TECHNICAL NOTE
Year : 2011  |  Volume : 36  |  Issue : 1  |  Page : 46-50
 

Development of departmental standard for traceability of measured activity for I-131 therapy capsules used in nuclear medicine


Medical Physics Unit, National Oncology Center, Royal Hospital, Oman

Date of Submission16-May-2010
Date of Decision07-Jul-2010
Date of Acceptance13-Sep-2010
Date of Web Publication12-Jan-2011

Correspondence Address:
Ramamoorthy Ravichandran
Chief Medical Physicist, National Oncology Center, Royal Hospital, PB1331, PC 111, Muscat
Oman
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.75472

Rights and Permissions

 

   Abstract 

International Basic Safety Standards (International Atomic Energy Agency, IAEA) provide guidance levels for diagnostic procedures in nuclear medicine indicating the maximum usual activity for various diagnostic tests in terms of activities of injected radioactive formulations. An accuracy of ± 10% in the activities of administered radio-pharmaceuticals is being recommended, for expected outcome in diagnostic and therapeutic nuclear medicine procedures. It is recommended that the long-term stability of isotope calibrators used in nuclear medicine is to be checked periodically for their performance using a long-lived check source, such as Cs-137, of suitable activity. In view of the un-availability of such a radioactive source, we tried to develop methods to maintain traceability of these instruments, for certifying measured activities for human use. Two re-entrant chambers [(HDR 1000 and Selectron Source Dosimetry System (SSDS)] with I-125 and Ir-192 calibration factors in the Department of Radiotherapy were used to measure Iodine-131 (I-131) therapy capsules to establish traceability to Mark V isotope calibrator of the Department of Nuclear Medicine. Special nylon jigs were fabricated to keep I-131 capsule holder in position. Measured activities in all the chambers showed good agreement. The accuracy of SSDS chamber in measuring Ir-192 activities in the last 5 years was within 0.5%, validating its role as departmental standard for measuring activity. The above method is adopted because mean energies of I-131 and Ir-192 are comparable.


Keywords: Activity calibration, iodine-131, re-entrant chambers, standards


How to cite this article:
Ravichandran R, Binukumar J P. Development of departmental standard for traceability of measured activity for I-131 therapy capsules used in nuclear medicine. J Med Phys 2011;36:46-50

How to cite this URL:
Ravichandran R, Binukumar J P. Development of departmental standard for traceability of measured activity for I-131 therapy capsules used in nuclear medicine. J Med Phys [serial online] 2011 [cited 2019 Jun 25];36:46-50. Available from: http://www.jmp.org.in/text.asp?2011/36/1/46/75472



   Introduction Top


International Basic Safety Standards (International Atomic Energy Agency, IAEA) [1] provide guidance levels for diagnostic procedures in nuclear medicine, indicating the maximum usual activity for various diagnostic tests in terms of injected activities of radioactive formulations. A study [2] correlating administered radiopharmaceutical activity with imaging outcome, indicates an accuracy of ± 10% in injected activity. A reasonable level of accuracy in administered activities is expected also in the therapeutic use of radioisotopes, [3] because there are other variables such as uptake in organs, biological half-life, dose response in heterogenous tumors, etc. Large-volume, calibrated, pressurized re-entrant type of ion chambers with 4π geometry are being used for this purpose along with suitable electrometers. Methods for standardization of 'activity of brachytherapy sources' are outlined in literature. [4],[5],[6],[7] Some of the important characteristics of these chambers are 1) linearity, 2) re-producibility and 3) long-term stability. For quality assurance of functioning of departmental isotope calibrators, periodic checking with long-lived 137 Cs check sources is recommended. [8],[9] In the Department of Nuclear Medicine we do not have such a check source, therefore, there is an urgent need to develop methods to maintain traceability to these instruments, for certifying measured activities for human use. This report highlights the method we adopted to achieve the above objective.


   Materials and Methods Top


Isotope calibrators

The Department of Nuclear Medicine has 3 re-entrant chambers with digital reading system-2-type Mark V (CalRad, Nuclear Associates, USA) for assay of diagnostic formulations, and 1 calibrator (Veenstra Instruments, USA) routinely used for Iodine-131 therapy capsule assay. The Mark V calibrations are traceable to cardinal imaging [(Accredited Dosimetry Calibration Laboratory, (ADCL), USA ]. Different isotopes can be selected by a function key on LCD control unit; and with multiplication factor 1.00, the activities are directly displayed for respective isotopes. Iodine-131 (I-131) therapy capsules are received from GE HealthCare Buchler, Amersham, UK. The activity of capsules is checked immediately upon arrival of the consignments in the department and compared with the certified activity provided by the supplier.

Activity measurement facility to establish traceability

The Department of Radiotherapy has 2 re-entrant chambers [HDR 1000, Standard Imaging, USA, and Selectron Source Dosimetry System (SSDS) chamber, Nucletron BV, Netherlands] for measurement of activities of high-dose-rate (HDR) remote-controlled after-loading Iridium-192 sources used for treatment with microSelectron (Nucletron). [Table 1] compares the specifications of all the three isotope calibrators, The collection efficiency stated by the manufacturers of these chambers is, Aion= 0.9996 and 0.9960 for HDR 1000 chamber and SSDS chamber respectively; and stated uncertainty in activity measurements for both the re-entrant chambers is within ± 2%.
Table 1 :Specifications of isotope-calibrators for estimation of I-131 activity


Click here to view


Calibration certificates from Accredited Dosimetry Calibration Laboratories (ADCL), provided along with these instruments [Table 1], represent multiplication factors on the displayed units, to arrive at true activity of the radioisotopes during routine measurements . The overall uncertainty in the calibration as expressed in calibration certificates is ± 10%. From August 2004 till February 2010, results of calibration in measuring activities of high-intensity Ir-192 sources of 12 Iridium-192 high-dose-rate brachytherapy sources with SSDS were in agreement within 0.4% ± 0.6% of stated activity. The calibration factor for SSDS (provided by PTW, Frieburg) to measure apparent activity of Ir-192 remained constant for more than 5 years (2.31x10 8 Ci/A in September,2004 and 2.316 x 10 8 Ci/A in September 2009). The calibration coefficient provided for HDR 1000 re-entrant chamber is 2.36 x10 11 μGy/A for beam quality I-125 (6711). The sensitivity of this chamber provided by the manufacturer is 8.7 nA/Ci. In a recent report, [10] the long-term stability of HDR 1000 chamber over a period of more than 3 years was found to be within ± 0.5%. In the above circumstances, it was decided to refer activity measurements from Mark V chamber, traceable to SSDS and HDR 1000 re-entrant chambers.

Measurement technique for establishing calibration traceability

In the absence of locally available Cs-137 standard check source, the department used the imported I-131 capsules received for thyroid treatments, to be measured simultaneously in Mark V, HDR 1000 and SSDS re-entrant chambers. Special nylon jigs were locally fabricated to fit HDR 1000 and SSDS chambers. The jigs are able to hold the Perspex tube provided for oral administration of the I-131 capsule inside the well chamber at exact position of 100% response (5.2 cm from the bottom). This had been locally checked by separate measurements. [Figure 1] shows the Mark V calibrator of the Department of Nuclear Medicine. [Figure 2] and [Figure 3] show the design of the nylon jigs, along with the measurement geometry. In [Figure 3], both chambers can be seen in use for measuring 'ion currents in nA' with same I-131 capsule sequentially.
Figure 1 :Mark V Isotope calibrator in the Department of Nuclear Medicine used for calibration of I-131 therapy capsules. L-Bench for keeping the original container of capsules and handling area are seen

Click here to view
Figure 2 :A nylon jig to hold the I-131 Perspex holder, along with the geometry of I-131 capsule for measurement inside the re-entrant HDR 1000 ion chamber

Click here to view
Figure 3 :Premier - 3000 and Unidos Electrometers, along with their respective HDR 1000 and SDS re-entrant chambers. The I-131 source container, which contains I-131 capsule, can also be seen. The Perspex holder to carry I-131 for inserting into the calibration ion chambers is also seen

Click here to view


Calculation of activity of capsules

With Mark V chamber , activity is directly read in MBq of GBq with keyboard setting selected for I-131 isotope. For HDR 1000 chamber and SSDS chambers, the readings in 'nA' are converted into 'mCi' by multiplying with a factor 2.36 Χ 10 2 mCi/nA and 2.31 Χ 10 2 mCi/nA x {gamma ray constants ratio of Ir-192 and I-131 (4.60/2.20)}, respectively. As apparent calibration factor of Ir-192 is not provided in SI units, calculation of activity is done in 'mCi' and converted to MBq subsequently.


   Results Top


The measured levels of activity of received consignments by Mark V Isotope calibrator during the last 5 years were within a limit of 10 % in 89.1% for MIBG (n= 46) and 63.3% for Ga-67 (n= 109). [Table 2] shows distribution of measured activities of I-131 capsules in the past. It was found that 89.7% of the capsules had activities within 10% of certificate values (stated activity on reference date). A separate analysis of measured values for I-131 capsules referred against GE HealthCare Buchler Amersham-stated activities on dispatch date, showed an agreement within - 3.0% ± 4.0% (n= 175). [Table 3] gives the activities measured by Mark-V isotope calibrator (A) compared with HDR 1000 chamber (B) and SSDS (C) measurements. The percentage deviations calculated against stated activity (on reference date) corrected for measurement date, are seen in the last three columns. Good agreement (within 2% in 4/12 measurements, within 5% in 5/12 measurements and within 10% in remaining 3/12 measurements) is seen between Mark-V and other two calibrators. Therapy level activities in GBq quantities (> 4000 MBq) have shown better agreement in measured values (6 out of 12) compared to diagnostic level < 230 MBq (4 out of 12) I-131 activities. Measured readings for estimation of activity in all three re-entrant chambers were average of more than 10 observations that had standard deviation within 0.1%.
Table 2 :Measured activity of I-131 capsules


Click here to view
Table 3 :Activity measured in three isotope calibrators


Click here to view



   Discussion Top


The above presentation has brought to light a method for validating the calibration of activity of I-131capsules used for therapeutic administrations, in the absence of a departmental check source for nuclear medicine isotope calibrators. Clinical requirement for estimation of activity with sealed sources such as Ir-192 for remote after-loading applications is more stringent, because the therapy-planning systems works out dose distributions based on the 'measured activity input' at the time of each source loadings. Therefore, the data of measured activities of I-131 correlated to these re-entrant chambers will definitely ensure traceability in the institution.

Our results of all measurements have shown assayed activities to be always higher than stated values [Table 1],[Table 2],[Table 3], because usually the manufacturers provide slightly higher activities for the use of patients (to avoid cancellations due to less activities than prescribed). An earlier report, [9] has indicated differences in activities at the department level, when compared to values quoted by manufacturers, and also various earlier reports have emphasized need for accuracy of the department calibrators.

The I-125 calibration factor is provided by Standard Imaging (ADCL) based on the TG 43 4 recommendations of American Association of Physicists in Medicine (AAPM). This expresses 'strength U', the definition includes specific gamma ray constant of the isotope. The calibration factor provided for HDR 1000 re-entrant chamber, viz., 2.36 x 10 11 μGy/A, actually refers to 'air kerma strength' U as in TG-43 report. Taking Gk = 0.0358 μGy.m 2 . MBq -1 .h -1 (Ref. TG 43); and 37 MBq/mCi , the conversion factor for HDR 1000 to measure I-131 activity in mCi works out to 2.59 x 10 2 mCi/nA { (1.78 x 10 2 mCi/nA) (gamma ray constants ratio of I-131 and I-125, 2.20/1.51). In all our measurements, we have applied numerically 2.36 x 10 2 mCi/nA for HDR 1000 calibrator, which is 9.0% less than the calculated value of 2.59 x 10 2 mCi/nA. It is thought that this may be empirically due to difference because of low-energy gamma 31.4 KeV for I-125 compared to 364 KeV for I-131, which may be due to difference in interactions in the chamber wall materials.

The uncertainty in calibration [Table 1] describes the dispersion of the measured values of activity. Type A uncertainties are usually evaluated by statistical methods, and type B uncertainties are based on scientific judgements of non-statistical values, usually influenced by application of correction factors or physical data obtained from literature. Type B uncertainties can be derived by estimating the limit beyond which the value of the factor is not going to lie. Based on similar estimates, an earlier report, [11] indicates an RMS error of 0.11% (n = 100) for SSDS isotope calibrators for measurements of HDR Ir-192 sources. The present activity estimates lie well within acceptable limits of uncertainties.

The results of this work have clearly brought out that using a selected number of I-131 sources as transfer check source, the response of two other calibrators could be 'standardized' as reference instruments at institutional level. This methodology could be adopted in many institutions, as most of the radiotherapy centers have high-dose-rate after-loading machines and a brachytherapy calibrator for activity calibration. Weekly measurements in these Ir-192 calibrators are being undertaken for quality assurance of high dose rate brachy machines, validating them as departmental standards.

Therefore, this report may indicate a method to routinely check nuclear medicine calibrators, as mean energy of Ir-192 is around 370 KeV, which is comparable to 364 KeV of I-131.


   Acknowledgements Top


The authors thank Director General, Royal Hospital and Head, Department of Nuclear Medicine, Royal Hospital, for kind permission for publication of this material. Engineering Department, Royal Hospital, is gratefully acknowledged for fabricating the special nylon jigs used in this study.

 
   References Top

1.International Basic Safety Standards for Protection against Ionizing Radiation and for the Safety of Radiation Sources. IAEA Safety Series No. 115, Vienna: IAEA; 1996.  Back to cited text no. 1
    
2.Robinson CN, Van Aswegen A, Julious SA, Nunan TO, Thomson WH, Tindale WB, et al. The relationship between administered radiopharmaceutical activity in myocardial perfusion scintigraphy and imaging outcome. Euro J Nucl Med Mol Imag 2008;35:329-35.  Back to cited text no. 2
    
3.Society of Nuclear Medicine Procedure Guideline for Therapy of Thyroid Disease with Iodine-131 (Sodium Iodide) Version 1.0. SNM Document, USA Approved Feb 2002.  Back to cited text no. 3
    
4.Goetsch SJ, Attix FH, Pearson DW, Thomadsen BR. Calibration of Ir-192 high dose afterloading systems. Med Phys 1991;18:462-7.  Back to cited text no. 4
[PUBMED]    
5.Goetsch SJ, Attix FH, DeWerd LA, Thomadsen BR. A new re-entrant ionization chamber for the calibration of Ir-192 high dose rate sources. Int J Radiat Oncol Biol Phys 1992;24:167-70.  Back to cited text no. 5
[PUBMED]  [FULLTEXT]  
6.American Association of Physicists in Medicine. Remote afterloading technology. Report of Task Group 41. AAPM Report No.41, New York: American Institute of Physics; 1993.  Back to cited text no. 6
    
7.Nath R, Anderson LL, Luxton G, Weaver KA, Williamson JF, Meigooni AS. Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No.43. American Association of Physicists in Medicine. Med Phys 1995;22:209-34.  Back to cited text no. 7
    
8.Elliott AT. Quality Control. In: Sharp PF, Gemmell HG, Smith FW, editors. Practical Nuclear Medicine. Oxford: IRL Press; 1989.  Back to cited text no. 8
    
9.Hare DL, Hendee WR, Whitney WP, Chaney EL. Accuracy of Well Ionization Chamber Isotope Calibrators. J Nucl Med 1974;15:1138-41.  Back to cited text no. 9
[PUBMED]  [FULLTEXT]  
10.Vandana S, Sharma SD. Long term stability of a well type ionization chamber used in calibration of high dose rate brachytherapy sources. J Med Phys 2010;35:100-3.  Back to cited text no. 10
[PUBMED]  Medknow Journal  
11.Baltas D, Geramani K, Ioannidis GT, Hierholz K, Rogge B, Kolotas C, et al. Comparison of calibration procedures for 192-Ir high dose rate brachytherapy sources. Int J Radiat Oncol Biol Phys 1999;43:653-61.  Back to cited text no. 11
[PUBMED]  [FULLTEXT]  


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 New technique for effective dose estimation using Monte Carlo simulation for the patients undergoing radioiodine therapy
Jokic, V.S., Kocovska, M.Z.
Measurement: Journal of the International Measurement Confederation. 2013; 46(1): 795-802
[Pubmed]



 

Top
Print this article  Email this article
  

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,105 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
    Introduction
    Materials and Me...
    Results
    Discussion
    Acknowledgements
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed3684    
    Printed108    
    Emailed0    
    PDF Downloaded70    
    Comments [Add]    
    Cited by others 1    

Recommend this journal