherapy.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed411    
    Printed28    
    Emailed0    
    PDF Downloaded46    
    Comments [Add]    

Recommend this journal