TECHNICAL NOTE
Year : 2017  |  Volume : 42  |  Issue : 3  |  Page : 140-143
 

Choice of a suitable dosimeter for photon percentage depth dose measurements in flattening filter-free beams


1 Beatson West of Scotland Cancer Centre, Glasgow, Scotland, UK
2 Edinburgh Cancer Centre, Edinburgh, Scotland, UK

Date of Submission13-Jan-2017
Date of Decision24-May-2017
Date of Acceptance14-Jun-2017
Date of Web Publication13-Sep-2017

Correspondence Address:
Francisco Cutanda Henríquez
Edinburgh Cancer Centre, Crewe Road South, EH4 2XU Edinburgh, Scotland
UK
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jmp.JMP_11_17

Rights and Permissions

 

   Abstract 


The International Atomic Energy Agency Technical Reports Series-398 code of practice for dosimetry recommends measuring photon percentage depth dose (PDD) curves with parallel-plate chambers. This code of practice was published before flattening filter-free (FFF) beams were widely used in clinical linear accelerators. The choice of detector for PDD measurements needs to be reassessed for FFF beams given the physical differences between FFF beams and flattened ones. The present study compares PDD curves for FFF beams of nominal energies 6 MV, 6 FFF, 10 MV, and 10 FFF from a Varian TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, USA) acquired with Scanditronix photon diodes, two scanning type chambers (both PTW 31010 Semiflex), two small volume chambers (Wellhofer CC04 and PTW 31016 PinPoint 3D), PTW 34001 Roos, Scanditronix Roos, and NACP 02 parallel-plate chambers. Results show that parallel-plate ion chambers can be used for photon PDD measurements, although for better accuracy, recombination effects should be taken into account.


Keywords: International Atomic Energy Agency Technical Reports Series-398, percentage depth dose, photon dosimetry, parallel-plate ion chamber


How to cite this article:
Vargas Castrillon S, Cutanda Henríquez F. Choice of a suitable dosimeter for photon percentage depth dose measurements in flattening filter-free beams. J Med Phys 2017;42:140-3

How to cite this URL:
Vargas Castrillon S, Cutanda Henríquez F. Choice of a suitable dosimeter for photon percentage depth dose measurements in flattening filter-free beams. J Med Phys [serial online] 2017 [cited 2019 Sep 20];42:140-3. Available from: http://www.jmp.org.in/text.asp?2017/42/3/140/214485





   Introduction Top


Linear accelerators with flattening filter-free (FFF) beams are becoming widespread, with Varian TrueBeam, and Elekta Versa being the most common ones since Siemens is no longer in the radiotherapy market. New radiotherapy techniques, such as volumetric modulated arc therapy and stereotactic body radiotherapy, allow the clinical use of unflattened beams. As it has been widely discussed in the literature, FFF linear accelerators have several potential advantages, including the fact that treatment times can be reduced as dose rates are much higher than the ones for flattened beams.[1],[2],[3]

The International Atomic Energy Agency (IAEA) published its Technical Reports Series-398 (TRS-398) code of practice “Absorbed Dose Determination in External Beam Radiotherapy” in 2000[4] when FFF linear accelerators were not yet in widespread clinical use. Nevertheless, this international code of practice is currently being utilized for FFF beams in different countries. The issue of whether or not corrections should be applied to the beam quality conversion factor, kq, to account for spectral differences of unflattened beams has been discussed in the literature, using both TPR20,10 or PDD(10)x as beam quality specifiers.[5] Some authors [6] recommend to subtract 0.5% from the beam quality conversion factor, kq, when using TPR20,10 as quality index but do not recommend any corrections when using PDD(10)x (0.4% being the maximum error). The National Physical Laboratory, which uses TPR20,10 as quality index, mentions in their 2014 calibration certificates that an additional small correction factor may be required but recommends to set it to unity while more data are collected, in the meantime increasing the expanded uncertainty from 1.4% to 1.6%. The IPEM topical report 1[3] introduces a correction factor of 0.997.

IAEA TRS-398 recommends the use of parallel-plate chambers to measure photon percentage depth doses (PDDs).[4] FFF linear accelerators produce radiation profiles which are peaked on the central axis; therefore, ion chambers used to measure central axis PDDs can experience partial volume averaging, which could lead to an underestimation of dose.[7] Since the dimensions of a parallel-plate chamber are larger than the dimensions of typical scanning chambers, a volume averaging effect (radial beam nonuniformity correction) could have a bigger influence on parallel-plate chambers than on scanning chambers; also, recombination and polarity effects could have an influence. An investigation into these effects is necessary if recommendations to measure photon PDDs with parallel-plate chambers are to be followed for FFF beams.

AAPM TG-51 does not make any specific recommendations regarding the use of parallel-plate chambers for PDD acquisition, but since PDD(10)x is used as the quality index, the accurate determination of this value is of great importance. AAPM TG-51 addendum explicitly mentions the advantages of using a parallel-plate chamber to measure PDDs.

The aim of this paper is to assess the use of parallel-plate ion chamber for photon PDD measurements in FFF beams, in comparison with other suitable dosimeters.


   Methods Top


Photon PDD curves for FFF beams of nominal energies 6 MV (6 FFF) and 10 MV (10 FFF) (Varian TrueBeam) were acquired with a Scanditronix photon diode, two scanning type chambers (both PTW 31010 Semiflex), two small volume chambers (Wellhofer CC04 and PTW 31016 PinPoint 3D), PTW 34001 Roos, Scanditronix Roos, and an NACP 02 parallel-plate chamber. PTW MP3 and PTW MP3-M water tanks, a PTW TANDEM electrometer, and PTW MEPHYSTO were used for all these measurements. IBA Blue Phantom 2 was also used for measurements with a second PTW Roos chamber and Scanditronix photon diodes. PDDs were measured at 100 cm source-to-surface distance (SSD), for fields of size: 5 cm × 5 cm, 10 cm × 10 cm, 20 cm × 20 cm, 30 cm × 30 cm, and 40 cm × 40 cm. PDDs were measured at all available dose rates for 6 FFF and 10 FFF as PDD curves smoothing vary with dose rate. A reference chamber (PTW Semiflex) was used for all scans to account for linear accelerator output variations, but for the diodes, where an appropriate diode was used instead.

Since there are conflicting findings in the literature about diode behavior with dose rate and energy dependence, diode measurements must be compared with ion chamber measurements. AAPM TG-106[8] does not recommend the use of diodes in large X-ray fields unless specific compensation or corrections with validated test result indicate; otherwise, our measurements in FFF beams show this diode behavior. There are also conflicting recommendations regarding PTW PinPoint 3D; according to a PTW technical note (D165.200.01/00), this chamber can be used up to a 30 cm × 30 cm field. Therefore, we restricted the use of these detectors to fields smaller or equal to 30 cm × 30 cm.

Dose-per-pulse in FFF beams is higher than for flattened ones, and this leads to higher values for ion recombination factors.[3],[9],[10],[11],[12] Since dose-per-pulse in Varian TrueBeam does not change with nominal dose rate, recombination factors do not change either. The two voltage technique was used to obtain these factors [7],[13],[14],[15],[16] for all field sizes, ion chambers, and polarities at six different depths, dmax, 50, 100, 150, 200, and 300 mm deep, with a PTW TANDEM electrometer and IBA CCU electrometer used for PDD scanning. All PDDs were measured for two different polarities as supplied by PTW TANDEM and IBA CCU electrometers. Recombination factors were also measured with a PTW UNIDOS E electrometer at the same depths and for the same field sizes, and yielded the same values within uncertainty estimates.

A decision was made to measure PDDs using two polarities as Sarkar et al.[13] reported significant differences in PDDs measured using very small ion chambers, which became larger with increasing depths. These discrepancies were found to be bigger for the smallest volume chamber, Exradin A16, but it was also significant for Wellhofer CC04, which is one of the chambers used in our study. A PTW PinPoint 3D ion chamber is also used in our study, with volume 0.016 cm,3 which is not as small as the Exradin A16, but it is still smaller than Wellhofer CC04, making the assessment of this chamber relevant. Exradin A16 and PTW PinPoint 3D differ greatly in one relevant feature of their design, PTW PinPoint chamber is built with an aluminum electrode, while Exradin A16 has a wire electrode. Due to this fact, Exradin A16 presents an overresponse for low-energy photons, effect which increases as the number of low-energy photons increases, e.g. as field size.

To correct the raw PDDs for recombination factors at different depths, the ion recombination measurements were fitted against depth using the least squares method. All PDDs were measured at all available dose rates, ranging from 400 MU/min to 1400 MU/min for 6 FFF and from 400 MU/min to 2400 MU/min for 10 FFF. Scans were found to be smoother at higher dose rates as it has been previously reported in the literature, but no reportable discrepancies were found for none of the parameters relevant for our study.


   Results Top


It was found that parallel-plate chambers showed the best PDD curves coincidence for both polarities, differences were below 0.1% for all depths [Figure 1], whereas cylindrical chambers could show differences up to 0.6% at 350 mm deep and a 30 cm × 30 cm field [Figure 2].
Figure 1: Percentage depth dose for a 10 FFF beam, measured with a PTW Roos parallel-plate ion chamber and both polarities. Field size 40 cm × 40 cm

Click here to view
Figure 2: Percentage depth dose for a 10 FFF beam, measured with a PinPoint 3D ion chamber and both polarities. Field size 30 cm × 30 cm

Click here to view


PDDs for one single polarity and different ion chambers were corrected for recombination and compared. The largest difference in PDD among different ion chambers (excluding PinPoint which was used up to 30 cm × 30 cm) was found at 350 mm deep for all field sizes, which would amount to: 0.6% for 6 FFF and 0.5% for 10 FFF for a 40 cm × 40 cm field between CC04 and PTW Roos. Differences increase with field size [Figure 3].
Figure 3: Percentage depth doses for a 10 × 10 field size, measured with a Photon Field Diode (PFD), 3D PinPoint ion chamber, Semiflex ion chamber, CC04 ion chamber, and PTW Roos parallel-plate ion chamber. (a) 6 FFF, (b) 10 FFF

Click here to view


PDDs measured with photon diodes led to the highest difference compared with ion chamber measurements, for both energies and for all field sizes, reaching 4% difference with PTW Semiflex chamber for 6 FFF, 40 cm × 40 cm and 350 mm deep, diodes should be compared with ion chambers in any case. Our results show that differences increase with depth and increasing field size, what could be related to the low energy overresponse of Scanditronix diodes and PTW PinPoint 3D ion chamber. The overresponse of diodes at low energies has been extensively reported in the literature.[8] Our results are consistent with these previous findings. Discrepancies are slightly higher than the ones found for flattened beams as the spectrum is softer for FFF beams.

PDDs measured with ion chambers have a very small dependence on dose rate, unlike diodes for which differences up to 1.5% can be found at the largest field size for 6 FFF.

For absolute dosimetry purposes, IAEA TRS-398 reference conditions for the determination of absorbed dose to water are SSD = 100 cm, 10 cm × 10 cm field at 10 cm deep.[4] Hence, the accurate measurement of the value of PDD(10) is relevant. Our results show that the maximum difference in PDD(10) measured with all different ion chambers, regardless of the polarity, is 0.3% for 6 FFF measured between PTW PinPoint 3D chamber and PTW Semiflex for a 30 cm × 30 cm field. The highest difference between ion chamber PDD(10) and diode PDD(10) is 1.3%.

Once PDDs are corrected to take into account recombination effects, we found differences between corrected and uncorrected PDDs that ranged from 1.2% for PTW Semiflex ion chamber to 2.4% for PTW Roos ion chamber, 2.5% for Scanditronix Roos, all measured for a 10 FFF, 40 cm × 40 cm at 350 mm deep. The NACP ion chamber gave consistent intermediate recombination results for all field sizes. The question of whether or not all PDDs must be corrected for recombination effects is something medical physicists should judge taking into account results for their detectors and beams. We checked two PTW Roos ion chambers and one Scanditronix Roos chamber, and in the case of these chambers, the percentage difference might be too high to be neglected. At 10 cm deep in a 10 cm × 10 cm field, the differences between PDD value for recombination corrected curves and noncorrected ones range from 0.5% for the Roos chamber to 0.2% for the PinPoint 3D chamber.

The differences between PDD(10) measured with diodes and ion chamber corrected for recombination effects PDD(10) are ~1.5% for all chambers for a 30 cm × 30 cm field and 6 FFF, whereas for 10 FFF, differences are ~2%.


   Conclusions Top


Parallel-plate chambers can be used for PDD measurements in Varian TrueBeam FFF beams, provided field sizes are large enough (at least 5 cm × 5 cm for a Roos-type ion chamber) to allow the use of parallel-plate chambers. They can be used in the case of a 10 cm × 10 cm field so that PDD(10) can be measured, and hence, absolute dose at dmax obtained for absorbed dose to water determination according to the TRS-398 code of practice. TRS-398 recommendations for PDD measurements can then be followed, and the determination of PDD(10)x according to TG-51 can also be carried out using this type of chamber. Parallel-plate chambers show an interesting and important property, which is a very small polarity dependence, and the smallest of all chambers included in our study. In addition, as mentioned in AAPM TG-51 addendum, they have a well-defined effective point of measurement, and they show good agreement with Monte Carlo calculations, especially in the region of dose buildup, which makes them good detectors for PDD measurements. There is one caveat though, regarding the use of parallel-plate ion chambers, i.e., correction for ion chamber recombination effects, as this type of chambers present the highest recombination effects dependence.

When obtaining recombination factors with values over 1% and noticeable variation with depth, such as the ones corresponding to PTW and Scanditronix Roos ion chambers for field sizes over 20 cm × 20 cm, the use of recombination corrected PDDs might be advisable, to prevent systematic errors that could be as high as 2.5% at 350 mm deep for large field sizes. In the case of Semiflex chambers, the difference between the uncorrected and the corrected PDD is not so high; clinical judgment should be used to decide whether or not corrections must be applied with this particular chamber or with any other ion chamber included in our study.

Acknowledgments

The authors are grateful to Dan Welsh for his careful revision of the manuscript.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Xiao Y, Kry SF, Popple R, Yorke E, Papanikolaou N, Stathakis S, et al. Flattening filter-free accelerators: A report from the AAPM Therapy Emerging Technology Assessment Work Group. J Appl Clin Med Phys 2015;16:5219.  Back to cited text no. 1
    
2.
Vassiliev ON, Titt U, Pönisch F, Kry SF, Mohan R, Gillin MT. Dosimetric properties of photon beams from a flattening filter free clinical accelerator. Phys Med Biol 2006;51:1907-17.  Back to cited text no. 2
    
3.
Budgell G, Brown K, Cashmore J, Duane S, Frame J, Hardy M, et al. IPEM topical report 1: Guidance on implementing flattening filter free (FFF) radiotherapy. Phys Med Biol 2016;61:8360-94.  Back to cited text no. 3
    
4.
Andreo P, Burns DT, Hohlfeld K, Huq MS, Kanai T, Laitano F, et al. Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. Technical Report Series No. 398. Vienna, Austria: IAEA; 2000.  Back to cited text no. 4
    
5.
Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, et al. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26:1847-70.  Back to cited text no. 5
    
6.
Xiong G, Rogers DW. Relationship between %dd(10)x and stopping-power ratios for flattening filter free accelerators: A Monte Carlo study. Med Phys 2008;35:2104-9.  Back to cited text no. 6
    
7.
Kry SF, Popple R, Molineu A, Followill DS. Ion recombination correction factors (P(ion)) for Varian TrueBeam high-dose-rate therapy beams. J Appl Clin Med Phys 2012;13:3803.  Back to cited text no. 7
    
8.
Das IJ, Cheng CW, Watts RJ, Ahnesjö A, Gibbons J, Li XA, et al. Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM. Med Phys 2008;35:4186-215.  Back to cited text no. 8
    
9.
Corns RA, Huang VW, Thomas SD. Pion effects in flattening filter-free radiation beams. J Appl Clin Med Phys 2015;16:376-85.  Back to cited text no. 9
    
10.
Lang S, Hrbacek J, Leong A, Klöck S. Ion-recombination correction for different ionization chambers in high dose rate flattening-filter-free photon beams. Phys Med Biol 2012;57:2819-27.  Back to cited text no. 10
    
11.
Wang Y, Easterling SB, Ting JY. Ion recombination corrections of ionization chambers in flattening filter-free photon radiation. J Appl Clin Med Phys 2012;13:3758.  Back to cited text no. 11
    
12.
Johnsen S. Ion chamber collection efficiency considerations for un-flattened X-ray beams. Med Phys 2008;35:2770.  Back to cited text no. 12
    
13.
Sarkar V, Wang B, Zhao H, Lynch B, James JA, McCullough KT, et al. Percent depth-dose distribution discrepancies from very small volume ion chambers. J Appl Clin Med Phys 2015;16:5230.  Back to cited text no. 13
    
14.
Nisbet A, Thwaites DI. Polarity and ion recombination correction factors for ionization chambers employed in electron beam dosimetry. Phys Med Biol 1998;43:435-43.  Back to cited text no. 14
    
15.
Boag JW, Currant J. Current collection and ionic recombination in small cylindrical ionization chambers exposed to pulsed radiation. Br J Radiol 1980;53:471-8.  Back to cited text no. 15
    
16.
Weinhous MS, Meli JA. Determining Pion, the correction factor for recombination losses in an ionization chamber. Med Phys 1984;11:846-9.  Back to cited text no. 16
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (578 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
   Methods
   Results
   Conclusions
    References
    Article Figures

 Article Access Statistics
    Viewed1067    
    Printed24    
    Emailed0    
    PDF Downloaded73    
    Comments [Add]    

Recommend this journal