TECHNICAL NOTE
Year : 2007  |  Volume : 32  |  Issue : 2  |  Page : 73-76
 

A portable organic plastic scintillator dosimetry system for low energy X-rays: A feasibility study using an intraoperative X-ray unit as the radiation source


1 School of Applied Sciences, RMIT University, Melbourne, Victoria 3000, Australia
2 Department of Physical Sciences, Peter MacCallum Cancer Institute, East Melbourne, Victoria 3001, Australia

Date of Submission09-Nov-2006
Date of Acceptance09-Mar-2007

Correspondence Address:
Kerry Williams
School of Applied Sciences, RMIT University, GPO Box 2476V, Melbourne, Victoria 3001
Australia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-6203.33245

Rights and Permissions

 

   Abstract 

The effective use of near water equivalent organic plastic scintillators (OPS) for radiation dosimetry with high-energy sources under laboratory conditions is recognized. In this work, an OPS-based dosimeter using a photodiode combined with improved solid state detection and signal processing techniques has been developed; it offers the potential for the construction of a stable and fully portable dosimeter which will extend the useful range of measurement beyond the usual MeV area and provide reliable readings down to sub-'100 keV' X-ray energy levels. In these experiments, the instrument described has been used for the dosimetry of INTRABEAM intraoperative radiotherapy (IORT) equipment at distances as low as 1.8 mm from the effective source, i.e., 0.2 mm from the X-ray probe surface. Comparison is shown with dosimetry measurements made using the calibrated reference ion chamber supplied by the IORT equipment manufacturer.


Keywords: Dosimeter, low energy intraoperative radiotherapy organic plastic scintillator


How to cite this article:
Williams K, Robinson N, Trapp J, Ackerly T, Das R, Kemp P, Geso M. A portable organic plastic scintillator dosimetry system for low energy X-rays: A feasibility study using an intraoperative X-ray unit as the radiation source. J Med Phys 2007;32:73-6

How to cite this URL:
Williams K, Robinson N, Trapp J, Ackerly T, Das R, Kemp P, Geso M. A portable organic plastic scintillator dosimetry system for low energy X-rays: A feasibility study using an intraoperative X-ray unit as the radiation source. J Med Phys [serial online] 2007 [cited 2019 Oct 17];32:73-6. Available from: http://www.jmp.org.in/text.asp?2007/32/2/73/33245


Radiation dosimetry procedures for intraoperative radiotherapy [1] (IORT) equipment are often performed with parallel plate ionization chamber type dosimeter, a device based on the principle of charges collected due to ionization within their sensitive air volume. [2] The main condition imposed on their satisfactory performance is the fulfillment of charge equilibrium. [2],[3] Thus reliable measurement close to the source can prove problematic. System manufacturer's data is provided only for distances of 5 mm and greater from the X-ray source. [4],[5] Dose measurements as close as 3 mm from the IORT source using a specially modified ion chamber [6] have been presented by Beatty et al. and by Ebert and Carruthers; [7] however, when measuring dose rates under clinical conditions using an 'ion chamber'-based dosimeter, readings may become unreliable as one approaches the near field due to the steep gradients encountered, the critical geometry of the chamber volume and field perturbation effects caused by its presence. The use of an organic plastic scintillator as the radiation detector can overcome these limitations and provide a seamless sequence of readings up to the surface of the IORT probe.

Detectors utilizing organic plastic scintillators are largely immune to environmental variables and also offer the advantage that they do not saturate at high dose rates. OPS offer the advantages of near water equivalence and offer minimal disturbance of the radiation field; they are not influenced by changes in ambient conditions of temperature and pressure, are uncritical as to angle of incidence of radiation [8],[9] and can provide an immediate direct reading. OPS detectors may be machined to very small dimensions of only a few cubic millimeters; yet they offer a wide dynamic response range with high spatial resolution and remain effective at the high dose rates encountered close to the radiation source. Organic plastic scintillators have previously been successfully tested as dosimeters for high-energy radiation, at MeV energies. [9] Details of the IORT probe and dosimetry techniques used in its clinical certification are described by Beatty; [6] they show the position of the gold target within the probe housing and also provide an explanation of the method used to achieve a virtual point source radiator.

The experimental setup of the PRS400 X-ray probe, water phantom and the OPS dosimeter, as depicted in [Figure - 1], shows the basic geometry of the system. With the OPS probe in contact with the beryllium window, the surface of the scintillation material will be ~1.8 mm from the effective source position within the IORT source probe housing. The OPS detector element fabricated from BICRON® BC-430 scintillator material forms the active tip of a probe. It consists of a disk of this material 4.5 mm in diameter and 2.5 mm thick. A 2 mm diameter light guide (BICRON® BC-98) 800 mm long is bonded to a recess machined in the edge of the scintillator disk as shown in [Figure - 1]. The surface of the probe is covered with a black acrylic coating to inhibit the entry of external light and to provide waterproofing.

The remote end of the plastic light-guide, coated with optical grease (BICRON® BC­630), is clamped to the window of a UDT Sensors® 5DPI photodiode, which constitutes the input device for the signal processing circuitry and data acquisition system. Photons enter the window of the photodiode and induce a minute electron current in the analogue buffer stage. The output of the buffer is connected directly to a dedicated analogue-to-digital (A/D) converter operating in the nano-amp region. The A/D conversion concept was developed around the LTC 1799 integrated circuit configured as a current-to-frequency converter and is used to generate a pulse train at TTL level, the frequency of which is proportional to the output signal level from the photodiode. The resulting signal is buffered and transferred to a personal computer via a USB-linked data acquisition system for storage, analysis and display. An overview of the electronic signal processing system and circuit functions is given in [Figure - 2]A, and a typical data stream from the instrument is shown in [Figure - 2]B.

Performance and linearity of the system were checked against the PTW Type N23342 parallel plate ion chamber reference supplied with the IORT equipment. The system was tested at various beam currents at a fixed distance from the source in air, and a further series of measurements was made at a fixed beam voltage and current in the water tank at varying distances from the source. For these tests, the probe element was attached to the outer surface of a plastic (solid water) holder housing the parallel plate ionization chamber. The probe disk was aligned edge-on to the radiation source and the complete assembly immersed in the water tank. The position of the OPS detector and ion chamber relative to the surface of the radiation source was adjusted in the water tank by use of external control arms that allow movement in three planes to an accuracy of ±0.25 mm.


   Results and Discussion Top


Prior to carrying out the series of measurements in the water tank, the apparatus was 'bench-tested' at 50 kVp, which is the set energy level used in the clinical situation. Ion chamber readings and OPS dosimeter measurements at a distance of 2 mm from the source surface (in air) were logged and compared over the available range of tube currents - from 5 mA to 40 mA. Numerical data showing the dose measured with the ion chamber and the equivalent output reading from the OPS dosimeter are shown in [Table - 1]. A graphical presentation of the OPS data from [Table - 1] is shown in [Figure - 3], together with dose readings from the ion chamber reference. A calibration factor for the particular configuration can be calculated from the data, which will allow interpolation of readings across the energy range of interest. When carrying out measurements over a broad range of X-ray energies or when physically changing the OPS probe assembly, a new calibration factor for the particular probe configuration would need to be established and entered into the data processing function.

A further series of measurements were taken in the water tank, and the results are shown in [Table - 2]. This data is presented graphically in [Figure - 4]. Note that the closest distance from the active area of the plastic scintillator to the surface of the source window is limited by the thickness of the titanium dioxide layer as also by the light blocking the plastic film covering the scintillator. The total thickness of the two layers is approximately 0.2 mm.

The measurements extend from this lower limit out to a distance of 16 mm from the source. Beyond this point when using the selected probe of volume 40 mm 3 , OPS dosimeter readings progressively fall to a level where due to the small volume of the scintillator element, background noise levels become significant.

Dose measurements previously made using a liquid water phantom by Beatty et al. were found to closely approximate an inverse cube function with distance. [6] A confirmation of this characteristic can be observed in the output of the OPS dosimeter plotted in [Figure - 4], which shows a plot of the dose rates measured by the OPS detector and ion chamber for distances up to 12 mm from the source surface. Across this range, the plot indicates very close agreement between the OPS measurements and the reference ion chamber readings. The resulting curves have been overlaid such that the correct relationship to the effective source position is maintained and additional measurements taken with the OPS dosimeter extend the readings down to the minimum possible distance of ~ 0.2 mm from the surface of the source probe. It can be seen that the curve continues to describe an inverse cube function, as would be expected.

Beyond 15 mm from the source, due to the relatively small volume of scintillator element used, the falling signal-to-noise ratio begins to limit resolution. Since dose rate measurement at these distances can effectively be covered using the ionization chamber method, the graphing of OPS dosimeter readings was limited in these tests to distances closer than 12 mm from the source and shows monotonic increase in dose rate up to about 65 Gy/min at the closest distance of approximately 0.2 mm from the external surface of the INTRABEAM X-ray probe.


   Conclusion Top


X-radiation emission from the PRS 400 intraoperative unit was measured using an organic plastic scintillator dosimeter over distances ranging from 0.2 mm to beyond 12 mm from the source surface. IORT probe manufacturer's data, measured using an ion chamber dosimeter, is provided for distances of 5 mm and greater from the X-ray source. The use of the instrument outlined in this work offers the possibility of another method of dosimetry for the PRS400 system and also extends the range of available dose rate readings down to distances of less than 5 mm from the source. The advantageous characteristics of OPS detectors allow reliable measurements to be made right up to the radiation window, where steep dose gradients and the beam-distorting effects of other types of dosimeters render results questionable. The facility to achieve reliable near water equivalent dose measurements at close distances from the radiation source using a portable detection instrument that is simple to apply and operate could be of considerable interest to users of IORT systems.

 
   References Top

1.Vaidya JS, Baum M, Tobias JS, D'Souza DP, Naidu SV, Morgan S, et al . Targeted intra-operative radiotherapy (Targit): An innovative method of treatment for early breast cancer. Ann Oncol 2001;12:1075-80.  Back to cited text no. 1  [PUBMED]  [FULLTEXT]
2.Hendee WR. Radiation therapy. Phys. 2 nd ed. Mosby: Ch.5 1996. p. 116.  Back to cited text no. 2    
3.Khan FM. The physics of radiation therapy. 2 nd ed. Williams and Wilkins: 2003, Ch.6.  Back to cited text no. 3    
4.Dinsmore M, Harte KJ, Sliski AP, Smith DO, Nomikos PM, Dalterio MJ, et al . A new miniature X-ray source for interstitial radiosurgery: Device description. Med Phy 1996;23:45-52.  Back to cited text no. 4    
5.Available from: http://www.zeiss.de.  Back to cited text no. 5    
6.Beatty J, Biggs PJ, Gall K, Okunieff P, Pardo FS, Harte KJ, et al . A new miniature X-ray device for interstitial radiosurgery: Dosimetry. Med Phy 1996;23:53-62.  Back to cited text no. 6    
7.Ebert MA, Carruthers B. Dosimetric characteristics of a low-kV intra-operative X-ray source: Implications for use in a clinical trial for treatment of low-risk breast cancer. Med Phys 2003;30:2424-31.  Back to cited text no. 7    
8.Clift MA, Sutton RA, Webb DV. Water-equivalence of plastic scintillators in megavoltage radiotherapy bremsstrahlung beams. Phys Med Biol 2000;45:1885-95.  Back to cited text no. 8    
9.Beddar AS, Mackie TR, Attix FH. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration. Phys Med Biol 1992;37:1883-900.  Back to cited text no. 9    


    Figures

  [Figure - 1], [Figure - 2], [Figure - 3], [Figure - 4]
 
 
    Tables

  [Table - 1], [Table - 2]


This article has been cited by
1 A SiPM based real time dosimeter for radiotherapic beams
A. Berra,V. Conti,D. Lietti,L. Milan,C. Novati,A. Ostinelli,M. Prest,C. Romanó,E. Vallazza
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2014;
[Pubmed] | [DOI]
2 Depth dose determination for a mixed radiation field using a thin plastic scintillator dosimetry system
Aydarous, A., Waker, A.
Nuclear Technology and Radiation Protection. 2012; 27(1): 20-27
[Pubmed]
3 A hybrid radiation detector for simultaneous spatial and temporal dosimetry
C. Poole, J. V. Trapp, J. Kenny, T. Kairn, K. Williams, M. Taylor, R. Franich, C. M. Langton
Australasian Physical & Engineering Sciences in Medicine. 2011;
[VIEW] | [DOI]
4 Internal calibration of gel dosimeters: A feasibility study
J V Trapp,T Kairn,S Crowe,A Fielding
Journal of Physics: Conference Series. 2009; 164: 012014
[Pubmed] | [DOI]
5 Internal calibration of gel dosimeters: A feasibility study
Trapp, J.V., Kairn, T., Crowe, S., Fielding, A.
Journal of Physics: Conference Series. 2009; 164(Art 012014)
[Pubmed]
6 Design of a versatile and low cost μvolt level A to D conversion system for use in medical instrumentation applications
Williams, K., Robinson, N.
Eurasip Journal on Advances in Signal Processing. 2008; Art 370171
[Pubmed]



 

Top
Print this article  Email this article
Previous article Next article

    

 
   Search
 
   Next article
   Previous article 
   Table of Contents
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (149 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
    Results and Disc...
    Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed4080    
    Printed120    
    Emailed2    
    PDF Downloaded289    
    Comments [Add]    
    Cited by others 6    

Recommend this journal