Investigation on the effect of sharp phantom edges on point dose measurement during patient-specific dosimetry with Rapid Arc


1 Department of Medical Physics, Tata Memorial Hospital, Mumbai, India
2 Department of Radiation Oncology, Tata Memorial Hospital, Mumbai, India
3 Department of Electronics, Brijlal Biyani Science College, Amravati, Maharashtra, India

Correspondence Address:
R A Kinhikar
Department of Medical Physics, Tata Memorial Hospital, Parel - 400 012, Mumbai, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.116373

Rights and Permissions

The objective of this work was to investigate and quantify the effect of sharp edges of the phantom on the point dose measurement during patient-specific dosimetry with Rapid Arc (RA). Ten patients with carcinoma of prostate were randomly selected for this dosimetric study. Rapid Arc plans were generated with 6 MV X-rays in the Eclipse (v 8.6.14) with single arc (clockwise). Dosimetry verification plans were generated for two phantoms (cylindrical and rectangular). The cylindrical phantom was solid water (diameter 34 cm) and the rectangular phantom was a water phantom (25 cm × 25 cm × 10 cm). These phantoms were pre-scanned in computed tomography (CT) machine with cylindrical ionization chamber (FC65) in place. The plans were delivered with Novalis Tx linear accelerator with 6 MV X-rays for both the phantoms separately. The measured dose was compared with the planned dose for both the phantoms. Mean percentage deviation between measured and planned doses was found to be 4.19 (SD 0.82) and 3.63 (SD 0.89) for cylindrical and rectangular phantoms, respectively. No significant dosimetric variation was found due to the geometry (sharp edges) of the phantom. The sharp edges of the phantom do not perturb the patient specific Rapid Arc dosimetry significantly.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1241    
    Printed34    
    Emailed1    
    PDF Downloaded109    
    Comments [Add]    

Recommend this journal