ORIGINAL ARTICLE
Year : 2018  |  Volume : 43  |  Issue : 4  |  Page : 247-254

Dosimetric characterization of an intensity-modulated X-ray brachytherapy system


1 Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland, USA
2 Department of NBICS Science and Technologies, Kurchatov Institute, Moscow, Russia
3 Institute for Diagnostic Imaging Research, University of Windsor, Windsor, Ontario, Canada
4 Department of Radiation Oncology, Duke University, Durham, North Carolina, USA

Correspondence Address:
Dr. Sung-Woo Lee
Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_52_18

Rights and Permissions

Purpose: An intensity-modulated X-ray brachytherapy system is being developed for various clinical applications. This new system makes it possible for clinical staff to control energy as well as dose rate for different tumor sites according to their sizes and radiobiological characteristics. Materials and Methods: This system is mainly composed of an X-ray tube, guide tube collimation, and secondary (pseudo) target. Due to its configuration, convenient modulations of fluorescent X-ray energy and intensity are possible. To observe applicability of this novel system for various primary and secondary target combinations, Monte Carlo simulation using MCNP5 was performed, and air measurements were done. As a primary and pseudo-target combination, silver–molybdenum (Ag-Mo), tungsten–neodymium (W-Nd), and tungsten–erbium (W-Er) were used for the calculation for dose profile. Specifically, a dose distribution was calculated around each of these target combinations. Dose distributions as a function of target angles were also calculated. The Ag-Mo combination was analyzed for Cartesian coordinates of xy, xz, and yz planes of the pseudo-target to observe dose distribution as a function of the angle of secondary target. Results: The results showed that radial dose fall-off of Ag-Mo was greater than commercially available brachytherapy sources (103Pd and 125I) due to its low characteristic X-ray energy. Conclusions: Dose distribution variance should be considered in beam modulation for clinical application. Dynamic movement of the pseudo-target is feasible and remains as a subject for future research.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed94    
    Printed1    
    Emailed0    
    PDF Downloaded28    
    Comments [Add]    

Recommend this journal