ORIGINAL ARTICLE
Year : 2018  |  Volume : 43  |  Issue : 4  |  Page : 230-235

Comparison of interfractional setup reproducibility between two types of patient immobilization devices in image-guided radiation therapy for prostate cancer


1 Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
2 Department of Radiation Oncology, Osaka International Cancer Institute, Osaka; Department of Radiation Oncology, Osaka University Graduate School of Medicine, Suita, Japan
3 Department of Radiation Oncology, Osaka International Cancer Institute, Osaka; Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan

Correspondence Address:
Mr. Yoshihiro Ueda
Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69, Otemae, Chuo-ku, Osaka 537-8511
Japan
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_20_18

Rights and Permissions

Purpose: The aim of this study is to compare the interfractional setup reproducibility of two types of patient immobilization devices for prostate cancer receiving image-guided radiation therapy (IGRT). Materials and Methods: The MOLDCARE (MC) involves hydraulic fixation, whereas the BlueBAG (BB) and Vac-Lock (VL) involve vacuum fixation. For 72 patients, each immobilization device was individually customized during computed tomography (CT) simulation. Before the treatment, bony registration was performed using orthogonal kV images and digitally reconstructed radiographs. The shift of the treatment couch was recorded as a benchmark in the first session. In subsequent sessions, the shifts from the benchmark were measured and analyzed. Soft-tissue registration was performed weekly by cone-beam CT and CT images, and the shifts were measured and analyzed. Results: In the superior-inferior and left-right directions, there were nearly no changes in the overall mean among the immobilization devices. In the anterior-posterior (AP) direction, the overall mean for the MC, BB, and VL were 0.34 ± 1.33, −0.47 ± 1.27, and −1.82 ± 1.65 mm, respectively. The mean shifts along the AP direction were approximately 1 mm more in patients immobilized on the BB and 2.5 mm more in those on the VL, compared to those on the MC, after the twentieth treatment. No significant changes were observed among the patients immobilized on those devices, respectively, in soft-tissue registration. Conclusion: It can be concluded that the settling of the vacuum fixation was caused by air leakage in the latter-half treatment, and the immobilization device type has no effect on the treatment-position reproducibility in IGRT.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed119    
    Printed2    
    Emailed0    
    PDF Downloaded22    
    Comments [Add]    

Recommend this journal