ORIGINAL ARTICLE
Year : 2014  |  Volume : 39  |  Issue : 1  |  Page : 24-31
 

Effective atomic numbers of some tissue substitutes by different methods: A comparative study


Department of Physics, Karnatak University, Dharwad, Karnataka, India

Date of Submission13-May-2013
Date of Decision04-Nov-2013
Date of Acceptance04-Nov-2013
Date of Web Publication20-Jan-2014

Correspondence Address:
Vishwanath P Singh
Health Physics Section, Kaiga atomic Power Station-3&4, NPCIL, Karwar, 581400, Karnatak
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.125489

Rights and Permissions

 

   Abstract 

Effective atomic numbers of some human organ tissue substitutes such as polyethylene terephthalate, red articulation wax, paraffin 1, paraffin 2, bolus, pitch, polyphenylene sulfide, polysulfone, polyvinylchloride, and modeling clay have been calculated by four different methods like Auto-Z eff, direct, interpolation, and power law. It was found that the effective atomic numbers computed by Auto-Z eff , direct and interpolation methods were in good agreement for intermediate energy region (0.1 MeV < E < 5 MeV) where the Compton interaction dominates. A large difference in effective atomic numbers by direct method and Auto-Z eff was observed in photo-electric and pair-production regions. Effective atomic numbers computed by power law were found to be close to direct method in photo-electric absorption region. The Auto-Z eff , direct and interpolation methods were found to be in good agreement for computation of effective atomic numbers in intermediate energy region (100 keV < E < 10 MeV). The direct method was found to be appropriate method for computation of effective atomic numbers in photo-electric region (10 keV < E < 100 keV). The tissue equivalence of the tissue substitutes is possible to represent by any method for computation of effective atomic number mentioned in the present study. An accurate estimation of Rayleigh scattering is required to eliminate effect of molecular, chemical, or crystalline environment of the atom for estimation of gamma interaction parameters.


Keywords: Atomic numbers, Compton scattering, gamma, mixture rule, tissue substitutes, Rayleigh scattering


How to cite this article:
Singh VP, Badiger N M. Effective atomic numbers of some tissue substitutes by different methods: A comparative study. J Med Phys 2014;39:24-31

How to cite this URL:
Singh VP, Badiger N M. Effective atomic numbers of some tissue substitutes by different methods: A comparative study. J Med Phys [serial online] 2014 [cited 2019 May 21];39:24-31. Available from: http://www.jmp.org.in/text.asp?2014/39/1/24/125489



   Introduction Top


Simulation of depth-dose distribution inside human organs and tissues is made by tissue equivalent materials. The tissue equivalent materials are called as tissue substitutes for various tissues, organs of the human body, and tissue components having similar properties with respect to ionizing radiation. Average soft tissues are mainly composed of low-atomic number (Z) such as H, C, N, O, and so on. [1] International Commission on Radiation Units (ICRU) report [2] describes various types of tissue substitutes which are used in medicine, radiation protection, and radiolobiology to calibrate the radiation detectors and application in nuclear engineering for realistic body phantom. [3],[4] The effective atomic number is most vital parameters for tissue equivalence, radiation absorption, radiation scattering, and shielding effectiveness for gamma and neutron for compound materials.

Waxes are organic compounds which consist of longalkyl chains. Polymers are made from the monomer propylene which is rugged and highly resistant to the chemicals. Polymers are low-Z, nonflammable, light weight, high durability, ease processing, economical, and stable against environment. Nowadays, radiation shielding for gamma rays as well as neutron ismade of polymer matrix lead shielding materials. Various types of waxes, plastics, and polymers are being used as tissue substitutes in field of medical, dosimetry, and radiological protection.

Investigators have studied effective atomic numbers of gaseous mixtures, [5] composite materials, [6] solutions, [7] dosimetric materials, [8],[9] and biological materials. [10],[11],[12],[13] Several photon interaction studies are reported for low-Z materials [14],[15],[16],[17] for X- and gamma-ray photons. Rubbers containing varying degree of carbon show a wide range of effective atomic numbers. [18] Mass attenuation coefficients of few common tissue substitutes have been reported. [19] Recently, alcohol tissue substitutes for human organs have been investigated. [20] The equivalence of tissue substitutes for experimental radiation physics has been reported by (⊠/ρ) substitute /(⊠/ρ) tissue for energy 0.01-100 MeV and linear attenuation coefficients at 60-80 keV for radiation characteristic of tissue substitutes. [21]

The bolus tissue substitute is easily made, inexpensive, and moldable during treatment. [22],[23] The paraffin is lipid equivalent material [2] and polyethylene terephthalate is a good skeleton-cartilage substitute. [2] Red articulation wax is used for dosimetry phantom preparation and pitch is very good skin equivalent material. The effective atomic number dependency upon photon energy is evaluated by various methods like Auto-Z eff , [24] direct method, [25] and logarithmic interpolation method. These methods use different input parameters (e.g. atomic cross-sections, atomic numbers, and attenuation coefficients) for computation of effective atomic numbers. The direct and interpolation methods use mixture rule [26],[27] for computation of mass attenuation coefficients of the compound or mixture. The mixture rule is simple additive law for weighted sum of mass attenuation of constituent elements. The photon interaction processes are energy dependent and photo-electric effect (E <0.1 MeV) is largely energy dependent. Appreciable differences between experiment and theoretical mass attenuation coefficients of polymers (59.5 keV), vitamins (30.82-59.54 keV), amino acids (8.04-80.99 keV), Au alloys (59.5 keV), and brass alloys (81 keV) are noted. [13],[28],[29],[30],[31] Effective atomic numbers of high-Z compounds like magnesium ferrite and borosilicate glass are observed to be following the mixture rule in low-energies. [32],[33] It is expected that the effective atomic numbers computed by different theoretical methods should be identical in magnitude for the compound materials at selected energies. Presently, such type of comparative study is not found in the literature and was studied first time.

In view of above, we have chosen some important tissue substitutes (1 ≤ Z < 18) given in [Table 1] for computation of effective atomic numbers by Auto-Z eff , direct, interpolation, and power law methods for various medical applications. This study will benefit for readily available effective atomic numbers of the tissue substitutes and choice for appropriate method for computation of effective atomic numbers.
Table 1: Elemental composition of tissues substitutes

Click here to view


Computational work and theoretical background

Mass attenuation coefficients and effective atomic numbers of compound materials are derived by mixture rule using mass attenuation coefficients and atomic cross-sections of the elements. In our study, we have computed the mass attenuation coefficients and effective atomic numbers of the tissue substitutes given in [Table 1]. These materials have been taken from the references. [2],[34],[35] The attenuation cross-section data can be found for 100 of elements in energy range of 1 keV-100 GeV by XCOM program. [36] The XCOM data have been transformed to user-friendly software package WinXCom [37] for the window platform which is easily exportable in the excel files. Using WinXCom, mass attenuation coefficients and attenuation cross-section data were generated for the elements in photon energy region 10 keV-20 MeV. The atomic numbers and atomic masses of the elements are taken from atomic weight of elements 2009, International union of Pure and Applied Chemistry. [38]

Mass attenuation coefficients



Auto-Z eff

Auto-Z eff is user-friendly software in visual basic for rapid computation of the energy-dependent effective atomic numbers, average atomic numbers, and spectral-weighted mean atomic numbers. Auto-Z eff surpasses dubious power-law approach. In this method, Z eff , Auto is determined via exploitation of the smooth correlation between atomic cross-section and atomic number. A matrix of cross-sections was constructed spanning atomic number Z = 1-100 for photon energies ranging between 10 keVand1 GeV and cross-sections of polyelemental media are calculated by linear additivity. The cross-sectional values are constructed with the cross-section matrix as a function of Z and an effective atomic number for any energy is obtained by interpolation of Z values between adjacent cross-section data. [24]

Direct method

Computation of the effective atomic number, Z eff, PI of the selected tissue substitutes for total gamma photon interaction has been carried out by practical formula. [25] The mass attenuation coefficients of the elements have been obtained from WinXcom computer program. The effective atomic number, Z eff, PI is given by



where ∫1 and ∫2 are the elemental cross-section (barn/atom) in between which the atomic cross-section ∫ of the tissue substitutes and Z 1 and Z 2 are atomic numbers of the elements (dimensionless) corresponding to the cross sections ∫1 and ∫2 , respectively.

Power law method

The effective atomic number, Z eff , PL , of a tissue substitute material by power law can be calculated according to the following equation:



where a 1 , a 2 ,… are the fractional contents of electron belonging to element Z 1 , Z 2 ,.………., respectively, n i is the numberof electrons, in one mole, belonging to each element Z i and N A is the Avogadro's number. The x values are in 2.94 [39] and 3.5 [40] ranges.

Errors

Auto-Z eff software calculates effective atomic numbers with errors of 1%-2% for photon energies 10 keV-1000 MeV. This errors increase to 25%-50% below10 keV. [24] The errors in mass attenuation coefficients of the elements is about 1% for low-Z (1 < Z < 8) in the energy region where Compton scattering dominates (30 keV-100 MeV). [41] Below 30 keV and above 100 MeV, the errors are as much as 5%-10%. For medium-Z elements sodium through copper, the errors are 1%-2% for energies 10 keV-1 MeV and 2%-3% for energies 1-100 MeV. Medical, biological, and industrial, applications and transportation tend to use sources with photon energies above 5 keV, so that the errors in our results may not have any practical impact. The power law method is an inaccurate method as the exponent of 2.94 relates to an empirical formula for the photoelectric process which incorporates a ''constant'' of 2.64 × 10−26 , which is in fact not a constant but rather a function of the photon energy. A linear relationship between Z 2.94 has been shown for a limited number of compounds for low-energy x-rays. [42]


   Result and Discussion Top


Variation of mass attenuation coefficients of selected tissue substitutes is shown in [Figure 1] for photon energy range 1 keV-100 GeV. The effective atomic numbers of the selected tissue substitutes by Auto-Z eff , direct, interpolation methods, and power law is shown in [Figure 2]a-j. The variations of these parameters were explained in detail in the next section.
Figure 1: Mass attenuation coefficients of tissue substitutes in photon energy range 1 keV-100 GeV

Click here to view
Figure 2: Effective atomic numbers of Red Articulation Wax, Polyethylene Terephthalate, Paraffin 1, Paraffin 2, Bolus, Pitch, Poly - Phenylene Sulfide, Poly-Sulfone, Poly-vinyl-chloride and Modeling Clay for photon energy ranges from 10 keV to 20 MeV by Auto-Zeff, direct, interpolation, and power Law

Click here to view


Mass attenuation coefficient

[Figure 1] shows the mass attenuation coefficient, ∝/ρ variation of tissue substitutes in photon energy range 1 keV-100 GeV. The [Figure 1] shows that the ∝/ρ values of the tissues substitutes decrease with increase in the photon energies in low-energy, minimum in intermediate-energy, and constant in high-energy region. The variation in ∝/ρ values with energy can be explained by partial photon interaction processes as the dependence of the total atomic cross-section on atomic number and photon energy. The interaction cross-section is directly proportional to Z 4-5/ E 3.5 for photo-electric absorption in low-energy; therefore, ∝/ρ values of the tissues substitutes reduces sharply. In Compton scattering region (intermediate-energy region), the interaction cross-section is dependent upon Z/E, whereas cross-section is directly proportional to Z 2 for pair-production (high-energy). The ∝/ρ values of all the selected tissues substitutes merges in Compton scattering (100 keV < E < 10 MeV) region.

The ∝/ρ values of Griffith breast tissue substitutes and breast of human body tissue [2] were compared with the selected tissue substitutes. It was found that the Polyethylene Terephthalate polyphenylene sulfide, and polysulfone, can replace Griffith tissue in photon energy region 0.10-10 MeV, whereas Polyethylene Terephthalate is most suitable for entire energy region

Effective atomic number

Effective atomic numbers of the selected tissue substitutes by different four methods (Auto-Z eff , direct, interpolation, and power law) is shown [Figure 2]a-j. From [Figure 2], it is observed that the effective atomic numbers computed by Auto-Z eff , direct and interpolation methods are in good agreement and almost identical in the energy region 0.1-5 MeV where the Compton interaction dominates. The effective atomic number values were found constant in the intermedium photon energy region, whereas significant variation was observed in the lower (0.01-0.1 MeV) as well as in the higher-energy regions (5-20 MeV). The effective atomic numbers computed by direct method were higher in photo-electric absorption and pair-production regions as compared with interpolation method. The effective atomic numbers calculated by Auto-Z eff were 6.01-6.23, 4.55-6.07, 5.90-6.27, 5.84-6.21, 5.96-6.28, 6.87-7.22, 5.09-9.07, 4.30-6.88, 5.34-11.02, and 7.63-8.14 for RAW, PETE, PF1, PF2, BOLUS, PITCH, PPS, PSU, PVC and MC- respectively, by direct method were 6.01-6.46, 4.55-6.81, 5.90-6.6, 5.84-6.52, 5.96-6.58, 6.87-7.63, 5.11-13.64, 4.30-9.71, 5.34-16.05, and 7.62-8.96 for RAW, PETE, PF1, PF2, BOLUS, PITCH, PPS, PSU, PVC and MC, respectively; whereas by interpolation method 6.00-6.29, 4.57-6.09, 5.91-6.34, 5.85-6.27, 5.97-6.35, 6.88-7.28, 5.11-9.08, 4.40-6.91, 5.37-11.02, and 7.64-8.18 for RAW, PETE, PF1, PF2, BOLUS, PITCH, PPS, PSU, PVC, and MC, respectively. The effective atomic numbers by power law is shown in the graphs for x equals to 2.94 and 3.5. The independency of effective atomic numbers on photon energy in Compton dominant region can be found in various literatures for materials containing low- and high-Z elements; however, photo-electric absorption and pair-production region are not found experimentally. [6],[7],[9] The effective atomic numbers computed by power law were found to be in the photo-electric absorption region close to direct method. In Compton scattering region, it was found that the effective atomic numbers by all the three methods were in good agreement.

The variation in effective atomic numbers of the tissue substitutes by Auto-Z eff , direct, and interpolation methods may be due to basic concept and input parameters for computation. In the Auto-Zeff effective atomic numbers are determined via exploitation of the smooth correlation between atomic cross-sections, atomic numbers, and mass attenuation coefficients. The cross-sections of polyelement compounds are calculated by linear additivity. The effective atomic numbers of the tissue substitutes was calculated by interpolation of Z of adjacent cross-section data of cross-section matrix as function of Z. Large variation in effective atomic number by direct method and interpolation method is observed because of different weight fractions are given as input in computation.

For tissue equivalence of the tissue substitutes, a graph for substitute (PVC) and tissue (cortical bone) has been plotted in [Figure 3]. The ratio of effective atomic numbers by direct, Auto-Z eff , and interpolation for substitute to tissue along with (∝/ρ)substitute/ (∝/ρ)tissue reported by White for photon energies 0.01, 0.1, 1, 10, and 100 MeV [21] are shown on abscissa; whereas radiation characteristics (effective atomic numbers and mass attenuation coefficients) are shown in the ordinate. It is evident from the [Figure 3] that the tissue equivalence of PVC is far away for (∝/ρ)substitute/ (∝/ρ)tissue method in low energy (<0.1 MeV) compared with effective atomic number method. In Compton scattering region, a good correlation is to be noted. With increase in the photon energy (>10 MeV), the deviations by both the methods were observed of same order. The ratio of Z eff, PL of PVC to cortical bone was found to be 1.05 and 1.03 for x values of 2.94 and 3.5 respectively. Therefore, the double-valued Z eff, PL using power law method shows insignificant variation of the radiation characteristics of PVC and cortical bone. The difference between the ratios of effective atomic numbers of tissue substitutes to tissues using above methods (direct, Auto-Z eff , and interpolation) was noted insignificant. Therefore, it can be concluded that the tissue equivalence of the tissue substitutes is possible to represent by using above methods (Auto-Z eff , direct, interpolation) for computation of effective atomic numbers.
Figure 3: Comparison of radiation characteristics (mass attenuation coefficients and effective atomic number) of poly - vinyl - chloride and cortical bone at photon energies 0.01, 0.1, 1, 10, and 100 MeV

Click here to view


In case of photo-electric absorption region, the interaction process is affected near absorption edges of the elements which may perturb the wave-function of the compound or mixture. The results support the conclusion elaborated in earlier several reports [41],[42] that the mixture rule is valid for the evaluation of photon attenuation coefficients for compounds in Compton scattering region. Therefore, mixture rule is limited to Compton scattering region and demand for further experimental data for smoothening of the interaction cross sections and it's applicability in photo-electric (10 keV < E < 100 keV) and pair-production regions (E > 10 MeV). From [Figure 2], it is evident that in photo-electric absorption region the effective atomic numbers by interpolation method and Auto-Z eff are far away from direct method. The experimental data for effective atomic numbers of various compounds/mixtures [13],[28],[29],[30],[31] and power law method results were found to be closest to the direct method; therefore, it is concluded that the direct method may be appropriate for computation of effective atomic numbers in photo-electric absorption region.

Discrepancies

The mass attenuation coefficients of chemical compound or mixture is evaluated by weighted sum of mass attenuation coefficients of the constituent elements using mixture rule. The mixture rule is not valid near k-absorption edge of the compounds and mixtures. [43],[44],[45],[46] The mixture rule doesnot consider the molecular, chemical, or crystalline environment of the atom which results in change in the atomic wave-function. With the exception of the fine structure region above absorption edge (>10 keV) errors from these sources are expected to be a few percent, whereas at low energies (10-100 eV), errors of as much as a factor of two can occur. In case of Auto-Z eff , the cross-sections of poly-elemental media are also calculated by linear additive method.

The study upon molecular, chemical, or crystalline environment of the atom [43],[44] reveals that there is requirement for revision of the cross-section and attenuation coefficients libraries in photo-electric absorption region. The reason behind is that the actual energy at which a particular photo-electric absorption edge of an element occurs is dependent on the chemical state of the absorbing atom and the nature of the chemical environment: For example, a chemical shift in the position of the k-edge in iron can be seen as the oxidation state of the atom is changed. [47] High-resolution studies show that within the edge region the structure depends on the electronic structure of the absorbing atom. The actual height of the edge is also sensitive to the atomic environment and hence there may be uncertainty in determining the k-jump ratios. [48] The errors in theoretical estimates of Rayleigh scattering may be the cause of part of the deviation in absorption coefficient are observed. [48] It may be concluded that the theoretical estimates for Rayleigh scattering for the photon energies very close to the K-edge are not accurate. An accurate estimation of Rayleigh scattering is required to eliminate effect of molecular, chemical, or crystalline environment of the atom. Therefore, the discrepancies in the effective atomic number may be diminished by considering the molecular, chemical, or crystalline environment of the atom for deriving the atomic cross-section, interaction and attenuation coefficients.[49]


   Conclusions Top


In the present study, we have compared effective atomic numbers of some important tissue substitutes by four methods (i.e., Auto-Z eff , direct method, interpolation method, and power law). The direct and power law methods are applicable in low-photon energy (10 keV < E < 100 keV) where photo-absorption dominate and intermediate-energy (0.1 MeV < E < 10 MeV) where Compton scattering interaction dominates. A large difference in effective atomic numbers by direct method and Auto-Z eff methods was observed in photo-electric and pair-production regions. An accurate estimation of Rayleigh scattering is required to eliminate effect of molecular, chemical, or crystalline environment of the atom. Therefore, the discrepancies in the effective atomic number may be diminished by considering the molecular, chemical, or crystalline environment of the atom for deriving the atomic cross-section, interaction, and attenuation coefficients. Direct method was found to be an appropriate method for deriving effective atomic numbers in photo-electric region (10 keV < E < 100 keV).

 
   References Top

1.General principles for the radiation protection of workers. Ann ICRP 1997;27.  Back to cited text no. 1
    
2.International Commission on Radiation Units and Measurements ICRU. Tissue substitutes in radiation dosimetry and measurement, Report No. 44, ICRU, Betheda; 1989.  Back to cited text no. 2
    
3.Indra Y, Daniel K, Oliver L, Willi AK. A formulation of tissue- and water-equivalent materials using the stoichiometric analysis method for CT-number calibration in radiotherapy treatment planning. Phys Med Biol 2012;57:1173-90.  Back to cited text no. 3
    
4.Constantinou C. Tissue substitutes for particulate radiations and their use in radiation dosimetry and radiotherapy," Ph.D. thesis, University of London, 1978.  Back to cited text no. 4
    
5.Singh VP, Badiger NM. Effective atomic numbers, electron densities and tissue equivalence of some gases and mixtures for dosimetry of radiation detectors. Nucl Technol Radiat Protect 2012;27:117-24.  Back to cited text no. 5
    
6.Prasanna KS, Manjunathguru V, Umesh TK. Effective atomic numbers of some H-, C-, O- based composite materials derived from differential incoherent scattering cross section. PRAMANA-J. Phys 2010;74:555-62.  Back to cited text no. 6
    
7.Kulwant S, Gagandeep K, Sandhu GK, Lark BS. Interaction of photons with some solutions. Rad Phys Chem 2001;61:537-40.  Back to cited text no. 7
    
8.Kiran TK Venkata KR. Effective atomic numbers for materials of dosimetric interest. Rad Phys Chem 1997;50:545-53.  Back to cited text no. 8
    
9.Shivaramu R, Vijayakumar L, Rajasekaran NR. Effective atomic numbers for photon energy absorption of some low-Z substances of dosimetric interest. Rad Phys Chem 2001;62:371-7.  Back to cited text no. 9
    
10.Kaginelli SB, Rajeshwari T, Sharanabasappa, Kerur BR, Kumar AS. Effective atomic numbers and electron density of dosimetric material. J Med Phys 2009;34:176-9.  Back to cited text no. 10
[PUBMED]  Medknow Journal  
11.Shivalinge G, Krishnaveni S, Ramakrishna G. Studies on effective atomic numbers and electron densities in amino acids and sugars in the energy range 30 1333 keV. Nucl Instr Meth Phys Res Sec B 2005;239:361-9.  Back to cited text no. 11
    
12.Manjunathguru V, Umesh TK. Effective atomic number and electron densities of some biologically important compounds containing H, C, N and O in energy range 145-1330 keV. J Phys B At Mol Opt Phys 2006;39:3969-81.  Back to cited text no. 12
    
13.Demir D, Tursucu A, Oznuluer T. Studies on mass attenuation coefficient, effective atomic number and electron densitiy of some vitamins. Radiat Environ Biophys 2012; 51:469-75.  Back to cited text no. 13
    
14.Tejbir S, Rajni, Updesh K, Parjit SS. Photon energy absorption parameters for some polymers. Ann Nucl Eng 2010;37:422-7.  Back to cited text no. 14
    
15.Tejbir S, Naresh K, Parjit SS. Chemical composition dependency of exposure buildup factors for some polymers. Ann Nucl Eng 2009;36:114-20.  Back to cited text no. 15
    
16.Murat K, Yuksel O. Energy absorption and exposure buildup factors for some polymer and tissue substitute materials: Photon energy, penetration depth and chemical composition dependency. J Radiol Prot 2011;31:117-28.  Back to cited text no. 16
    
17.Parthsaradhi K, Esposito A, Pelliccioni M. Photon attenuation coefficients in tissue equivalent compound. Appl Radiat Isot 1992;43:1481-4.  Back to cited text no. 17
    
18.Elias S, Zainal AS, Anuar A, Husin W. Determination of effective atomic number of rubber. Pertanika 1983:95-8.  Back to cited text no. 18
    
19.Ferreira CC, Ximenes RE, Garcia CAB, Viera JW, Maia AF. Total mass attenuation coefficient evaluation of ten materials commonly used to simulate human tissue, International Conference on Defect in Insulating Materials. J Phy Conf Series 2010; 249:012029.  Back to cited text no. 19
    
20.Singh VP, Badiger NM. Study of effective atomic numbers and electron densities, kerma of alcohols: Phantom and human organ tissue substitutes. Nucl Technol Radiat Protect 2013;28:137-45.  Back to cited text no. 20
    
21.White DR. Tissue substitutes in experimental radiation physics. Med Phys 1978;5:467-79.  Back to cited text no. 21
[PUBMED]    
22.Jorge ET, Luis FV, Eleni M. Development and dosimetric characterization of a tissue substitute (Bolus) for use in linear accelerator electron radiotherapy. AIP Conf Proc 2003;682:118-23.  Back to cited text no. 22
    
23.Seppälä T, Collan J, Auterinen I, Serén T, Salli E, Kotiluoto P, et al. A dosimetric study on the use of bolus materials for treatment of superficial tumors with BNCT. Appl Radiat Isot 2004;61:787-91.  Back to cited text no. 23
    
24.Taylor ML, Smith RL, Dossing F, Franich RD. Robust calculation of effective atomic numbers: The Auto-Zeff software. Med Phys 2012;39:1769-78.  Back to cited text no. 24
[PUBMED]    
25.Manohara SR, Hanagodimath SM, Thind KS, Gerward L. On the effective atomic number and electron density: A comprehensive set of formulas for all types of materials and energies above 1 keV. Nucl Instru Meth Phys Res B 2008;266:3906-12.  Back to cited text no. 25
    
26.Deslattes RD. Estimates of x-ray attenuation coefficients for the elements and their compounds. Acta Cryst 1969;A25, Part 1:89-93.  Back to cited text no. 26
    
27.Jackson DF, Hawkes DJ. X-ray attenuation coefficients of elements and mixtures. Phys Rep 1981;70:169-233.  Back to cited text no. 27
    
28.Nil K, Merve C, Nihat AI., Mass attenuation coefficients, effective atomic numbers and electron densities of some polymers. Radiat Prot Dosim 2013; 153:127-34.  Back to cited text no. 28
    
29.Turºucu A, Önder P, Demir D, Öznülüer T. Studies on mass attenuation coefficient, effective atomic number and electron density of some amino acids. Int J Phys Sci 2013;8:147-56.  Back to cited text no. 29
    
30.Han I, Demir L. Studies on effective atomic numbers, electron densities and mass attenuation coefficients in Au alloys. J X-ray Sci Technol 2010;18:39-46.  Back to cited text no. 30
    
31.El-Kateb AH, Rizk RA, Abdul-Kader AM. Determination of atomic number cross-sections and effective atomic number for some alloys. Ann Nucl Ener 2000;27:1333-43.  Back to cited text no. 31
    
32.Kadam RH, Alone ST, Bichile GK, Jadhav KM. Measurement of atomic number and mass attenuation coefficient in magnesium ferrite. Pranama J Phys 2007;68:869-74.  Back to cited text no. 32
    
33.Chanthima N, Kaewkhao J. Investigation on radiation shielding parameters of bismuth borosilicate glass from 1 keV to 100 GeV, Ann Nucl Energy 2013; 55: 23-8.  Back to cited text no. 33
    
34.Wang M, Yue CY, Chua B. Production and evaluation of hydroxyapatite reinforced polysulfone for tissue replacement, J Mater Sci Mater Med 2001;12:821-6.  Back to cited text no. 34
    
35.Fang L, Wen MC, Zhou Y. Structure research on carrying frame in portable exoskeleton make by polyphenylene sulfide. Adv Mater Res 2012;591-593:96-101.  Back to cited text no. 35
    
36.Berger MJ, Hubbell J H Seltzer S M, Chang J, Coursey J S, Sukumar R, Zucker D S, Olsen K (2010) XCOM: photon cross sections database, NIST standard reference database. http://www.nist.gov/pml/data/xcom/index.cfm.  Back to cited text no. 36
    
37.Gelward L, Guilbert N, Jensen KB, Levring H. WinXcom-a program for calculating X-ray attenuation coefficients. J Radi Phys Chem 2004;71:653-4.  Back to cited text no. 37
    
38.Michael EW, Michal B. Atomic weight of elements 2009, (IUPAC Technical Report). Pure Appl. Chem. 2011; 83:359-96.  Back to cited text no. 38
    
39.Mayneord WV. Acta Int Union Against Cancer 1937;2:271.  Back to cited text no. 39
    
40.Harold EJ, Cunningham JR. The physics of radiology. In: USA: Charles C Thomas Publisher; 1983.  Back to cited text no. 40
    
41.Hubbell JH. Photon cross sections, attenuation coefficients and energy absorption coefficients from 10 keV to 100 GeV, 1969, NSRDS-NBS 29.  Back to cited text no. 41
    
42.Spiers W. Effective atomic number and energy absorption in tissues. Br J Radiol 1946;19:52-63.  Back to cited text no. 42
    
43.Deslattes RD. 1969;186:1.  Back to cited text no. 43
    
44.Kerur BR, Thontadarya SR, Hanumaiah B. X-ray attenuation coefficients at 6.46 keV and the validity of the mixture rule for compounds. X-Ray Spectrometry 1993;22:13-6.  Back to cited text no. 44
    
45.Turgut U, Þimþek, Büyükkasap E, Ertuðrul M. X-Ray attenuation coefficients at different energies and the validity of the mixture rule for compounds around the absorption edge. Spectrochimica Acta Part B Atom Spectroscopy 2002;57:261-6.  Back to cited text no. 45
    
46.Sharanabasappa, Chitralekha A, Kerur BR, Anilkumar S. Study of non-validity of mixture rule near K-absorption edges by X-ray spectrometric technique. J Radioanalyt Nucl Chem 2012;291:895-900.  Back to cited text no. 46
    
47.Turgut U, Simsek O, Buyukkasap E. Measurement of mass attenuation coefficients in some Cr, Co and Fe compounds around the absorption edge and the validity of the mixture rule, PRAMANA 2007;69:199-207.  Back to cited text no. 47
    
48.Jackson DF. Chemical effects in X-ray transmission measurements. Nucl Instrum Meth 1982;193:387.  Back to cited text no. 48
    
49.Kerur B R, Thontadarya S R, Hanumaiah B., Anomalous X-ray attenuation coefficients around the absorption edges using Mn-Kx, Cu-K X-rays. Appl Radiat.Isot 1994; 45: 159-63.  Back to cited text no. 49
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (2,630 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Result and Discu...
   Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed3541    
    Printed69    
    Emailed0    
    PDF Downloaded125    
    Comments [Add]    

Recommend this journal