ORIGINAL ARTICLE
Year : 2007  |  Volume : 32  |  Issue : 4  |  Page : 175-184
 

Computer simulations and models for the performance characteristics of spectrally equivalent X-ray beams in medical diagnostic radiology


Department of Physics, Obafemi Awolowo University 220005, Ile-Ife Osun, Nigeria

Correspondence Address:
Akintunde A Okunade
Department of Physics, Obafemi Awolowo University 220005, Ile-Ife Osun
Nigeria
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.37483

Rights and Permissions

 

   Abstract 

In order to achieve uniformity in radiological imaging, it is recommended that the concept of equivalence in shape (quality) and size (quantity) of clinical X-ray beams should be used for carrying out the comparative evaluation of image and patient dose. When used under the same irradiation geometry, X-ray beams that are strictly or relatively equivalent in terms of shape and size will produce identical or relatively identical image quality and patient dose. Simple mathematical models and software program EQSPECT.FOR were developed for the comparative evaluation of the performance characteristics in terms of contrast (C), contrast-to-noise ratio (CNR) and figure-of-merit (FOM = CNR 2 /DOSE) for spectrally equivalent beams transmitted through filter materials referred to as conventional and k-edged. At the same value of operating potential (kVp), results show that spectrally equivalent beam transmitted through conventional filter with higher atomic number ( Z -value) in comparison with that transmitted through conventional filter with lower Z -value resulted in the same value of C and FOM. However, in comparison with the spectrally equivalent beam transmitted through filter of lower Z -value, the beam through filter of higher Z -value produced higher value of CNR and DOSE at equal tube loading (mAs) and kVp. Under the condition of equivalence of spectrum, at scaled (or reduced) tube loading and same kVp, filter materials of higher Z -values can produce the same values of C, CNR, DOSE and FOM as filter materials of lower Z -value. Unlike the case of comparison of spectrally equivalent beam transmitted through one conventional filter and that through another conventional filter, it is not possible to derive simple mathematical formulations for the relative performance of spectrally equivalent beam transmitted through a given conventional filter material and that through k-edge filter material.


Keywords: Contrast, elemental filters, medical diagnostic radiology


How to cite this article:
Okunade AA. Computer simulations and models for the performance characteristics of spectrally equivalent X-ray beams in medical diagnostic radiology. J Med Phys 2007;32:175-84

How to cite this URL:
Okunade AA. Computer simulations and models for the performance characteristics of spectrally equivalent X-ray beams in medical diagnostic radiology. J Med Phys [serial online] 2007 [cited 2020 Jul 8];32:175-84. Available from: http://www.jmp.org.in/text.asp?2007/32/4/175/37483


The use of tube voltage (peak kilovoltage, kVp), effective energy, homogeneity coefficient and half-value layer (HVL) as measures of beam quality and quantity in medical diagnostic radiology continues to produce diversities in results obtained for image quality and patient dose. As a result of this, the use of equivalent spectral as a measure of beam quality and as a tool for the standardization of medical X-ray imaging has been reported. [1],[2] Among other factors, the quality and quantity of X-ray beams for medical diagnostic purpose depend on filter material and filter thickness. The standardization of beam quality can be achieved by providing a ranking for filter materials and filter thicknesses when the technology of X-ray tube is the same. By careful choice of thickness, two different elemental filters can be made to transmit beams that are simultaneously of the same/close shape (quantity) and size (quality). [3],[4],[5],[6] When used under the same irradiation condition, beams that are strictly equivalent or relatively equivalent in terms of shape and size will produce identical or relatively identical image quality and patient dose. The results of theoretical and experimental approaches to the comparative ranking of the transmissions of filter materials on the basis of closeness of hardening (size) and attenuation (quantity) properties have been reported earlier in literature. [7],[8],[9] Due to the differences in attenuation and absorption properties, two different elemental filters that can independently produce beams of exactly equal shape and size do not exist.

In this paper, computer simulations were carried out to obtain spectrally equivalent X-ray beams using elemental filters - which include those referred to as conventional, such as aluminum and copper; and k-edge, such as gadolinium and tungsten. Analytical models similar to those earlier reported [6] for quantitative parameters were derived for the evaluation of differences in parameters for image quality, such as contrast-to-noise ratio (CNR) and figure-of-merit (FOM) for filter materials that transmit spectrally equivalent beams. For the purpose of clinical assessment of the performance characteristics of spectrally equivalent beams, iterative computer software EQSPECT.FOR was developed for the evaluation of these models. Also, by least square methods, parameters which can be used for the evaluation of thicknesses of other filter materials that are 'spectrally' equivalent to a specified thickness of copper are presented. The definitions of some of the terms used in the computer simulations and analytical formulations that were derived are presented in [Table - 1].

Theoretical formulations

In this work, T (t m ) of filter material m at 'spectrally' equivalent thickness of t m is given by



C is defined as



E a and E b were calculated respectively using equations of the forms [10]



CNR and FOM were respectively calculated as



and



The absorbed dose in patient (water phantom of thickness t w centimeters) was determined using equation of the form [11]



where ε(t w ) could be obtained using an equation of the form



Details of the derivation of Eq. (7) can be found in the work reported by Okunade. [11]

From the earlier work reported, [6] for a given filter material m in comparison with aluminum filter at spectrally equivalent thickness,



and



The values of λm resulting in best fit to Eq. (9) are reported in [Table - 1] of Okunade [6] for filter materials m with atomic numbers ranging from 12 to 39 and kVp ranging between 50 and 140 when reference filter is assumed to be aluminum.

From the simulated data and curve-fitting exercises carried out in this work,



Calculations carried out using copper as reference filter in comparison with filter material m resulted in equation of the form



where Δ= λm t cu

The values of λm resulting in best fit to Eq. (11) are reported in [Table - 2] of this text. Thus from Eqs. (5), (9) and (10), for spectrally equivalent beams emanating from filter material m and aluminum filter at 'spectrally' equivalent thicknesses respectively, FOM is of the form



and the CNR for beam transmitted through filter material m in comparison with that through a reference filter (aluminum) is of the form



In order to obtain spectral that is exactly equal in shape and size (or that will produce exactly the same values of transmittance, T, contrast, C, dose, DOSE, contrast-to-noise ratio, CNR, and figure-of-merit, FOM), the spectral data generated from filter material, say copper, in comparison with those from aluminum at 'spectrally' equivalent thickness have to be scaled across the entire energy spectrum by using the factor α. Mathematically, for this, we can write,



Note that the values of number of photons (per energy bin E i ) transmitted by a filter with higher Z -value are greater than those transmitted by filter of lower Z -values when the tube is operated at the same kVp and tube loading. The use of this scaled spectral results in models of the form



The thickness of a given material, t eq , that is 'spectrally' equivalent to a given aluminum thickness t Al is of the form [6]



The values of ωn resulting in best fits to Eq. (16) when Z -value ranges between 12 and 39 are obtainable from [Table - 5] of Okunade. [6] When Z -values are 12, 13 and 14, the value of the thickness (t eq ) of other filter material that is 'spectrally' equivalent to copper thickness t cu can be obtained from equation of the form



When Z -values are 23, 26, 28, 30, 32 and 39, the value of the thickness (t eq ) of other filter material that is 'spectrally' equivalent to copper thickness t cu can be obtained from equation of the form



The values of ωn and αn resulting in best fits to Eqs. (17) and (18) are shown in [Table - 3] of this text.


   Methods Top


Simulations of beams of the same/close shape and size, contrast, contrast-to-noise ratio, dose and figure-of-merit

In order to simulate X-ray beams of the same/close shape and size ( spectrally equivalent beams), calculations were carried out for the matching of hardening and attenuation from aluminum and copper filters (reference filters) and other filter materials that are different from aluminum and copper. The matching exercises involved the use of the numerical algorithms earlier reported by Jennings. [8] The thicknesses of aluminum filter used were between 1 and 6 mm at 1 mm increments while those of copper were between 0.05 and 0.3 mm at 0.05 mm increments. The incident X-ray spectrals on the filter materials were determined by using polynomial functions earlier reported by Boone and Seibert. [12] Unfiltered tungsten anode spectra φ0 (E i ) in units of photons/mm 2 at energy E i in keV were calculated for values of kVp ranging between 50 and 140 in 10 kVp steps.

The transmission, contrast, contrast-to-noise ratio, figure-of-merit and dose were calculated using equations presented in the theory section above. A total of 10 values of kVp, 8 alternative conventional filter materials (other than aluminum/copper) and 6 reference filter thicknesses of aluminum/copper were considered. Simulations were carried out for the evaluation of T, C, CNR, FOM and DOSE for water phantom of thicknesses ranging between 5 and 30 cm at 5 cm increments. Also considered in these simulations are three contrast media - namely, calcium, barium and iodine; and three intensifying screens, which include CaWO 4 , CsI and Gd 2 O 2 S. By similar simulation exercise, the k-edge filter materials investigated in comparison with aluminum/copper filters include lanthanium, gadolinium, holmium, thulium and tungsten. The interaction data of photons for water (patient phantom) and those for elemental filters that were considered were taken from McMaster et al ., [13] while those for the contrast media and intensifying screens were from Hubbell and Seltzer. [14] All the models reported in this work were fitted using the curve-fit least-square routines implemented in the computer package GRAF4WIN. [15] [Figure - 1] shows the schematic diagram for the implementation of the computer simulations.

Computer code EQSPECT.FOR

A FORTRAN source codes EQSPECT.FOR was developed for the execution of the simulations of same/close shape and size of beams, C, CNR, FOM and DOSE. In these computer codes, the algorithms reported [12] for the simulations of energy distribution of X-ray beams permit the specification of values of kVp other than those of multiples of 10 kV (say 86 kV). The interpolations for values of lm at values of kVp other than multiples of 10 kVp were carried out using cubic spline computer routines (named DEFSAL and DEFSCU) that were embedded in EQSPECT.FOR. A brief description of the main steps in the execution of EQSPECT.FOR is presented in [Table - 4].


   Results Top


[Figure - 2] shows the results of comparison of T, CNR, [CNR] 2 , C, DOSE and FOM for beams transmitted through aluminum/copper and selected alternative filter materials at 'spectrally' equivalent thicknesses. Specifically, for peak voltage of 86 kV, [Figure - 3] shows the results of comparison of the spectra transmitted through 3.7 mm Al and 9.3 mm Al and those through the respective spectrally equivalent thicknesses of copper, 0.1 mm Cu and 0.25 mm Cu, at equal tube loading. [Table - 5] shows the results of comparison of performance characteristics of other filter materials relative to copper filter. For the purpose of comparison with the work of Nagel, [9] these results are for 0.2 mm Cu filter, imaging of patient phantom of thickness 20 cm and contrast simulated with iodine (10 mg/cm 2 ) and Gd 2 O 2 S intensifying screen (80 mg/cm 2 ). [Figure - 4] shows the comparison of spectral shape and size obtained by forcing the same shape and size for aluminum/copper and gadolinium filters using the algorithms reported by Jennings. [8] The results of the comparison of transmission, contrast, CNR, [CNR 2 ] DOSE and FOM obtained by the matching of spectral shape and size of beams transmitted by aluminum/copper and lanthanium are shown in [Figure - 5]. To further provide verification of the validity of the formulations (Eqs. 8-15) presented in this work, the results of comparison carried out for 0.088 mm Gd filter and 0.097 mm Cu filter using EQSPECT.FOR are presented in [Table - 6].


   Discussion Top


The value of CNR for the beam filtered by 0.1 mm Cu exceeds that of the beam filtered by 3.7 mm Al by 7.0% for the imaging of a 20-cm patient phantom with the contrast simulated with iodine (12 mg/cm 2 ) and CsI intensifying screen (60 mg/cm 2 ). For this same imaging, the value of CNR for beam filtered by 0.25 mm Cu exceeds that for beam filtered by 9.3 mm Al by 19.0%. For the case of 3.7 mm Al and 0.10 mm Cu filtrations, the values of T Cu , DOSE Cu and [CNR Cu ] 2 exceed those of T Al , DOSE Al and [CNR Al ] 2 by 14.5% respectively while C Cu ~ C Al and FOM Cu ~ FOM Al . This is in good agreement with the result from earlier work, [8] which reported a surplus of 14.8% for T Cu over T Al . At spectrally equivalent thicknesses, the difference in exposure time obtained experimentally for these aluminum and copper filters to produce beam of exactly the same shape and size was 15.6%, with the former filter requiring more exposure time. [8] This is in reasonable agreement with the theoretical formulations [Eqs. (8) and (9) of this text], which yield 14.5% as the value of D Δ 100% (where Δ = λm t ref ). With the beams from these filters (copper and aluminum at 'spectrally' equivalent thicknesses) numerically made exactly equal in shape and size {φAl (E i ) = αφCu (E i )}, the results obtained by simulations were such that T Cu ~ T Al , DOSE Cu ~ DOSE Al , CNR Cu ~ CNR Al , [CNR Cu ] 2 ~ [CNR Al ] 2 , C Cu ~ C Al and FOM Cu ~ FOM Al.

In the case of beam filtered by 9.3 mm Al in comparison with that filtered by 0.25 mm Cu and for spectral without scaling (same mAs and heat capacity), the values of T Cu , DOSE Cu and [CNR Cu ] 2 exceed those of T Al , DOSE Al and [CNR Al ] 2 by 41.9% respectively while C Cu ~ C Al and FOM Cu ~ FOM Al . Theoretical result presented [8] shows this surplus value to be 41.6%. With the transmitted spectrum through the copper filter theoretically scaled {φAl (E i ) = αφCu (E i )} in comparison with aluminum filter, the results obtained by simulations were such that T Cu ~ T Al , DOSE Cu ~ DOSE Al , [CNR Cu ] 2 ~ [CNR Al ] 2 , CNR Cu ~ CNR Al , C Cu ~ C Al and FOM Cu ~ FOM Al .

Results of spectrometric measurements carried out by Jennings [8] show that 4.08 mmAl, 0.11 mmCu and 0.10 mmY transmit spectrally equivalent beams. For 4.08 mm Al in comparison with 0.11 mm Cu and without the scaling of spectral (same mAs and heat capacity), using Eqs. (8-11), the values of T Cu , DOSE Cu and [CNR Cu ] 2 exceed those of T Al , DOSE Al and [CNR Al ] 2 by 16.2% respectively while C Cu ~ C Al and FOM Cu ~ FOM Al . The same surplus was found for yttrium in comparison with aluminum at the respective 'spectrally' equivalent thicknesses of 0.10 mm Y and 4.08 mm Al while C Y ~ C Al and FOM Y ~ FOM Al . This is in good agreement with the results from the work of Jennings, [8] which reported a difference of 15.3% (experimental) and 16.7% (theoretical) for T Cu and T Al with copper transmitting surplus. Also, reported by the same author [8] was a difference of 15.4% (experimental) and 16.4% (theoretical) for T Y and T Al with yttrium filter transmitting surplus. With the transmitted spectrum through the copper filter scaled (mAs and heat capacity scaled by the factor a), the results obtained from the simulations carried out in this work were such that T Al ~ T Cu ~ T Y , DOSE Al ~ DOSE Cu ~ DOSE Y , [CNR Al ] 2 ~ [CNR Cu ] 2 ~ [CNR Y ] 2 , C Al ~ C Cu ~ C Y and FOM Al ~ FOM Cu ~ FOM Y .

Using scaled spectrum {α φCu (E i )}, operating the tube with copper filter at reduced tube loading/heat capacity (α x mAs Al or 1.35α x mAs Al x kVp) results in same value of T, DOSE, C, CNR and FOM when compared with operating the tube with aluminum filter at higher tube loading/heat capacity (mAs Al or 1.35 mAs Al kVp). Note that whether the beams are scaled or not scaled, once the spectral shape is the same, the values of contrast and FOM will be the same. However, beams filtered by materials with higher values of atomic numbers are more intense (bigger in shape) and produce higher values of CNR and dose to patient than those filtered by material of lower atomic numbers at 'spectrally' equivalent thicknesses when the tube is operated at the same values of kVp, tube loading and heat capacity [Figure - 2],[Figure - 3]. For a given filter material when compared with aluminum/copper at 'spectrally' equivalent thicknesses, only the shapes of transmitted spectra are the same. Depending on the Z -value, the sizes are not the same when the tube is operated at same values of kVp, tube loading and heat capacity [Figure - 3]. At respective 'spectrally' equivalent thicknesses, the benefits derivable from the use of higher Z -value filter material in comparison with those from the use of lower Z -value include higher intensity, lower filter-generated scattered radiation and shorter tube current or exposure time. This can translate into an increase in the lifespan of tube (more filament hours by operating at reduced mAs) and reduction in the cost of dealing with the detrimental effects of heat and scattered radiation. Also, extra images could be obtained using the gain in exposure time and tube loading/heat capacity.

The spectrum obtained by numerically scaling the values of αφCu (E i ) using the factor α (Eq. 14) is in agreement with that transmitted by 3.7 mm Al [Figure - 3]. The scaled spectrum obtained from copper filter {αφCu (E i )} and that directly obtained from aluminum {φAl (E i )} were found to produce the same contrast, CNR, DOSE and FOM. Though not carried out in this work, experimentally at the same kVp, the spectra obtained from filter material with higher Z -value (say copper) in comparison with those with lower Z -value (say aluminum) could be obtained at the scaled intensity and dose (scaled to that obtainable from aluminum filter) by operating the tube at scaled tube loading and heat capacity. This scaling factor provides for the means of obtaining strictly equivalent (same shape and size) spectra from two different elemental filter materials (conventional filters) at the same value of kVp [Figure - 3].

The validity of Eq. (15) had been reported in Okunade [6] using the case of peak voltage of 86 kV and comparison of 0.10 mm Cu with the corresponding 'spectrally' equivalent aluminum thickness, 3.7 mm Al (reference filter). Also, [Figure - 3] shows that the formulation presented as T Al = α x T Cu is valid. Hence all the formulations in Eq. (15) are valid. [Table - 5] shows that the values of 'spectrally' equivalent thicknesses obtained using Eqs.(16-18) are in good agreement with those earlier reported in literature. [8],[9] Apart from providing comparison for aluminum/copper and any given filter material irrespective of atomic number, EQSPECT.FOR can be used to obtain relative performance for any arbitrary pair of filter materials, say silicon and iron. This is accomplishable by specifying the reference filter material as silicon and the alternative filter material as iron.

Generally, it is well known that it is not possible to have a complete spectral matching for conventional and k-edge filter materials. The k-edge discontinuity interferes substantially in the range of useful energies, resulting in beams with significant difference in shape and size. Unlike the case of the comparison of conventional filter materials (aluminum/copper) with other conventional filter materials, large values of the root-mean-square errors [8] were obtained when conventional filter materials were compared with k-edge filter materials. This implies a large difference between the actual transmission of aluminum/copper and the scaled transmission of the k-edge filters. It is not possible to derive simple mathematical formulations of the form with factors α, δ and Δ for the comparison of performance of beams from conventional filter materials with those from k-edge filter materials since spectrals transmitted by these pair of filter materials differ in quality and quantity [Figure - 4]. Theoretically, the transmission below the k-edge in terms of quality and quantity could not be matched in these comparisons. However, using Eq. (9), the results presented in [Table - 6] for the comparison of the overall quantity of photons in terms of the ratios of T and DOSE from beams transmitted by 0.097 mm Cu and 0.088 mm Gd are in reasonable agreement with experimentally measured ratio of exposure time reported by Nagel. [9]


   Conclusions Top


Models and software program have been presented for the comparative evaluation of the performance characteristics (contrast, contrast-to-noise ratio and FOM) of beams transmitted at spectral equivalent thicknesses by elemental filter materials (conventional and k-edge filters). Under this condition of spectral equivalence, the mathematical formulations presented show that when compared with one of lower Z -value, conventional filter material with higher Z -value offers opportunity for extended exposure time. This extended exposure time can translate into acquiring extra images using the gain in exposure time, tube loading and heat capacity. The computer software EQSPECT.FOR developed in this work can be used as a tool for the comparative investigation of filter materials with a view to determine alternative beams that could yield optimum performance in clinical radiological practice. The results from this comparative investigation can aid in the design of X-ray machines in such a way as to produce a reference shape and size of X-ray beam with which others can be compared. This could assist in the achievement of some degree of uniformity in clinical X-ray diagnostic practice.

 
   References Top

1.Boone JM. Equivalent spectra as an index of beam quality. Med Phys 1986;13:861-8.  Back to cited text no. 1    
2.Boone JM. Three parameter equivalent spectra as an index of beam quality. Med Phys 1988;15:304-10.  Back to cited text no. 2    
3.Thoraeus R. The study of the ionization method for measuring the intensity and absorption of Roentgen rays and efficiency of different filters used in therapy. Acta Radiol 1932;15:1-86.  Back to cited text no. 3    
4.Thoraeus R. Tin filters in roentgen therapy. Acta Radiol 1934;15:225-36.  Back to cited text no. 4    
5.Okunade AA. Numerical models for comparing filter materials for diagnostic radiology. Radiat Phys Isot 2002;65:1-9.  Back to cited text no. 5    
6.Okunade AA. Parameterized algorithm for quantitative differentials in spectrally equivalent medical diagnostic x-ray beams. Med Phys 2005;32:1785-95.  Back to cited text no. 6    
7.Nagel HD. Aluminum equivalence of materials used in diagnostic radiology and its dependence on beam quality. Phys Med Biol 1986;31:1381-99.  Back to cited text no. 7    
8.Jennings RJ. A method for comparing beam-hardening filter materials for diagnostic radiology. Med Phys 1988;15:588-99.  Back to cited text no. 8    
9.Nagel HD. Comparison of the performance of characteristics of conventional and k-edge filters in general diagnostic radiology. Phys Med Biol 1989;31:1269-87.  Back to cited text no. 9    
10.Birch R, Marshall M, Ardran GM. Catalogue of spectral data for diagnostic x-rays. The Hospital Physicist's Association Scientific Report Series - 30, 1979.  Back to cited text no. 10    
11.Okunade AA. Effective dose as a limiting quantity for the evaluation of primary barriers for diagnostic x-ray facilities. Health Phys 2005;89:S100-16.  Back to cited text no. 11    
12.Boone JM, Seibert JA. An accurate method for computer-generated tungsten anode x-ray spectra from 30 to 140 kV. Med Phys 1997;24:1661-70.  Back to cited text no. 12    
13.McMaster WH, Del Grande NK, Mallett JH, Hubbell JH. Compilation of X-ray cross section," University of California, Lawrence Radiation Laboratory Report No UCRL-50174, Sec, II, Rev 1 (NTIS), 1969.  Back to cited text no. 13    
14.Hubbell JH, Seltzer SM. Tables of x-ray mass attenuation coefficients and mass energy-absorption coefficients 1 keV to 20 MeV for elements 1-92 and 48 additional substances of dosimetric interest. National Institute of Standard and Technology Internal Report NISTIR 1995. p. 5632.  Back to cited text no. 14    
15.GRAF4WIN. A computer program for graphics and curve fitting GRAPHER() for Windows v1.2 Golden Software Incorporated, Golden, CO: USA; 1993-1994.  Back to cited text no. 15    


    Figures

  [Figure - 1], [Figure - 2], [Figure - 3], [Figure - 4], [Figure - 5]
 
 
    Tables

  [Table - 1], [Table - 2], [Table - 3], [Table - 4], [Table - 5], [Table - 6]



 

Top
Print this article  Email this article
Previous article Next article

    

 
   Search
 
   Next article
   Previous article 
   Table of Contents
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (240 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
    Methods
    Results
    Discussion
    Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed3373    
    Printed135    
    Emailed0    
    PDF Downloaded245    
    Comments [Add]    

Recommend this journal