The effects of cutouts on output, mean energy and percentage depth dose of 12 and 14 MeV electrons


1 Young Researchers Club, Science and Research Branch, Islamic Azad University, Tehran; Department of Medical Physics, Imam Hosein Hospital, Shahid Beheshti Medical University, Tehran, Iran
2 Department of Medical Physics, Imam Hosein Hospital, Shahid Beheshti Medical University, Tehran, Iran
3 Department of Medical Radiation, Islamic Azad University, Science and Research Branch, Tehran, Iran

Correspondence Address:
Navid Khaledy
Department of Radiotherapy, Imam Hossain Hospital, Madani St., Tehran
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.89970

Rights and Permissions

Electron field-shaping cerrobend cutouts on the linear accelerator applicator have some effects on the output and percentage depth dose. These effects which arise from the lateral scatter nonequilibrium are particularly evident in higher energies and in cutouts with smaller radius. Dose measurements for circular, square, and triangular cutouts as well as open field was performed in a 10 × 10 cm applicator, using plane parallel type ion chamber with a 100 cm source surface distance. The Percentage Depth Doses curves were drawn and the outputs were measured for each of these cutouts. The output factors, normalized to open 10 × 10 cm field, varied between 0.891 and 0.996 depending on the energy, cutout shape, and cavity area. With the use of cutouts, R 100 shifted toward the surface. The shifts ranged from 9 to 0 mm and from 13 to 0 mm for 12 and 14 MeV, respectively, depending on the shape and cavity area. For R 90 , R 80 , and R 50 the ranges for observed shifts narrowed down and practically no shifts were observed for R 20 . We present these changes in the form of predictive formulas, which would be useful in clinical applications.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed3439    
    Printed116    
    Emailed2    
    PDF Downloaded139    
    Comments [Add]    
    Cited by others 2    

Recommend this journal