ORIGINAL ARTICLE
Year : 2013  |  Volume : 38  |  Issue : 1  |  Page : 22-29
 

Dosimetric comparison of intensity modulated radiotherapy isocentric field plans and field in field (FIF) forward plans in the treatment of breast cancer


Department of Radiotherapy, Royal Hospital, Muscat, Oman

Date of Submission03-Jul-2012
Date of Decision29-Nov-2012
Date of Acceptance01-Dec-2012
Date of Web Publication29-Jan-2013

Correspondence Address:
Ramamoorthy Ravichandran
Department of Radiotherapy, Medical Physics Unit, National Oncology Center, Royal Hospital, Muscat, Sultanate of Oman PBox 1331, PC 111, Seeb, Muscat
Oman
Login to access the Email id


PMID: 23531607

Get Permissions

 

   Abstract 

The present study is aimed at comparing the planning and delivery efficiency between three-dimensional conformal radiotherapy (3D-CRT), field-in-field, forward planned, intensity modulated radiotherapy (FIF-FP-IMRT), and inverse planned intensity modulated radiotherapy (IP-IMRT). Treatment plans of 20 patients with left-sided breast cancer, 10 post-mastectomy treated to a prescribed dose of 45 Gy to the chest wall in 20 fractions, and 10 post-breast-conserving surgery to a prescribed dose of 50 Gy to the whole breast in 25 fractions, with 3D-CRT were selected. The FiF-FP-IMRT plans were created by combining two open fields with three to four segments in two tangential beam directions. Eight different beam directions were chosen to create IP-IMRT plans and were inversely optimized. The homogeneity of dose to planning target volume (PTV) and the dose delivered to heart and contralateral breast were compared among the techniques in all the 20 patients. All the three radiotherapy techniques achieved comparable radiation dose delivery to PTV-95% of the prescribed dose covering > 95% of the breast PTV. The mean volume of PTV receiving 105% (V 105 ) of the prescribed dose was 1.7% (range 0-6.8%) for IP-IMRT, 1.9% for FP-IMRT, and 3.7% for 3D-CRT. The homogeneity and conformity indices (HI and CI) were similar for 3D-CRT and FP-IMRT, whereas the IP-IMRT plans had better conformity index at the cost of less homogeneity. The 3D-CRT and FiF-FP-IMRT plans achieved similar sparing of critical organs. The low-dose volumes (V 5Gy ) in the heart and lungs were larger in IP-IMRT than in the other techniques. The value of the mean dose to the ipsilateral lung was higher for IP-IMRT than the values for with FiF-FP-IMRT and 3D-CRT. In the current study, the relative volume of contralateral breast receiving low doses (0.01, 0.6, 1, and 2Gy) was significantly lower for the FiF-FP-IMRT and 3D-CRT plans than for the IP-IMRT plan. Compared with 3D-CRT and IP-IMRT, FiF-FP-IMRT proved to be a simple and efficient planning technique for breast irradiation. It provided dosimetric advantages, significantly reducing the size of the hot spot and minimally improving the coverage of the target volume. In addition, it was felt that FiF-FP-IMRT required less planning time and easy field placements.


Keywords: Breast cancer, contralateral breast, field in field breast plan, intensity-modulated radiation therapy


How to cite this article:
Al-Rahbi ZS, Al Mandhari Z, Ravichandran R, Al-Kindi F, Davis CA, Bhasi S, Satyapal N, Rajan B. Dosimetric comparison of intensity modulated radiotherapy isocentric field plans and field in field (FIF) forward plans in the treatment of breast cancer. J Med Phys 2013;38:22-9

How to cite this URL:
Al-Rahbi ZS, Al Mandhari Z, Ravichandran R, Al-Kindi F, Davis CA, Bhasi S, Satyapal N, Rajan B. Dosimetric comparison of intensity modulated radiotherapy isocentric field plans and field in field (FIF) forward plans in the treatment of breast cancer. J Med Phys [serial online] 2013 [cited 2014 Nov 27];38:22-9. Available from: http://www.jmp.org.in/text.asp?2013/38/1/22/106601



   Introduction Top


In patients with early breast cancer, local standard therapy is breast-conserving surgery followed by radiotherapy to the whole breast or, in the case of high-risk patients after mastectomy, radiotherapy to the chest wall with or without drainage areas. [1] Continuing information on the complete biological issues in breast cancer, coupled with the input of modern technology, helps to alleviate treatment-related morbidity and to improve the treatment outcome. [2] The radiotherapy techniques in the treatment of breast cancer vary in different institutions, but, in general, the issue of radiation dose delivery to the chest wall after total mastectomy or to the breast following breast conservation surgery remains complex. In the conventional breast irradiation technique, the beam arrangement consists of two opposing tangential glancing portals, [3] which allows acceptable coverage of the breast tissue while minimizing the dose to the adjacent critical structures (i.e., ipsilateral lung, contralateral breast, and heart). Physical or dynamic wedges are usually added to these tangential beams in order to compensate for the rapid changes in external contours and to improve the dose uniformity to the entire breast. The risk of contralateral breast cancer has been discussed in recent studies, [4],[5],[6] which emphasize the need for reduction of radiation dose to the contralateral breast using physical wedges, avoiding cerroband half beam blocks, and using asymmetric jaws and some form of intensity modulation. [4] Our earlier works had addressed the efficacy of three-dimensional computerized tomography (3D-CT)-based treatment planning and field shaping, as well as better dose conformity by applying multileaf collimator (MLC) optimized tangential beams using field-in-field techniques, [7],[8] in achieving better radiotherapy plans. Dose-related morbidity due to irradiation of heart tissue has been reported in a few studies earlier. [9],[10] Inversely planned intensity modulated radiotherapy (IMRT) has proved its efficacy in various sites, where there are constraints in dose delivery in general, and restricting minimum dose to critical structures in particular. One of our recent publications had addressed the issue of spatial specificity in dose delivery in the treatment of gastric cancers by applying IMRT plans, thereby achieving significant reduction in the dose to organs at risk (OARs), such as liver, kidney, and spinal cord. [11] Several single-institution studies and two randomized trials for breast cancer have reported that IMRT improves the dose homogeneity and decreases the acute skin toxicity as well as the dose to the contralateral breast compared with conventional tangential techniques with wedges. In these studies, most IMRT plans were created using the field-in-field (or forward-planned [FP] IMRT) technique. [12] Conformal electron irradiation for optimization has also been attempted. [13] The present work aims to compare the dose delivery parameters of the isocentric variable-angle multi-field IMRT plan vis-à-vis other simple plans with parallel opposed tangential fields in the treatment of breast cancer.


   Materials and Methods Top


Patients

The radiotherapy treatment data of 20 patients with left-sided breast cancer treated between October 2009 and March 2010 at the National Oncology Center, Muscat, Oman were selected. Ten patients who underwent mastectomy received a prescribed dose of 45 Gy to the chest wall in 20 fractions. The other 10 patients who had breast-conserving surgery received a prescribed dose of 50 Gy to the whole breast in 25 fractions. Both groups were treated with 3D-CRT. New treatment plans with IMRT were created.

Target volumes

The target volumes (the chest wall or the whole breast) and sensitive structures, such as the heart, ipsilateral lung, contralateral lung, and contralateral breast, were delineated in 5-mm-thick CT slices [Figure 1]. The field borders were clinically defined with radiopaque wires during simulation and also delineated according to the location of the tumor, extent of breast tissue, and adequate set-up margins. The field borders extended up to midline medially, lower border of clavicle superiorly, and laterally and inferiorly 2 cm beyond the palpable breast tissue. To prevent the inverse IMRT planning algorithm from delivering too high a dose to the skin, intact breast and chest wall PTVs were restricted to 5 and 3 mm under the skin surface, respectively, to exclude the build up region from the PTVs. [14]
Figure 1: Patient anatomy showing breast planning target volume (PTV) and OARs

Click here to view


Treatment planning

Computerized radiation treatment planning system (RTPS) Eclipse (version 6.5, M/s Varian Ag, Palo Alto, USA) was used for treatment planning. On the basis of the FiF-FP-IMRT plans created by combining two open fields with three to four segments in two tangential beam directions, the patients had received radiotherapy using 6-MV photon beams in Clinac 2300-CD (Varian Ag, USA), with millennium 120 MLC as tertiary collimator at 100 cm source to skin distance (SSD). Inversely planned IMRT (IP-IMRT) treatment plans were generated in the CT images of the already treated patients as 'study plans' with sliding window dynamic IMRT employing segmental multi-fields delivery, using Helios optimization software. Eight different beam directions were selected to create IP-IMRT isocentric plans and were inversely optimized. Pencil beam calculation algorithm was used for optimization. Final dose was calculated, with heterogeneity corrections, using Equivalent TAR algorithm after determining the leaf sequencing. The target doses and dose-volume limits restricted for critical structures are listed in [Table 1].
Table 1: Target doses and dose-volume constraints of the organs at risk


Click here to view


Comparison of treatment techniques

Plans of the three different treatment techniques were compared for evaluation of dosimetric parameters. Details of the beam arrangements and objectives of plans are described below:

3D-Conformal Plan

In this conventional planning technique, the beam arrangement consisted of two parallel opposing tangential beams ensuring the best possible coverage of the breast tissue and minimizing the dose to the adjacent critical structures (i.e., ipsilateral lung, contralateral breast, and heart). The "isocenter" of the treatment machine is positioned at the centre point of the midline joining two parallel opposing fields. Physical or dynamic wedges were then added to both tangential beams in order to improve the dose uniformity to the PTV, and to compensate for the rapid changes in external contours, as shown in [Figure 2] and [Figure 3]. Efforts were made to minimize volumes of heart and lung that unavoidably get included within the field borders.
Figure 2: Dose distribution for the two tangential fields without wedges

Click here to view
Figure 3: Dose distribution for the two tangential fields with wedges

Click here to view


Field-in-Field-Forward-planned-IMRT (FiF-FP-IMRT)

Two open tangential fields were created in this technique, according to the geometry defined during simulation to achieve uniform dose distribution to the breast volume (adequate coverage to the tumor bed), limiting the volume of the heart receiving a dose > 20 Gy not to exceed 5%, minimizing hot spot regions, and limiting dose to the ipsilateral lung and contralateral breast. The " isocenter" of the treatment machine is positioned at the same point as for the 3D-conformal plan. Initially, equal weights were assigned to the two open fields, and the corresponding dose distribution was calculated. By viewing the 95% dose cloud in a beam's eye view projection of the treatment fields, subfields were manually designed to boost the area not included in the dose cloud. The shape of each subfield was iteratively modified with aided visualization of 105% dose clouds in the beam's eye view. The number of subfields varied from three to four. Either 6- or 15-MV photons were selected for the subfields depending on separation of fields.

Inversely planned isocentric IMRT (IP-IMRT)

The IP-IMRT optimized plans were generated to achieve the same objectives described for the FiF-FP-IMRT plan. Eight different photon beam directions from 120 to 300, at intervals of 25 were chosen for this technique [Figure 4]. Beam directions at 120 to 300 remained identical to the initial tangential plans. The "isocenter" of the treatment machine is positioned at the same point as in the 3D-CRT and FiF-FP-IMRT plans. In addition to selection of PTVs 5 mm and 3 mm below the skin, a 5-mm bolus was added to the skin surface as per recommendations. [11]
Figure 4: Isocentric IP-IMRT plan with eight photon fields in breast treatments

Click here to view


Evaluation of plans

The treatment plans generated were compared objectively using the dose volume histograms (DVHs) for PTVs and different Organs at Risk (OARs) regions of interest. In the PTV, the values of V 105 , V 99, V 95 , V 90 , D mean , D max , homogeneity index (HI), conformity index (CI) were compared for all these three techniques. For lung OAR, the values of V mean , V 5 , V 20 ; for heart, V mean , V 5 , V 30 ; for ipsilateral lungs, V mean dose value, and V 5 , V 2 , V 1 , V 0.6 , V 0.01 doses for contralateral breast were evaluated and compared the dose to PTV.

The following parameters were used to evaluate the plans objectively:

  1. Relative volume of breast PTV receiving 105% of the prescription dose (V 105% ) (represent the extension of hot-spot regions within the breast).
  2. Mean (D mean ) and maximum dose (D max ) delivered to the target volume.
  3. Target volume receiving 90% and 99% of the dose, (V 90% and V 99% ).
  4. Homogeneity index (HI) in PTV defined by the relation

    HI = (D Max - D Min )/D Mean .....(1)
  5. Conformity index (CI) defined as per relation

    CI = (PTV RI /PTV) (PTV RI /V RI ) .....(2)

    (PTVRI is the 95% of the planning target volume, which should be covered by 95% of the prescribed dose)
  6. Relative volume of a given tissue receiving 20 Gy or 30 Gy (V 20 Gy or V 30 Gy , respectively) and the mean doses. V 20 Gy for the lungs (combined as one organ) and the V 30 Gy for the heart.
  7. Relative volume of the heart and lungs receiving low dose (5 Gy).
  8. Relative volume of the contralateral breast receiving dose (0.05, 0.6, 1, 2, and 5 Gy) and the mean dose.


For comparing each parameter, the statistical significance was calculated by P value analysis from Student's t-test.


   Results Top


Dosimetric characteristics

All the three plans, 3D-CRT, FiF-FP-IMRT, and IP-IMRT, achieved comparable good dose coverage, delivering prescribed dose more than 95% to > 95% of the breast PTV. In all these techniques, 105% of dose (hot regions) was observed in less than 5% of the target volume. Mean volume of PTV breast receiving 105% was 1.7% (range 0-6.8%) for IP-IMRT, 1.9% for FP-IMRT, and 3.7% for 3D-CRT. The results on estimated dosimetric parameters are shown in [Table 2].
Table 2: Comparison of average dosimetric characteristics for planning target volume for the 3D-CRT, FiF-FP-IMRT, and IP-IMRT plans


Click here to view


From [Table 2] it can be seen that the 3D-CRT and FiF-FP-IMRT plans had D max in the range 105-108% (mean 107%). For the IP-IMRT plan, D max ranged from 105-117% (mean 111%). The 3D-CRT and FiF-FP-IMRT plans provided comparable HI and CI. The IP-IMRT plan had a better conformity with the same HI. There was no statistical significance in D max between 3D-CRT and FiF-FP-IMRT.

[Figure 5]a-b show the average dose volume histograms (DVHs) for heart and lungs for the three treatment techniques of breast irradiation. It can be observed that the minimum and mean dose to the heart and lungs are much higher in IP-IMRT than in 3D-CRT and FiF-FP-IMRT. [Figure 6]a-b illustrate the DVHs for ipsilateral lung and contralateral breast. In these OARs also the minimum and mean doses are much higher for IP-IMRT plans. [Figure 7]a-b highlight the low-dose distribution compared for the 5 Gy dose level.
Figure 5: (a) Mean DVH for Heart for three plans, (b) Mean DVH for Lungs for three Plans

Click here to view
Figure 6: (a) Mean DVH for Ipsilateral lung, (b) Mean DVH for contralateral breast

Click here to view
Figure 7: (a) 5Gy dose distribution for the 3D-CRT and FiF-FP-IMRT plans, (b) 5Gy dose distribution for the IP-IMRT treatment plan

Click here to view


It is observed that the volumes receiving low dose (V 2Gy ) are much larger in the IP-IMRT technique.

The comparison of the average dosimetric parameters for OARs for the three treatment techniques are shown in [Table 3]. The 3D-CRT and FiF-FP-IMRT plans were equivalent in sparing critical organs. For the IP-IMRT planning technique, the average V 20 Gy for the lungs and the mean V 30 Gy for the heart were found to be less than that in the 3D-CRT and FiF-FP-IMRT plans. In addition, D mean for heart and lungs were higher for IP-IMRT than for FiF-FP-IMRT and 3D-CRT. The low-dose volumes (V 5 Gy ) in the heart and lungs were larger in the IP-IMRT than in the other techniques. The value of the mean dose to the ipsilateral lung was higher for IP-IMRT than the values for the FiF-FP-IMRT and 3D-CRT. The relative volume of contralateral breast receiving low doses (0.01, 0.6, 1, and 2Gy) [Table 4] was significantly lower for the FiF-FP-IMRT and 3D-CRT plans than for the IP-IMRT plan.
Table 3: Comparison of average dosimetric characteristics for organs at risk for 3D-CRT, FiF-FP-IMRT and IP-IMRT plans


Click here to view


Although the differences between 3D-CRT and FiF-FP-IMRT planning techniques were not statistically significant, a small increase in the dose to OARs was present in the 3D-CRT plans.


   Discussion Top


The present work was aimed to compare the planning and dose delivery efficiency among three techniques of radiotherapy to the chest wall/breast, namely 3-dimensional conformal radiotherapy (3D-CRT), field-in-field-forward planned-intensity modulated radiotherapy (FiF-FP-IMRT), and inverse planned-intensity modulated radiotherapy (IP-IMRT). [Table 1] shows that the FiF-FP-IMRT results are comparable to the 3D-CRT technique in terms of breast coverage, delivering 95% of the prescribed dose to > 95% of the breast PTV, the mean dose delivered to the breast PTV and the hot-spot regions. However, in the IP-IMRT plans, the mean maximum doses were more than 110% of the prescribed dose. The results of our study match with other similar studies. [14],[15] The authors of this study [14] have opined that the treatment plan with IMRT photons is inferior to 3D-CRT technique, because with 3D-CRT technique the mean radiation doses to the heart, contralateral breast, and lung doses are comparatively less for same PTV coverage and similar OAR shielding. They [14] applied 5-mm-thick bolus on chest wall in each photon beam inferior to match line. We have used a bolus for our IMRT study plans based on the recommendations from earlier work. [13] Efficacy of FIF technique versus plain tangential field is clearly brought out by Sasaoka and Futami. [16] The field-in-field technique significantly reduced the maximum dose, the volumes receiving > 107% of the prescription dose, and the HI for the PTV for dose evaluation compared with the tangential field technique. For each dosimetry of the OARs, excluding the contralateral breast, the field-in-field technique significantly reduced the maximum dose and the volumes receiving > 10, 30, and 50 Gy of the prescribed dose. The volume receiving < 1 Gy of the prescription dose for the contralateral breast was significantly decreased using the field-in-field technique. In addition, the dose distribution using the field-in-field technique in the target volume was less sensitive to the effects of breast motion during normal breathing. In the clinical outcome, the field-in-field technique significantly reduced the Radiation Therapy Oncology Group (RTOG) grade II acute skin toxicity compared with the tangential field technique (3.1 vs. 10.6%).

The 3D-CRT and FiF-FP-IMRT techniques required fewer monitor units (MU) to deliver a given dose, compared with IP-IMRT. In the IP-IMRT, the number of MUs is found to be about four times greater than in the FiF-FP-IMRT plans (1160 MU vs. 293 MU), and about three times greater than in the 3D-CRT plans (1160 MU vs. 443 MU). The afore-mentioned comparison also implies that the Full-size tableMUs for FiF-FP-IMRT tangent plans were less than those for 3D-CRT (443 MU and 293 MU, respectively) because of the use of wedges in the 3D-CRT.

Increased MU results in excess of machine and treatment times (for a dose rate of 300 MU/min), as well as higher machine leakage and total body stray radiation dose. The IP-IMRT for breast treatment is time consuming and requires advanced planning skills. The average planning time to generate the IP-IMRT plans was 45-60 min, whereas the 3D-CRT took about 20-30 min and the FiF-FP-IMRT 15-20 min. In contrast to the 3D-CRT and FiF-FP-IMRT plans, the IP-IMRT required verification of pretreatment patient-specific QA measurements. The additional QA time must be taken into account when considering the total workload per plan. Compared with the IP-IMRT and 3D-CRT plans, the higher-quality FiF-FP-IMRT plans are likely to be generated in a shorter time without requiring a high level of planning ability.

When we attempted to decrease dose to the OARs, the maximum dose increased to the hot-spot regions, which were small regions near the surface of the skin as shown in [Figure 8]. This may be because of the limitation of the dose optimization algorithm. In some IMRT studies, [13] doses up to 130% (60 Gy) were accepted in the breast PTV in attempt to minimize the dose delivered to the OARs. Furthermore, compared with IP-IMRT, the 3D-CRT and FIF-FP-IMRT plans resulted in a significantly smaller hot spot within the breast volume (V 105% ). This may have impact on improved cosmetic outcomes.
Figure 8: Islands of Hot spots with the IP-IMRT plans near the surface of the skin

Click here to view


The 3D-CRT and FiF-FP-IMRT plans provided almost the same HI and CI (P values: 0.50, 0.49, respectively), whereas the IP-IMRT plans had less HI, but better CI since this plan had more than two tangential fields [Table 2]. The 3D-CRT and FiF- FP-IMRT plans were also equivalent in sparing sensitive structures. It can also be seen that the average DVHs for the critical structures were nearly identical in the FP-IMRT and 3D-CRT plans [Figure 5] a-b.

Although the mean volume of the lung (both lungs combined as one organ) receiving 20 Gy (V 20 Gy ) and the mean volume of the heart receiving 30 Gy (V 30 Gy ) were lower for the IP-IMRT technique, both the 3D-CRT and FIF-FP-IMRT plans resulted in a lower mean volume of the lung and heart receiving low dose (5 Gy) Refer to [Table 2] and [Table 3]. Moreover, compared with IP-IMRT, the 3D-CRT and FIF-FP-IMRT plans resulted in a significantly smaller mean volume of the heart and lung. In addition, in the IP-IMRT, the value of the mean dose to the ipsilateral lung was higher than with the FiF-FP-IMRT and 3D-CRT [Figure 7]a-b. The FiF-IP-IMRT keeps the threshold of only 2% of heart volume receiving 30 Gy, which originates from the penumbral region of tangential fields, which is in comparison with the earlier reports [16] [Table 3].

More recently, many reports [14] have proposed a series of contralateral breast dose-volume thresholds (V 0.05 Gy , V 0.6 Gy , V 2.0 Gy , and the mean dose), which could serve as the maximum dose objectives in treatment planning to limit the risk of second malignancy. In our study, compliance with these dose volume metrics was significantly better with the FiF- FP-IMRT and 3D-CRT than with the IP-IMRT.

The IP-IMRT plans increased the total MUs, thereby increasing the volume of normal tissues exposed to the very low dose. It is not known whether this could result in a greater probability of radiation-induced complications and secondary malignancy. The decrease in the number of monitor units leads to the reduction in leakage dose. The selected patients in this study were treated with the 3D-CRT plans, which were generated by an expert physicist, and the typical planning time was 20-30 min. For the FiF-FP-IMRT, it took about 15-20 min. In contrast, the IP-IMRT plans required about 50-60 min. This time could increase considerably for someone with lesser experience. Moreover, the treatment delivery (including patient setup) required approximately 10-15 min for both the 3D-CRT and FP-IMRT and about 15-20 min for the IP-IMRT. Hot spots in the IP-IMRT plans have some implications in cosmetic outcomes. [17] However. this is not addressed in the present work.

The afore-mentioned facts objectively bring out the fact that the IP-IMRT for breast is not justified because of increased planning time, need for more MUs, advanced planning skills, increased D max value in PTV, excess dose to ipsilateral lung, contralateral breast, and islands of hot spots. Therefore, it is summarized that, compared with the 3D-CRT and IP-IMRT, the FiF-FP-IMRT [8] is a simple and efficient planning technique for breast irradiation. It provides dosimetric advantages, significantly reducing the size of the hot spot and minimally improving the coverage of the target volume. In addition, the FiF-FP-IMRT requires less planning time and is less dependent on the planner's skills.


   Conclusion Top


FiF-FP- IMRT is a simple and efficient planning technique for breast irradiation. Compared with the 3D-CRT and IP-IMRT, the FiF-FP-IMRT is efficient in reducing the hot-spot regions within the breast volume, which could result in better cosmetic outcomes. Moreover, implementation of this technique in clinical practice is straight forward, resulting in an overall reduction of doses for OARs. The planning time is reduced and the technique is user friendly.


   Acknowledgments Top


The authors acknowledge Director General, Royal Hospital for allowing the publication.

 
   References Top

1.Veronesi U, Marubini E, Marian L, Galimberti V, Luini A, Veronesi P, et al. Radiotherapy after breast conserving surgery in breast cancer: Long term result of randomized trial. Ann Oncol 2001;12:997-1003.  Back to cited text no. 1
    
2.Giantonio BJ, Catalano PJ, Meropol NJ, O'Dwyer PJ, Mitchell EP, Alberts SR, et al. National comprehensive cancer network (NCCN). Breast Cancer V. 2.2010. NCCN Clinical practice guidelines in oncology.   Back to cited text no. 2
    
3.Recht A. Breast cancer: Stages T 1 and T 2 . In: Gunderson LL, Tepper JE, editors. 2 nd ed. Clinical Radiation Oncology. Elsevier Publs: Churchill Livingstone; 2007. p. 1475-95.  Back to cited text no. 3
    
4.Unnithan J, Macklis RM. Contralateral breast cancer risk. Radiother Oncol 2001;60:239-46.  Back to cited text no. 4
    
5.Yadav BS, Sharma SC, Patel FD, Ghoshal S, Kapoor RK. Second primary in the contra-lateral breast after treatment of breast cancer. Radiother Oncol 2008;28:171-6.  Back to cited text no. 5
    
6.Kelly CA, Wang XY, Chu JC, Hartsell WF. Dose to contralateral breast: A comparison of four primary breast irradiation techniques. Int J Radiat Oncol Biol Phys 1996;34:727-32.  Back to cited text no. 6
    
7.Khalifa A, Somangali SS, Satyapal N. Comparative study of two-dimensional and three-dimensional planning for external irradiation after breast conservation surgery. New Egypt J Med 2006;34:63-70.  Back to cited text no. 7
    
8.Murthy KK, Sivakumar SS, Davis CA, Ravichandran R, El Ghamrawy K. Optimization of dose distribution with multi-leaf collimator using field-in-field technique for parallel opposing tangential beams of breast cancers. J Med Phys 2008;33:60-3.  Back to cited text no. 8
[PUBMED]  Medknow Journal  
9.Lind PA, Pagnanelli R, Marks LB, Borges Neto S, Hu C, Zhou SM, et al. Myocardial perfusion changes in patients irradiated for left-sided breast cancer and correlation with coronary artery distribution. Int J Radiat Oncol Biol Phys 2003;55:914-20.  Back to cited text no. 9
    
10.Correa CR, Das IJ, Litt HI, Ferrari V, Hwang W, Solin LJ, et al. Association between tangential beam treatment parameters and cardiac abnormalities after definitive radiation treatment for left-sided breast cancer. Int J Radiat Oncol Biol Phys 2008;72:508-16.  Back to cited text no. 10
    
11.Murthy K, Shukeili K, Kumar S, Davis C, Chandran R, Namrata S. Evaluation of dose coverage to target volume and normal tissue sparing in the adjuvant radiotherapy of gastric cancers: 3D-CRT compared with dynamic IMRT. Biomed Imaging Interv J 2010;6:e29.  Back to cited text no. 11
    
12.Wilfried DN, Yan W, Gary E. Practical IMRT Planning. In: Thomas B Rupert Schmidt U, Wilfried DN, David EW, editors. Chapter 5. Image-Guided IMRT. New York: Springer Pubs; 2006. p. 48-50.  Back to cited text no. 12
    
13.Descovich M, Fowble B, Bevan A, Schechter N, Park C, Xia P. Comparison Between Hybrid Direct Aperture Optimized Intensity-Modulated Radiotherapy and Forward Planning Intensity-Modulated Radiotherapy for Whole Breast Irradiation. Int J Radiat Oncol Biol Phys 2010;76:91-9.  Back to cited text no. 13
    
14.van der Laan HP, Korevaar EW, Dolsma WV, Maduro JH, Langendijk JA. Minimising contralateral breast dose in post-mastectomy intensity-modulated radiotherapy by incorporating conformal electron irradiation. Radiother Oncol 2010;93:235-40.  Back to cited text no. 14
    
15.Sasaoka M, Futami T. Dosimetric evaluation of whole breast radiotherapy using field-in-field technique in early-stage breast cancer. Int J Clin Oncol 2011;16:250-6.  Back to cited text no. 15
    
16.Landau D, Adams EJ, Webb S, Ross G. Cardiac avoidance in breast radiotherapy: A comparison of simple shielding techniques with intensity-modulated radiotherapy. Radiother Oncol 2001;60:247-55.  Back to cited text no. 16
    
17.Taylor ME, Perez CA, Halverson KJ, Kuske RR, Philphott GW, Garcia DM. Factors influencing cosmetics results after conservation therapy for breast cancer. Int J Radiat Oncol Biol Phys 1995;31:753-64.  Back to cited text no. 17
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]


This article has been cited by
1 Improvement of dose distribution with irregular surface compensator in whole breast radiotherapy
Hideki, F., Nao, K., Hiroyuki, H., Hiroshi, K., Haruyuki, F.
Journal of Medical Physics. 2013; 38(3): 115-119
[Pubmed]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (2,339 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
   Acknowledgments
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2750    
    Printed80    
    Emailed3    
    PDF Downloaded101    
    Comments [Add]    
    Cited by others 1    

Recommend this journal