ORIGINAL ARTICLE
Year : 2016  |  Volume : 41  |  Issue : 4  |  Page : 224-233
 

On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator


1 Department of Radiotherapy and Oncology, Kasturba Medical College Hospital (An Associated Teaching Hospital of Manipal University), Mangalore, Karnataka, India
2 Department of Radiation Oncology, Manipal Hospital, Bengaluru, Karnataka, India
3 Department of Medical Radiation Physics, School of Allied Health Sciences, Manipal University, Manipal, Karnataka, India
4 Department of Radiological Physics, SMS Medical College, Jaipur, Rajasthan, India

Date of Submission21-Jun-2016
Date of Decision19-Oct-2016
Date of Acceptance19-Oct-2016
Date of Web Publication5-Dec-2016

Correspondence Address:
Challapalli Srinivas
Department of Radiotherapy and Oncology, Kasturba Medical College Hospital (An Associated Teaching Hospital of Manipal University), Attavar, Mangalore - 575 001, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.195186

Rights and Permissions

 

   Abstract 

Transmission of radiation fluence through patient's body has a correlation to the planned target dose. A method to estimate the delivered dose to target volumes was standardized using a beam level 0.6 cc ionization chamber (IC) positioned at electronic portal imaging device (EPID) plane from the measured transit signal (S t ) in patients with cancer of uterine cervix treated with three-dimensional conformal radiotherapy (3DCRT). The IC with buildup cap was mounted on linear accelerator EPID frame with fixed source to chamber distance of 146.3 cm, using a locally fabricated mount. S t s were obtained for different water phantom thicknesses and radiation field sizes which were then used to generate a calibration table against calculated midplane doses at isocenter (D iso,TPS ), derived from the treatment planning system. A code was developed using MATLAB software which was used to estimate the in vivo dose at isocenter (D iso,Transit ) from the measured S t s. A locally fabricated pelvic phantom validated the estimations of D iso,Transit before implementing this method on actual patients. On-line dose estimations were made (3 times during treatment for each patient) in 24 patients. The D iso,Transit agreement with D iso,TPS in phantom was within 1.7% and the mean percentage deviation with standard deviation is −1.37% ±2.03% (n = 72) observed in patients. Estimated in vivo dose at isocenter with this method provides a good agreement with planned ones which can be implemented as part of quality assurance in pelvic sites treated with simple techniques, for example, 3DCRT where there is a need for documentation of planned dose delivery.


Keywords: Cervix cancer; clinical dosimetry; conformal radiotherapy; on-line dose verification


How to cite this article:
Putha SK, Saxena P U, Banerjee S, Srinivas C, Vadhiraja B M, Ravichandran R, Joan M, Pai K D. On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator. J Med Phys 2016;41:224-33

How to cite this URL:
Putha SK, Saxena P U, Banerjee S, Srinivas C, Vadhiraja B M, Ravichandran R, Joan M, Pai K D. On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator. J Med Phys [serial online] 2016 [cited 2019 Apr 22];41:224-33. Available from: http://www.jmp.org.in/text.asp?2016/41/4/224/195186



   Introduction Top


Outcome of radiation therapy treatment depends on the accuracy of dose delivery to the specified target volume. Uncertainty in dose delivery includes patient and machine specific parameters, for example, outline of patient structure, positioning error, variations in patient geometry from time of planning to the time of treatment, organ motion (internal and external), accuracy of treatment planning system (TPS) dose algorithm, and random variations in linear accelerator (linac) output. With the increase in complexity of treatment techniques such as three-dimensional conformal radiotherapy (3DCRT), intensity modulated radiotherapy, and volumetric modulated arc therapy there is a need to verify the planned dose delivery to the target site as part of verification process. Various national and international organizations recommend the need for in vivo dosimetry to be carried out on a regular basis as important part of quality assurance. [1],[2],[3],[4]

The methods for in vivo dosimetry involve physical placement of detectors inside the patient cavities such as mouth, esophagus, vagina and rectum. [5] Next closest approach is by placing dosimeters, such as diodes, thermoluminescent dosimeter (TLD), or metal oxide semiconductor field effect transistors (MOSFET), on the patient's skin or inside the patient to derive the dose at specific points within the patient. However, in clinical practice, placing detectors on the patient's skin at the entrance and/or exit surface of the beam is not always easy or feasible and requires some extra setup time in the treatment room. These detectors offer disadvantage owing to periodic calibrations, correction for temperature, pressure, photon energy dependence, accurate positioning, and estimation of the photon fluence perturbation inside the patient. This explains why "in vivo0" dosimetry remains limited to the first session of treatment and also why it is still not implemented in every radiotherapy center. [6] Another method, as part of in vivo dosimetry, transmission of radiation flux/signal measurement at a point far away from the patient was used in the days of deep X-rays and telecobalt beam treatments when rotation therapy was introduced in radiotherapy. [7] A simple and practical method was devised to obtain the in vivo dose in 60 pelvic treatments on the beam central axis using the signal measured by a small air ion chamber positioned along with beam central axis at typical distance of the electronic portal imaging device (EPID) from the source. [8] In another work, an estimation of in vivo dose from back projection of transit signal (S t ) measurement with EPID along the central axis of photon beam in patients undergoing conformal radiotherapy for various localizations was presented. [9] During the treatment session, a transit dose is measured with the EPID which is calibrated against an ionization chamber (IC) and the dose in the patient is estimated from the back projection of the portal dose and also validated for different beam energies. These methods can potentially be used to identify changes in tissue thickness and setup errors if any. The improvement of dose delivery accuracy and its efficacy in clinical radiotherapy is highlighted in another report. [10]

International Atomic Energy Agency (IAEA) report recommended implementation of in vivo dosimetry in sites with regular body contours such as the pelvis and for simple techniques not involving high dose gradients. [11] In this direction, a few reports describe in vivo dosimetry with chemical dosimeter and IC by placement of detectors in vaginal cavity of patients receiving external beam radiotherapy treatment for carcinoma uterine cervix. [12],[13] As the above method needs patient's consent and was time consuming, better methodology was found necessary. The present work illustrates our attempt in positioning an ion chamber, positioned on locally fabricated mount fixed at EPID level, along the beam central axis, and estimating the in vivo dose at isocenter in a group of carcinoma of uterine cervix patients treated with 3DCRT by accounting water equivalent path length principle.


   Materials and Methods Top


Treatment machine

A medical linear accelerator (model Elekta Compact) having a single 6 MV photon energy, motorized wedge, and 40 pairs Multi-Leaf Collimator (MLCi2) having leaf thickness of 1 cm at 100 cm iso-center available at our center was used. This is routinely operated at 350 monitoring units (MU) per minute, having output 1 centi-Gray (cGy)/MU at isocenter.

Estimation of in vivo dose thorough transit method

S t is measured at a faraway point along the central axis of the beam, at a distance of 46.3 cm from the isocenter of linac. A metallic mount made up of iron (ρ ≈ 7.85 g/cc) was fabricated locally and fixed at the level of EPID assembly. This carries a 0.65 cc IC (Model FC65-G from IBA Dosimetry, Germany) with build-up cap on an extended acrylic sheet (ρ ≈ 1.065 g/cc) of 8 mm thickness which facilitates no metallic interference in the path of beam along the central axis. [Figure 1] shows the schematic diagram of the transmission dose measurement setup showing the metallic mount fixed on the linac's EPID frame and the IC with its build-up cap placed on an acrylic sheet. The IC with buildup cap was inserted in an acrylic holder box so that a fixed target to chamber distance of 146.3 cm is maintained for all gantry angulations of linac with this geometry. [Figure 2] shows the metallic mount fixed to EPID frame of linac. Therefore, the chamber accurately measures S t using a Dose 1 electrometer (IBA Dosimetry, Germany) in nano Coulombs (nC), when media (water phantom or patient) is positioned on the linac treatment couch at the iso-center. The geometry of S t measurements is shown in [Figure 3]. S t s for 100 MU delivery were taken with a water phantom (dimensions of 40 cm ×40 cm ×40 cm) positioned on treatment couch in isocenteric condition for various water thicknesses (ranging from 6 to 30 cm with a step size of 2 cm) and for square fields (from 5 to 25 cm at intervals of 5 cm). These readings were corrected for temperature, pressure, phantom (base), and couch transmission [Annexure [Table 1].
Figure 1: Schematic diagram of the transmission dose measurement setup showing the metallic mount fixed on the linac's EPID frame and the IC with its build - up cap placed on an acrylic sheet

Click here to view


Figure 2: Metallic mount (with ionization chamber in transit position) fixed to electronic portal imaging device of linac

Click here to view
Figure 3: Schematic diagram representing the source to center of ionization chamber (with buildup cap) distance (1.463 m) along central axis kept on an acrylic sheet and water phantom kept at iso - center on treatment couch of linac

Click here to view




Computerized TPS (CMS XiO ® , Elekta Ltd, UK, version 4.80.02) which has Clarkson, convolution, superposition, and fast superposition algorithms, was commissioned with beam data measurements of the linac. Superposition algorithm was used for all dose calculations in this study. Quality assurance of TPS with regard to the calculated dose under different clinical situations was studied in a water phantom, and the measurements were found to be within ±1.5% of TPS calculated values. "Virtual water" phantoms (ρ = 1.0 g/cc) of dimensions 40 cm × 40 cm, with the above-selected thickness were generated in TPS. The midplane doses at isocenter in TPS (D iso,TPS ) in cGy for 100 MU delivery were recorded from the virtual water phantoms, for various water thicknesses, and field size combinations. A calibration table was generated by taking the ratio of S t versus D iso,TPS (in nC/cGy) for corresponding water thickness and field size combinations [Annexure [Table 2] and [Table 3].





A code was developed with MATLAB software (version R2015) to estimate the in vivo dose at isocenter (D iso,Transit ) through S t . This has a database of above measured S t values and calibration table. Input parameters such as equivalent square field size (Z), temperature, pressure, couch attenuation factor, delivered MUs, S t reading (in nC), and TPS calculated dose, were given through input window. Corrected transit reading (nC) was calculated from the entered MUs. Water equivalent depth (W eq ) was referred from the interpolated values of equivalent square field size (Z) and corrected S t by searching the analogous index. The estimated D iso,Transit value corresponding to Z and W eq was picked up from the nC/cGy ratio interpolated data.

Validation of MATLAB code with pelvic phantom

Before implementing this method on actual patients, a medium sized female pelvis phantom was used to verify the efficacy of the above-developed code for estimation of D iso,Transit through S t . This phantom was locally fabricated out of acrylic sheets with inbuilt bladder, rectum tissue structures in wax, femoral heads (Teflon material), and uterus cervix complex (target) that simulated actual patient. This has provision to hold a compact ion chamber (Model CC01 from IBA Dosimetry, Germany) in annular slots. [Figure 4] shows pelvic phantom in transverse view with CC01 IC slots. This simulates the treatment setup for carcinoma cervix with a nearly accurate anatomical geometry.
Figure 4: Pelvic phantom transverse view with CC01 ionization chamber slots

Click here to view


The phantom was scanned under computed tomography (CT) machine (Wipro GE, Model: High Speed) with CC01 in the target slot. Scanned CT images were exported to Focalsim contouring station (Elekta Ltd., Crawley, UK) through digital imaging and communications in medicine (DICOM) network. Contouring of target volume (position of CC01 chamber) was done in all transverse slices.

The contoured images were transferred to CMS XiO ® (Elekta Ltd, Crawley, UK) version 4.80.02 TPS for beam placement and dose calculations. A set of four beams with gantry angles 270°, 90° (with field size 15 cm ×12 cm), 0°, and 180° (with field size 12 cm ×12 cm), placing isocenter at the center of the target were placed. A dose of 200 cGy was prescribed to the target which was normalized to the 100% isodose line encompassing the target, and a box technique treatment plan was generated. The mean dose calculated by TPS around the region of the target was noted from the dose volume histogram.

The S t and the dose at isocenter were taken simultaneously with FC65 and CC01 chambers respectively for all four beams with pelvic phantom positioned under linac in an iso-centric condition during planning execution. Measured dose at isocenter with CC01 (D iso,phantom ) was compared with the D iso,Transit obtained from MATLAB code. To check the uniformity of absolute dose measured with CC01 and FC65 chambers, readings were taken separately following IAEA protocol TRS-398 in a water phantom, and a deviation of 0.2% was found from the measurements. [14]

Estimation of in vivo dose at isocenter through transit signal in patients

A total of 24 patients diagnosed with carcinoma uterine cervix who were treated by 3DCRT were included in this study. All patients were immobilized with "Vacloc" device (Klarity Medical, USA) in supine position. CT simulation was carried out for all patients for treatment planning, and the scanned serial images were exported to contouring station using DICOM network. Contouring of tumor volumes and normal structures was done by radiation oncologist. The created clinical tumor volume (CTV) encompassed the gross tumor volume together with nodal basins. The planning target volume (PTV) has a selected margin of 5 mm to CTV to account for inter-fractional and geometric positional uncertainties as per data from the investigating institution. Contoured image data set was transferred to TPS for beams placement. The isocenter was placed at the midplane of physical separation (both anterior-posteriorly and laterally) of patient's planning CT slice which has a central axis for 3DCRT planning in all patients where it falls inside the PTV. Therefore, it was ensured that the dose at the geometric center of the patient (which is same as isocenter) is identical to the homogeneous prescribed tumor dose (D iso,TPS ) in the 3DCRT planning slice. Four-field (270°, 0°, 90°, and 180°) box technique with all beams conformed to PTV by MLC with a margin of 5 mm was used for all patients. Field infield (subfield) and/or wedge technique were used, as and when required, to reduce the hot spots around target region. A 3DCRT plan was generated with a dose prescription of 2.0 Gray per fraction that was normalized to 100% isodose line covered to PTV. The plan was evaluated and approved by the radiation oncologist and was exported to record and verification (R and V) system (MOSAIQ ® ) for scheduling and execution.

The lateral and anterior digitally reconstructed radiograph images were exported to EPID (iViewC - camera based) for positional verification before treatment execution. Before the treatment execution, verification of patient's treatment setup under linac was checked with iViewC. Translational (x, y, and z) shifts of 3 mm margin were accepted, and necessary couch corrections were applied as and when required and the scheduled 3DCRT plan was subsequently executed. S t readings were recorded during treatment for all fields with IC (in transmission geometry as described earlier) and measurements were repeated 3 times (with an interval of 5-6 fractions) during 3DCRT in all 24 patients (n = 72 measurements). Schematic representation of the position of IC with buildup cap vis-ΰ-vis treatment head and representative treatment slice of the patient at four gantry angles (180°, 270°, 0°, and 90°) under transit study condition which has isocenter at the midplane of patient's planning CT slice is shown in [Figure 5]. The method of dose estimates is illustrated in Annexure [Table 4].



Figure 5: Schematic representation of position of ionization chamber with buildup cap vis -à - vis linac treatment head, and transverse slice of representative treatment slice of patient used as model at four angles (180°, 270°, 0°, and 90°) under transit study condition

Click here to view


During transit measurements with patients, the simultaneous measurement of in vivo dose at isocenter could not be performed as it was done in pelvic phantom. S t s were measured only for confirmed fields but not for wedged and subfields. The percentage deviation of D iso,Transit with D iso,TPS calculated from the following equation was noted.



The method of estimating 'in vivo' dose at isocentre using transit signal is outlined in Annexure [Table 5].




   Results Top


[Figure 6] shows the calibration curve of nC/cGy ratio for 100 MU delivery versus thickness of water for different field sizes for estimation D iso,Transit . Interpolation (for different field sizes and water equivalent thicknesses) was carried out using polyfit and polyval functions available in the MATLAB code to determine the D iso,Transit in routine clinical situations, is shown.
Figure 6: Plot of nano Coulombs/centi - Gray ratio versus thickness of water (in cm) for 100 monitoring units delivery for different field sizes

Click here to view


The in vivo dose values (D iso,Transit , D iso,Phantom and D iso,TPS ) obtained with pelvic phantom for four fields are shown in [Table 1]. The standard deviation of D iso,Transit with D iso,TPS and D iso,Phantom from individual fields was within ±1.7 and 0.6%, respectively. However, the mean percentage deviation of D iso,Transit with D iso,TPS and D iso,Phantom combined from all fields was 0.9 and 0.4% respectively with pelvic phantom. [Table 2] shows the results of estimated D iso,Transit (72 measurements) of 24 patients taken 3 times during treatments along with D iso,TPS values and the percentage deviation (calculated from equation 1) which was varied from −4.84% to 3.65%. The mean percentage deviation with standard deviation (SD) of estimated D iso,Transit with D iso,TPS is −1.37% ±2.03% (n = 72) observed from this group of patients. [Figure 7] shows the percentage deviation of estimated D iso,Transit with D iso,TPS in a group of 24 patients, during 3DCRT.
Figure 7: Percentage deviation of estimated Diso,Transit with Diso,TPS in a group of 24 patients (taken 3 times for each patient during the course of three - dimensional conformal radiotherapy)

Click here to view
Table 1: The percentage deviation of in vivo dose at isocenter obtained through transit method (Diso,Transit) with the values from TPS (Diso,TPS) and measured with CC01 (Diso,mid) in pelvic phantom


Click here to view
Table 2: The estimated Diso,Transit values (obtained in three fractions of all 24 patients) and corresponding Diso,TPS along with the percentage deviation


Click here to view



   Discussion Top


The transit dosimetry in different geometries has been attempted by various researchers. [7],[8],[9],[10] Goldenberg et al. [7] did a study to determine lung correction factors under actual treatment conditions in 8 patients treated for the middle third of the esophagus with cobalt-60 unit rotational therapy. Transit dose measurements by IC were related to direct measurements (temperature corrected) using intraluminal dosimeter and obtained results within 3%. Francois et al. [9] verified dose delivered on 38 patients treated with conformal therapy by a transit method. Central axis doses estimated by their formalism were compared with a measured dose which was found within the accepted tolerance of classical in vivo dosimetry (SD of 3.5%). A multicenter dose audit using similar transit measurement has shown an efficacy of dose estimation within an action level accuracy of 4%-5% with a measured deviation of 4% for thorax/pelvic treatments. [10] An earlier work [15] described the method of obtaining in vivo dose through transmission measurements with IC in a group of carcinoma cervix patients treated under telecobalt unit. An in vivo dosimetry system [16] used the measurement of transmission dose after phantom study, applied clinically on 11 patients who were treated for pelvic site with and without bone correction done in TPS. Mean errors were between −5.20% and +2.20% for AP-PA without bone correction, and between −0.62% and +3.32% with bone correction. For lateral fields, mean errors were between −10.80% and +3.46% without bone correction and between −0.55% and +3.50% with bone correction. It was brought out that transmission method is a useful form of in vivo dosimetry, because of noninvasiveness and simplicity with no additional efforts. The above authors emphasized that if bone corrections are not applied, the variation in transmission measurement can be as much as 10%. Even without any patient involved, their dosimetry variation of output was 2% over the course of patient treatments. The algorithm used in our study takes care of the inhomogeneity corrections in TPS, and the proposed method was validated in an in-house locally fabricated pelvic phantom before implementing in actual patients. In a recent study, Camilleri et al. [17] examined the feasibility of using an EPID-based in vivo dosimetry method that enables a point dose delivered to the patient to be calculated from the S t acquired with an EPID in a group of 53 patients treated by 3DCRT at pelvic site for 211 in vivo dose verifications. Excellent agreement was found between dose reconstructed at the isocenter and dose calculated by TPS with a mean deviation of −1.0% ±2.2% (1 SD). In vivo dose estimates obtained in this study are in good agreement with the accuracy levels obtained through various methods (done in pelvic malignancies) which are published in literature. [9],[10],[12],[13],[15],[16],[17] Our experience is limited to only symmetric fields where the central axis goes through the "ion chamber centre" in the exit position. If it is an asymmetric field or very long elongated field, the calibration cannot be extrapolated. The described method should be read with this limitation, and adequate caution must be exercised while using elongated fields. Wedged fields were not included in the present study, because of calibration differences for skewed isodose patterns. We plan to take up these aspects as a separate study in the future.


   Conclusion Top


The in vivo dosimetry has the potential to be useful in clinical settings, especially when treating regions not in close proximity to large tissue inhomogeneity. The proposed method in this study for in vivo dose verification has an added advantage because it is very simple to implement and can be implemented in clinics since the detector is kept away from the patient during transit measurements for all gantry angles. This work is not limited to only four-field box technique but also applicable to parallel opposed, three-field or five-field techniques as well. There are no difficulties arising pertaining to the positioning, reproducibility, and calibration. Measurements can be repeated during several sessions giving the opportunity to build new strategies for the validation by statistical evaluation of the data. The described method is applicable only when isocenter is at the mid-plane, both anterio-posteriorly and laterally. In situations when this is not the case, this method is not applicable. As the treatment planning algorithms have a few assumptions and extrapolations in dose calculations and display, our method will estimate the true path lengths encountered in the real patient, and therefore check the validity in delivered doses. This transmission method can quantify gross errors easily. We could estimate the dose to mid-point in the central axis with reasonable accuracy. However, this method is not a substitute to pretreatment quality assurance, but serves as an adjuvant dose verification, to track dose delivery and to catch gross errors which may be harmful for patients. This study also addresses a scientific problem where alternatives to EPID sophisticated dosimetry methods for transmission measurements are not available. With further work, this method may also be used to evaluate dose distribution variations throughout treatment fractions due to inter-fractional variability (weight loss, swelling) and help as a guide to adaptive radiotherapy.

Acknowledgment

Authors deeply acknowledge to Mr. Arun Kumar ES, Medical Radiation Dosimetrist, Department of Radiotherapy, Malabar Cancer Center, Moozhikkara P. O., Thalasseri, Kerala for his role in compiling the data, Mr. Jitendra Tirakala, Research associate, Department of Aerodynamic Engineering, Technical University, Munich, Germany in preparation of MATLAB Code for in vivo dose estimations and Dr. S.D. Sharma, Scientific Officer (G) and Head, Medical Physics Section, Associate Professor (Physical and Health Sciences), HBNI, RP and AD, BARC CT and CRS, Anushaktinagar, Mumbai, for his technical suggestions throughout our study.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
ICRU Report 24. Determination of Dose in Patient Irradiated by Means of X or Gamma Rays in Radiotherapy Procedures. International Commission on Radiation Units and Measurements. Bethesda; 1976.  Back to cited text no. 1
    
2.
Kutcher GJ, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ, et al. Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40. Med Phys 1994;21:581-618.  Back to cited text no. 2
    
3.
Procedures in external radiation therapy dosimetry with electron and photon beams with maximum energies between 1 and 50 MeV. Recommendations by the Nordic Association of Clinical Physics (NACP). Acta Radiol Oncol 1980;19:55-79.  Back to cited text no. 3
    
4.
Tripartite committee - The Faculty of Radiation Oncology, The Australian Institute of Radiography, The Australasian College of Physical Scientists and Engineers in Medicine. Developing standards for radiation treatment services in Australia. Sydney: RANZCR 2005. Available from: http://www.ranzcr.edu.au/documents/download.cfm/ Discussion%20Paper%20-%20Final.pdf?txtLibraryID=ranzcr&txtFile Name=ACF4E02.pdf. [Last accessed on 2008 Aug 18].  Back to cited text no. 4
    
5.
Podgorsak EB. Radiation Oncology Physics: A Handbook for Teachers and Students. Vienna: International Atomic Energy Agency; 2005.  Back to cited text no. 5
    
6.
Celi S, Costa E, Wessels C, Mazal A, Fourquet A, Francois P. EPID based in vivo dosimetry system: Clinical experience and results. J Appl Clin Med Phys 2016;17:6070.  Back to cited text no. 6
    
7.
Goldenberg DB, Rao UV, Lott S, Digel J. Dosimetric considerations in cobalt 60 rotational therapy for esophageal lesions. A comparison of transit and intraluminal dose measurement. Am J Roentgenol Radium Ther Nucl Med 1969;105:518-22.  Back to cited text no. 7
    
8.
Piermattei A, Fidanzio A, Azario L, Grimaldi L, D'Onofrio G, Cilla S, et al. Application of a practical method for the isocenter point in vivo dosimetry by a transit signal. Phys Med Biol 2007;52:5101-17.  Back to cited text no. 8
    
9.
Francois P, Boissard P, Berger L, Mazal A. In vivo dose verification from back projection of a transit dose measurement on the central axis of photon beams. Phys Med 2011;27:1-10.  Back to cited text no. 9
    
10.
Malicki J, Litoborski M, Kierzkowski J, Kosicka G. How the implementation of an in-vivo dosimetry protocol improved the dose delivery accuracy in radiotherapy? Rep Pract Oncol Radiother 2003;8:S348.  Back to cited text no. 10
    
11.
Development of Procedures for In vivo Dosimetry in Radiotherapy: IAEA. Human Health Reports Series No. 8, Viennna, Austria: International Atomic Energy Agency (IAEA); 2013.  Back to cited text no. 11
    
12.
Srinivas C, Shenoy KK, Dinesh M, Savitha KS, Kasturi DP, Supe S, et al. In vivo dosimetric study of carcinoma of uterine cervix with FBX solution in external beam therapy. J Med Phys 1999;24:47-50.  Back to cited text no. 12
  Medknow Journal  
13.
Srinivas C, Kumar PS, Ravichandran R, Banerjee S, Saxena PU, Kumar ES, et al. Dose verification in carcinoma of uterine cervix patients undergoing 3D conformal radiotherapy with Farmer type ion chamber. J Med Phys 2014;39:247-50.  Back to cited text no. 13
  Medknow Journal  
14.
IAEA. Absorbed Dose Determination in External Beam Radiotherapy. Technical Report Series TRS 398. Vienna: International Atomic Energy Agency; 2000.  Back to cited text no. 14
    
15.
Srinivas C, Shenoy KK, Shetty J, Solomon JG, Ravichandran R, Supe S, et al. In-vivo dosimetry: Confirmation of dose delivery through transit dose measurements in carcinoma of uterine cervix patients undergoing teletherapy. J Med Phys 2004;29:177-8.  Back to cited text no. 15
    
16.
Kim BK, Chie EK, Huh SN, Lee HK, Ha SW. Clinical application of in vivo dosimetry system in radiotherapy of pelvis. J Radiat Prot Res 2002;27:37-49.  Back to cited text no. 16
    
17.
Camilleri J, Mazurier J, Franck D, Dudouet P, Latorzeff I, Franceries X. Clinical results of an EPID-based in-vivo dosimetry method for pelvic cancers treated by intensity-modulated radiation therapy. Phys Med 2014;30:690-5.  Back to cited text no. 17
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,546 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed964    
    Printed22    
    Emailed0    
    PDF Downloaded94    
    Comments [Add]    

Recommend this journal